Evaluation of the Comparability of Wantai Wan200+ Instrument with Routine Laboratory Assays for 21 Different Analytes
Ilaria Talli, Andrea Padoan, Chiara Cosma, Giulia Furlan, Martina Zaninotto, Lucio Marchioro, Paola Galozzi, Daniela Basso, Mario Plebani

TL;DR
This study compared the Wantai Wan200+ instrument with other lab methods for 21 analytes and found it to be comparable for most, though some showed suboptimal results.
Contribution
The study provides a comprehensive evaluation of the Wantai Wan200+ instrument's performance against established methods for multiple analytes.
Findings
An excellent comparability profile was found for 11 analytes, such as total PSA.
Ten analytes showed suboptimal comparability, indicating potential issues with consistency.
Biological variation metrics were used to assess the clinical relevance of observed biases.
Abstract
Background: We compared the performance of 21 different assays performed by the Wantai Wan200+ (Wantai BioPharm, Beijing, China) with respect to other methods in use at the University Hospital of Padova (AOPD), Italy. Methods: The plasma (P) or serum (S) of 5027 leftover samples, collected from May to Sept 2023, was either analyzed or frozen at −20 °C. Beckman DXI800 (DXI), Roche Cobas 8000 e801 (RC), Snibe Maglumi 4000 plus (SM), DiaSorin Liaison XL (DL) and Binding Site Optilite (BS) equipment were used at the AOPD. P-procalcitonin (PCT), DXI; P-Troponin I (TnI), DXI; S-CA125, DXI; S-free PSA (f-PSA), DXI; S-total PSA (t-PSA), DXI; S-IL6, SM; P-Troponin T (TnT), RC; P-NT-proBNP, RC; P-Neuron-Specific Enolase (NSE), RC; S-CA15-3, DL; S-CA19-9, DL; S-AFP, DL; and S-CEA, DL were tested in fresh samples. P-Myoglobin (Myo), DXI; P-Cyfra21-1, RC; S-β2 microglobulin (B2MIC), BS; S-HE4, SM;…
Genes, proteins, chemicals, diseases, species, mutations and cell lines named across the full text — each resolved to its canonical identifier and authoritative record.
Click any figure to enlarge with its caption.
Figure 1
Figure 2
Figure 3Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsClinical Laboratory Practices and Quality Control · Reliability and Agreement in Measurement · Meta-analysis and systematic reviews