# Stability of Proximal Femoral Osteotomies in Pediatric Bone Models Fixed with Flexible Intramedullary Nails and Evaluated by the Finite Element Method

**Authors:** Mário Augusto Ferreira Cruz, José Vinícius Lima Santana, Leonardo Rigobello Battaglion, José Batista Volpon

PMC · DOI: 10.1055/s-0044-1785467 · 2024-04-10

## TL;DR

This study uses computer modeling to show that flexible nails are less stable for fixing bone cuts near the hip in children compared to cuts in the thigh bone shaft.

## Contribution

The paper introduces a finite element analysis of flexible intramedullary nail stability in pediatric proximal femoral osteotomies.

## Key findings

- Trochanteric osteotomies showed 40% higher maximum stress compared to diaphyseal osteotomies.
- Subtrochanteric osteotomies had the largest displacement (0.5mm) under load.
- Flexible intramedullary nails showed lower stability in proximal femoral regions.

## Abstract

Objective
 To evaluate the stability of osteotomies created in the subtrochanteric and trochanteric regions in a pediatric femur model fixed by flexible intramedullary rods.

Method
 Tomographic sections were obtained from a pediatric femur model with two elastic titanium rods and converted to a three-dimensional model. This model created a mesh with tetrahedral elements according to the finite element method. Three virtual models were obtained, and osteotomies were performed in different regions: mediodiaphyseal, subtrochanteric, and trochanteric. A vertical load of 85N was applied to the top of the femoral head, obtaining the displacements, the maximum and minimum main stress, and the equivalent Von Mises stress on the implant.

Results
 With the applied load, displacements were observed at the osteotomy site of 0.04 mm in the diaphyseal group, 0.5 mm in the subtrochanteric group, and 0.06 mm in the trochanteric group. The maximum stress in the diaphyseal, subtrochanteric, and trochanteric groups was 10.4 Pa, 7.52 Pa, and 26.4 Pa, respectively. That is around 40% higher in the trochanteric group in regards to the diaphyseal (control). The minimum stress of the bone was located in the inner cortical of the femur. The equivalent Von Mises stress on the implants occurred at osteotomy, with a maximum value of 27.6 Pa in the trochanteric group.

Conclusion
 In both trochanteric and subtrochanteric osteotomies, fixation stability was often lower than in the diaphyseal model, suggesting that flexible intramedullary nails are not suitable implants for proximal femoral fixations.

Objetivo
 Avaliar a estabilidade de osteotomias criadas nas regiões subtrocantérica e trocantérica em modelo de fêmur pediátrico, fixadas por hastes intramedulares flexíveis.

Método
 A partir de um modelo de fêmur pediátrico com duas hastes elásticas de titânio, foram obtidos cortes tomográficos que foram convertidos para um modelo tridimensional. Neste modelo foi criado uma malha com elementos tetraédricos, de acordo com o método dos elementos finitos. Foram obtidos três modelos virtuais, e realizadas osteotomias em regiões diferentes: mediodiafisária, subtrocantérica e trocantérica. Foi aplicado um carregamento vertical de 85N no topo da cabeça do fêmur, obtidos os deslocamentos, a tensão máxima e mínima principal e tensão equivalente de Von Mises no implante.

Resultados
 Com o carregamento aplicado foram observados deslocamentos no local da osteotomia de 0,04mm no grupo diafisário, 0,5mm no subtrocantérico e 0,06mm no trocantérico. A tensão máxima principal foi 10,4Pa, 7,52Pa e 26,4Pa nos grupos diafisário, subtrocantérico e trocantérico, respectivamente. Ou seja, a tensão máxima foi em torno de 40% maior no grupo trocantérico, em relação ao diafisário (controle). A face de tensão mínima do osso localizou-se na cortical interna do fêmur. A tensão equivalente de Von Mises nos implantes ocorreu na osteotomia, com valor máximo de 27,6Pa no grupo trocantérico.

Conclusão
 Tanto nas osteotomias no nível trocantérico, quanto subtrocantérico, a estabilidade da fixação foi muitas vezes menor que no modelo diafisário, sugerindo que as hastes intramedulares flexíveis não são implantes adequados para as fixações proximais do fêmur.