# Clinical compatibility of magnetic resonance imaging with magnetic intramedullary nails: a feasibility study

**Authors:** Joseph D. Femino, Samuel R. Barnes, Scott C. Nelson, Lee M. Zuckerman

PMC · DOI: 10.1007/s00402-024-05210-y · Archives of Orthopaedic and Trauma Surgery · 2024-02-14

## TL;DR

This study shows MRI is generally safe with magnetic intramedullary nails, though some precautions are needed.

## Contribution

The study provides clinical evidence that MRI can be safely used with magnetic nails, contradicting manufacturer warnings.

## Key findings

- No adverse symptoms or hardware complications were observed in 12 patients who underwent MRI with magnetic nails.
- MRI scans at both 1.5 T and 3.0 T did not cause implant activation or migration in clinical settings.
- Patients spent an average of 84.7 minutes in the MRI suite without complications.

## Abstract

The use of magnetic resonance imaging (MRI) with a magnetic intramedullary lengthening nail in place is contraindicated per the manufacturer due to the concern of implant activation and migration. A prior in vitro study did not confirm these complications only noting that a 3.0 T MRI weakened the internal magnet. Therefore, a retrospective analysis of patients who underwent an MRI with a magnetic nail in place was performed to determine if any adverse effects occurred in the clinical setting.

A retrospective review of all patients who underwent an MRI with a magnetic lengthening nail in place was performed. The time spent being imaged in the MRI, number of times the patient entered the MRI suite, and the images obtained were recorded. Radiographs were performed before and after the MRI to determine if any hardware complications occurred. The patients were monitored for any adverse symptoms while they were in the suite.

A total of 12 patients with 13 nails were identified. Two patients underwent imaging with a 3.0 T MRI while the remaining 10 underwent imaging with a 1.5 T MRI. Each patient entered the MRI suite 2.1 times and spent an average of 84.7 min being imaged in the MRI (range 21–494). No patients noted any adverse symptoms related to the nail while in the suite and no hardware complications were identified.

MRI appears to be safe with a magnetic nail in place and did not result in any complications. Given the manufacturer’s recommendations, informed consent should be obtained prior to an MRI being performed and a 3.0 T MRI should be avoided when possible if further activation of the nail is required.

## Full-text entities

- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC10965604/full.md

## Figures

3 figures with captions in the complete paper: https://tomesphere.com/paper/PMC10965604/full.md

## References

41 references — full list in the complete paper: https://tomesphere.com/paper/PMC10965604/full.md

---
Source: https://tomesphere.com/paper/PMC10965604