# The Effect of Prone and Supine Limb Positioning on the Radiographic Evaluation of Posterolateral Plate Fixation of the Posterior Malleolus

**Authors:** Preemal Patel, Zoe Little, Phillip Beak, Rachel Williams, Alex Trompeter

PMC · DOI: 10.1007/s43465-023-01066-3 · Indian Journal of Orthopaedics · 2024-01-22

## TL;DR

This study shows that knee flexion during prone imaging affects how implant positions appear in radiographs, which could impact surgical decisions.

## Contribution

The study identifies that knee flexion, not prone positioning itself, causes radiographic discrepancies in implant placement.

## Key findings

- No significant difference in plate tip-joint line distance between prone and supine views.
- Knee flexion in prone position causes the plate tip to appear 6.5–8.5 mm more proximal in radiographs.
- Radiographic discrepancies are due to fluoroscopy beam angle changes from knee flexion, not projectional differences.

## Abstract

To facilitate the posterolateral approach to the posterior malleolus patients are often positioned prone initially, then turned supine to complete fixation at the medial malleolus. We sought to define observed differences in the radiographic appearance of implants relative to the joint line, in prone and supine positions.

A 3.5 mm tubular plate and a 3.5 mm posterior distal tibial periarticular plate were applied sequentially to 3 individual cadaveric legs, via a posterolateral approach. The tubular plate was positioned to simulate buttress fixation and the posterolateral plate placed more distally. Each limb was secured on a custom jig and radiographs were taken on a mobile c-arm fluoroscopy machine with a calibration ball. A series of prone AP, supine PA and mortise radiographs were taken. Prone radiographs were also taken in different degrees of caudal tilt to simulate knee flexion which occurs in practice, during intraoperative positioning. Plate tip-joint line distances were measured and Mann–Whitney U tests performed.

There was no statistically significant difference in plate tip-joint line distance when comparing equivalent prone and supine views (PA/AP or mortise). However, significant differences in apparent implant position were noted with alterations in caudal tilt. When taking a prone image, when the knee is flexed to 20 degrees, the plate tip will appear 6.5–8.5 mm more proximal than in the equivalent supine image where the knee is extended and the fluoroscopy beam is orthogonal to the anatomic axis of the tibia.

Observed differences in radiographic appearance of metalwork in the prone and supine position are most likely due to knee flexion and the resulting variation in the angle of the fluoroscopy beam, rather than projectional differences between supine and prone views. Surgeons should be alert to this when analysing intraoperative images.

## Full-text entities

- **Diseases:** Malleolus (MESH:D064386)
- **Species:** Homo sapiens (human, species) [taxon 9606]

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/PMC10899138/full.md

## Figures

4 figures with captions in the complete paper: https://tomesphere.com/paper/PMC10899138/full.md

## References

8 references — full list in the complete paper: https://tomesphere.com/paper/PMC10899138/full.md

---
Source: https://tomesphere.com/paper/PMC10899138