# Nuclear quadrupole resonance spectroscopy with a femtotesla diamond   magnetometer

**Authors:** Yaser Silani, Janis Smits, Ilja Fescenko, Michael W. Malone, Andrew F., McDowell, Andrey Jarmola, Pauli Kehayias, Bryan Richards, Nazanin Mosavian,, Nathaniel Ristoff, Victor M. Acosta

arXiv: 2302.12401 · 2025-04-09

## TL;DR

This paper demonstrates a femtotesla sensitivity RF magnetometer based on NV-doped diamond, enabling detection of nuclear quadrupole resonance signals at room temperature with potential applications in various fields.

## Contribution

The authors developed a femtotesla RF magnetometer using NV-doped diamond and flux concentrators, achieving unprecedented sensitivity and successful NQR detection at room temperature.

## Key findings

- Sensitivity of ~70 fT/Hz^{1/2} at 0.35 MHz
- Detection of 14N NQR signal in sodium nitrite at room temperature
- NQR frequency shifts linearly with temperature (-1.00 kHz/K)

## Abstract

Sensitive Radio-Frequency (RF) magnetometers that can detect oscillating magnetic fields at the femtotesla level are needed for demanding applications such as Nuclear Quadrupole Resonance (NQR) spectroscopy. RF magnetometers based on Nitrogen-Vacancy (NV) centers in diamond have been predicted to offer femtotesla sensitivity, but published experiments have largely been limited to the picotesla level. Here, we demonstrate a femtotesla RF magnetometer based on an NV-doped diamond membrane inserted between two ferrite flux concentrators. The device operates in bias magnetic fields of 2-10 microtesla and provides a ~300-fold amplitude enhancement within the diamond for RF magnetic fields in the 0.07-3.6 MHz range. The magnetometer's sensitivity is ~70 fT s^{1/2} at 0.35 MHz, and the noise floor decreases to below 2 fT after 1 hour of acquisition. We used this sensor to detect the 3.6 MHz NQR signal of 14N in sodium nitrite powder at room temperature. NQR signals are amplified by a resonant RF coil wrapped around the sample, allowing for higher signal-to-noise ratio detection. The diamond RF magnetometer's recovery time after a strong RF pulse is ~35 us, limited by the coil ring-down time. The sodium-nitrite NQR frequency shifts linearly with temperature as -1.00 +/- 0.02 kHz/K, the magnetization dephasing time is T2* = 887 +/- 51 us, and a spin-lock spin-echo pulse sequence extends the signal lifetime to 332 +/- 23 ms, all consistent with coil-based NQR studies. Our results expand the sensitivity frontier of diamond magnetometers to the femtotesla range, with potential applications in security, medical imaging, and materials science.

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/2302.12401/full.md

## Figures

15 figures with captions in the complete paper: https://tomesphere.com/paper/2302.12401/full.md

## References

75 references — full list in the complete paper: https://tomesphere.com/paper/2302.12401/full.md

---
Source: https://tomesphere.com/paper/2302.12401