An automated QC Station for the Calibration of the Mu2e Calorimeter Readout Units
E. Sanzani, C. Bloise, S. Ceravolo, F. Cervelli, F. Colao, M., Cordelli, G. Corradi, S. Di Falco, E. Diociaiuti, S. Donati, C. Ferrari, R., Gargiulo, A. Gioiosa, S. Giovannella, V. Giusti, D. Hampai, F. Happacher, M., Martini, S. Miscetti, L. Morescalchi, D. Paesani, D. Pasciuto

TL;DR
This paper presents an automated QC station designed for calibrating Mu2e calorimeter readout units, evaluating key parameters to ensure quality and performance.
Contribution
The paper introduces a novel automated QC station for Mu2e calorimeter units, detailing its hardware, software, and initial testing results.
Findings
Successful calibration of Readout Units using the QC station
Evaluation of gain, charge, and photon detection efficiency
First measurement results demonstrating system effectiveness
Abstract
The Mu2e calorimeter will employ Readout Units, each made of two Silicon Photomultipliers arrays and two Front End Electronics boards. To calibrate them, we have designed, assembled and put in operation an automated Quality Control (QC) station. Gain, collected charge and photon detection efficiency are evaluated for each unit. In this paper, the QC Station is presented, in its hardware and software aspects, summarizing also the tests performed on the ROUs and the first measurement results.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsRadiation Detection and Scintillator Technologies · Particle Detector Development and Performance · Neutrino Physics Research
