TL;DR
This paper validates the Fastcat CBCT simulator against real kV and MV CBCT images from a Varian Truebeam linac, demonstrating its accuracy and speed for clinical imaging simulations.
Contribution
It provides the first experimental validation of the Fastcat CBCT simulator with detailed modeling and comparison to actual linac images, confirming its accuracy and efficiency.
Findings
Contrast within 14 and 9 HUs for kV and MV simulations
Detector MTF within 4.2% and 2.5%
Dose within 2.4% and 1.6% of MC calculations
Abstract
Purpose: To experimentally validate the Fastcat cone beam CT (CBCT) simulator against kV and MV CBCT images acquired with a Varian Truebeam linac. Methods: kV and MV CBCT images of a Catphan 504 phantom were acquired using a 100 kVp beam with the on-board imager (OBI) and a 6 MV treatment beam with the electronic portal imaging device (EPID), respectively. The kV Fastcat simulation was performed using detailed models of the x-ray source, bowtie filter, a high resolution voxelized virtual Catphan phantom, anti-scatter grid, and the CsI scintillating detector. Likewise, an MV Fastcat CBCT was simulated with detailed models for the beam energy spectrum, flattening filter, a high resolution voxelized virtual Catphan phantom, and the GOS scintillating detector. Experimental and simulated CBCT images of the phantom were compared with respect to HU values, contrast to noise ratio (CNR),and…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Code & Models
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.