Interpretation of Confocal ISO 21073: 2019 confocal microscopes: Optical data of fluorescence confocal microscopes for biological imaging- Recommended Methodology for Quality Control
Glyn Nelson, Laurent Gelman, Orestis Faklaris, Roland Nitschke, Alex, Laude

TL;DR
This paper explains how to perform quality control on confocal microscopes according to ISO 21073 standards, ensuring optimal performance for biological imaging.
Contribution
It provides a detailed methodology for implementing ISO 21073 QC procedures and suggests additional methods for comprehensive microscope assessment.
Findings
Established a standard QC protocol for confocal microscopes
Outlined additional QC methods for enhanced system evaluation
Emphasized importance of maintenance for optimal imaging performance
Abstract
The performance of a confocal imaging system may be no better than a general-purpose widefield system if it is not properly maintained or quality controlled. The publication of ISO 21073, 'Confocal microscopes- Optical data of fluorescence confocal microscopes for biological imaging', set a standard for the minimal Quality Control (QC) that should be performed for Confocal Microscopes. Here we describe methodology for performing the QC requirements to satisfy ISO 21073, as well as suggesting other QC methods that should be performed to obtain a minimum level of information about the microscope system.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsAdvanced Fluorescence Microscopy Techniques · Optical Imaging and Spectroscopy Techniques · Photoacoustic and Ultrasonic Imaging