Evaluation Of The PETsys TOFPET2 ASIC In Multi-Channel Coincidence Experiments
Vanessa Nadig, David Schug, Bjoern Weissler, and Volkmar Schulz

TL;DR
This study evaluates the TOFPET2 ASIC's multi-channel timing performance in PET systems, demonstrating promising coincidence time resolutions and analyzing the impact of configuration parameters across different SiPM and scintillator setups.
Contribution
It provides a comprehensive analysis of the TOFPET2 ASIC's performance with various configurations and SiPM types, extending previous studies with detailed multi-channel results.
Findings
Achieved CRTs as low as 219.9 ps with Hamamatsu SiPMs.
Achieved CRTs as low as 216.1 ps with Broadcom SiPMs.
Identified the trigger delay and integrator gain as key performance factors.
Abstract
Aiming to measure the difference in arrival times of two coincident gamma-photons with an accuracy in the order of 200 ps, time-of-flight positron emission tomography systems commonly employ silicon photomultipliers (SiPMs) and high-resolution digitization electronics, application specific integrated circuits (ASICs). This work evaluates the performance of the TOFPET2 ASIC, released by PETsys Electronics S.A. in 2017, dependent on its configuration parameters in multi-channel coincidence measurements. SiPM arrays fabricated by different vendors (KETEK, SensL, Hamamatsu, Broadcom) were tested in combination with the ASIC. Scintillator arrays featuring different reflector designs and different configurations of the TOFPET2 ASIC software parameters were evaluated. The benchtop setup used is provided with the TOFPET2 ASIC evaluation kit by PETsys Electronics S.A.. Compared to existing…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.