Calibration of the charge and energy loss per unit length of the MicroBooNE liquid argon time projection chamber using muons and protons
MicroBooNE collaboration: C. Adams, M. Alrashed, R. An, J. Anthony, J., Asaadi, A. Ashkenazi, S. Balasubramanian, B. Baller, C. Barnes, G. Barr, V., Basque, M. Bass, F. Bay, S. Berkman, A. Bhanderi, A. Bhat, M. Bishai, A., Blake, T. Bolton, L. Camilleri, D. Caratelli

TL;DR
This paper presents a data-driven calibration method for the MicroBooNE liquid argon TPC, improving energy measurement accuracy and particle identification by correcting for various detector effects using cosmic-ray muons and stopping protons.
Contribution
The paper introduces a novel calibration technique that accounts for multiple detector effects, enhancing energy scale precision and particle identification in liquid argon TPCs.
Findings
Achieved 2% absolute energy scale uncertainty in data.
Improved proton selection efficiency by 2%.
Enhanced energy loss measurement accuracy for ionizing particles.
Abstract
We describe a method used to calibrate the position- and time-dependent response of the MicroBooNE liquid argon time projection chamber anode wires to ionization particle energy loss. The method makes use of crossing cosmic-ray muons to partially correct anode wire signals for multiple effects as a function of time and position, including cross-connected TPC wires, space charge effects, electron attachment to impurities, diffusion, and recombination. The overall energy scale is then determined using fully-contained beam-induced muons originating and stopping in the active region of the detector. Using this method, we obtain an absolute energy scale uncertainty of 2\% in data. We use stopping protons to further refine the relation between the measured charge and the energy loss for highly-ionizing particles. This data-driven detector calibration improves both the measurement of total…
| MC | Data | |
|---|---|---|
| Fitted value of | (5.077 0.001) ADC/e | (4.113 0.011) ADC/e |
| 15.0/18 0.84 | 5.12/18 0.28 |
| particle type | purity | efficiency | purity before | efficiency before |
|---|---|---|---|---|
| calibration | calibration | |||
| selected muons | ||||
| selected protons |
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Calibration of the charge and energy loss per unit length of the MicroBooNE liquid argon time projection chamber using muons and protons
C. Adams
M. Alrashed
R. An
J. Anthony
J. Asaadi
A. Ashkenazi
S. Balasubramanian
B. Baller
C. Barnes
G. Barr
V. Basque
M. Bass
F. Bay
S. Berkman
A. Bhanderi
A. Bhat
M. Bishai
A. Blake
T. Bolton
L. Camilleri
D. Caratelli
I. Caro Terrazas
R. Carr
R. Castillo Fernandez
F. Cavanna
G. Cerati
Y. Chen
E. Church
D. Cianci
E. O. Cohen
J. M. Conrad
M. Convery
L. Cooper-Troendle
J. I. Crespo-Anadón
M. Del Tutto
D. Devitt
A. Diaz
L. Domine
K. Duffy
S. Dytman
B. Eberly
A. Ereditato
L. Escudero Sanchez
J. Esquivel
J. J. Evans
R. S. Fitzpatrick
B. T. Fleming
N. Foppiani
D. Franco
A. P. Furmanski
D. Garcia-Gamez
S. Gardiner
V. Genty
D. Goeldi
S. Gollapinni
O. Goodwin
E. Gramellini
P. Green
H. Greenlee
R. Grosso
L. Gu
W. Gu
R. Guenette
P. Guzowski
P. Hamilton
O. Hen
C. Hill
G. A. Horton-Smith
A. Hourlier
E.-C. Huang
R. Itay
C. James
J. Jan de Vries
X. Ji
L. Jiang
J. H. Jo
R. A. Johnson
J. Joshi
Y.-J. Jwa
G. Karagiorgi
W. Ketchum
B. Kirby
M. Kirby
T. Kobilarcik
I. Kreslo
I. Lepetic
Y. Li
A. Lister
B. R. Littlejohn
S. Lockwitz
D. Lorca
W. C. Louis
M. Luethi
B. Lundberg
X. Luo
A. Marchionni
S. Marcocci
C. Mariani
J. Marshall
J. Martin-Albo
D. A. Martinez Caicedo
K. Mason
A. Mastbaum
N. McConkey
V. Meddage
T. Mettler
K. Miller
J. Mills
K. Mistry
A. Mogan
T. Mohayai
J. Moon
M. Mooney
C. D. Moore
J. Mousseau
M. Murphy
R. Murrells
D. Naples
R. K. Neely
P. Nienaber
J. Nowak
O. Palamara
V. Pandey
V. Paolone
A. Papadopoulou
V. Papavassiliou
S. F. Pate
A. Paudel
Z. Pavlovic
E. Piasetzky
D. Porzio
S. Prince
G. Pulliam
X. Qian
J. L. Raaf
A. Rafique
L. Ren
L. Rochester
H.E. Rogers
M. Ross-Lonergan
C. Rudolf von Rohr
B. Russell
G. Scanavini
D. W. Schmitz
A. Schukraft
W. Seligman
M. H. Shaevitz
R. Sharankova
J. Sinclair
A. Smith
E. L. Snider
M. Soderberg
S. Söldner-Rembold
S. R. Soleti
P. Spentzouris
J. Spitz
M. Stancari
J. St. John
T. Strauss
K. Sutton
S. Sword-Fehlberg
A. M. Szelc
N. Tagg
W. Tang
K. Terao
R. T. Thornton
M. Toups
Y.-T. Tsai
S. Tufanli
T. Usher
W. Van De Pontseele
R. G. Van de Water
B. Viren
M. Weber
H. Wei
D. A. Wickremasinghe
Z. Williams
S. Wolbers
T. Wongjirad
K. Woodruff
M. Wospakrik
W. Wu
T. Yang
G. Yarbrough
L. E. Yates
G. P. Zeller
J. Zennamo
C. Zhang
Abstract
We describe a method used to calibrate the position- and time-dependent response of the MicroBooNE liquid argon time projection chamber anode wires to ionization particle energy loss. The method makes use of crossing cosmic-ray muons to partially correct anode wire signals for multiple effects as a function of time and position, including cross-connected TPC wires, space charge effects, electron attachment to impurities, diffusion, and recombination. The overall energy scale is then determined using fully-contained beam-induced muons originating and stopping in the active region of the detector. Using this method, we obtain an absolute energy scale uncertainty of 2% in data. We use stopping protons to further refine the relation between the measured charge and the energy loss for highly-ionizing particles. This data-driven detector calibration improves both the measurement of total deposited energy and particle identification based on energy loss per unit length as a function of residual range. As an example, the proton selection efficiency is increased by 2% after detector calibration.
1 Introduction
Liquid argon time projection chamber (LArTPC) technology provides both particle tracking and energy loss per unit length reconstruction with high resolution. The basic working principle of a LArTPC neutrino detector is that neutrinos first interact with an argon nucleus and produce charged and neutral secondary particles. The charged secondary particles travel through liquid argon and mainly lose their energy by ionizing and exciting argon atoms. The ionization electrons travel under an applied electric field to a set of anode wire planes. The charge is measured on the anode wire planes in order to reconstruct particle trajectories and energies. Excited argon atoms also produce scintillation light, which is detected by the photon detectors. In this paper, we address the calibration of the charge collected at the anode wires for the MicroBooNE LArTPC [1].
MicroBooNE is a 170-ton LArTPC with dimensions m3 (horizontal drift dimensionheightlength). The nominal electric field inside the TPC is 0.273 kV/cm, which leads to a nominal electron drift velocity of 0.11 cm/s. The drifted charge from particle interactions is read out in three planes with a plane spacing and a wire pitch of 3 mm. The 3456 collection plane wires are vertical and the angle between the induction and collection plane wires is 60 degrees. The readout window size is 4.8 ms and the ADC sampling rate is 2 MHz. Scintillation photons are observed by 32 photo-multipliers (PMTs) [2]. MicroBooNE started collecting neutrino events in Booster Neutrino Beam (BNB) at Fermilab in October 2015.
The first step of energy loss per unit length () reconstruction in LArTPC detectors involves the extraction of charge information from the signals (waveforms) on the anode plane wires. The MicroBooNE experiment uses several signal processing techniques [3, 4] including noise filtering and signal deconvolution for charge extraction. However, the total charge extracted in this way normally does not equal the total charge produced from ionization for a number of reasons. The reasons are distortions in detector response due to cross-connected TPC channels [5], space charge effects (SCE) [6], electron attachment to impurities [7, 8, 9, 10, 11], diffusion [12] and recombination [13, 14]. To trace back the exact amount of charge released from the original interaction, we have to correct for each of these effects starting from the ionization charge per unit length in the direction of the particle’s travel, , as reconstructed from the signal collected on the TPC wires. Throughout this paper refers to the length of a short segment of the particle track, which should not be confused with an infinitesimal distance in x (drift) direction of the TPC.
The data-driven detector calibration consists of several steps. In the first step we correct the position- and time-dependence of the detector response to ionization charge using data from cosmic ray muons (CR) which enter the detector from either the anode or cathode side and exit through the opposite face (\saycrossing CR). This correction process is known as the calibration or equalization calibration of the detector. Once the detector response is corrected to be uniform throughout the TPC and in time, we determine a calibration for converting to , by using stopping muons from neutrino interactions. Here stopping muons from neutrino interactions refers to the muons produced in the charged current interactions of neutrinos which ultimately decay inside the detector with a Bragg peak in their profile. This process is known as the overall calibration of the detector. In the last step, stopping protons from neutrino interactions are used to further refine the relation between the measured charge and the energy loss for highly ionizing particles.
The MicroBooNE coordinate system is shown in figure 1 (left). MicroBooNE uses a right-handed coordinate system in which the axis is vertical, the axis is horizontal, perpendicular to the anode and cathode planes, and the is also horizontal and along the beam direction. The TPC signal formation in MicroBooNE is illustrated in figure 1 (right), where the induction wire planes are referred to as the “U” and “V” planes and the collection wire plane is referred to as the “Y” plane [1]. In this analysis we focus on calibrating only the collection wire plane of the detector, as this is the wire plane predominantly used for calorimetry in a LArTPC. The calibration of induction planes is generally more difficult because the response of the induction wires is highly dependent on the angle of the tracks relative to the wires. This is caused by the cancellation of overlapping induction signals for large-angle tracks due to the bipolar signal shape. New techniques are being developed to improve the reconstruction of the induction plane signals [3, 4], which will allow reliable measurements on induction plane wires in the future.
The two-step calibration procedure presented in this paper is similar to calibration techniques developed for other calorimeters such as the MINOS detectors [15].
2 dQ/dx Calibration
2.1 Introduction
The goal of the calibration procedure is to make the detector response uniform in space and over time. There are many effects that can produce a nonuniform detector response. The dominant effects are described in the following sections.
2.1.1 Cross-connected TPC channels
Cross-connected wires, which can distort the electric field between wire planes [4], affect the response of roughly 20% of the detector volume. Figure 2 shows the distribution of in the collection plane as a function of Y and Z coordinates in data. The highlighted diagonal region represents a “shorted-U” region where multiple U plane channels are shorted to one or more V plane channels. The highlighted vertical region represents a “shorted-Y” region where multiple Y plane channels are shorted to one or more V plane channels. By calibrating out these effects we recover the affected areas for useful physics.
2.1.2 Space charge effects (SCE)
Since MicroBooNE is a surface-based detector, there is a significant flux of CR tracks in the detector volume. Because of this, there is a significant accumulation of slow-moving positive argon ions inside the detector, which is enough to distort the uniformity of the drift electric field. These distortions in the electric field inside the TPC have two significant effects:
- •
distortions in the magnitude of the drift electric field compared to a uniform electric field, and
- •
distortions in the electric field direction compared to the nominal direction parallel to the x-axis.
When the magnitude of the electric field is nonuniform, the recombination of electrons and ions is affected. This recombination effect [13] is sensitive to changes in the electric field. When the drift electric field is relatively low, recombination becomes dominant compared to the recombination at higher electric fields. Moreover, space charge effects can lead to spatial distortions in the trajectories of reconstructed particle tracks and electromagnetic showers. The positive argon ions built up in the detector tend to drag ionization electrons closer to the middle of the detector. The space charge effects are expected to be stronger at the detector edges transverse to the drift direction. The cumulative effect of the space charge in track reconstruction leads to squeezing of the reconstructed track in transverse directions and bending towards the cathode. See Figure 3.
Figure 4 shows how space charge effects [6] implemented in the MicroBooNE simulation change values over the entire drift distance of the MicroBooNE LArTPC. As seen in the right plot of figure 4, the values closer to the cathode are higher compared to that at the anode. The accumulation of positive ions causes the electric field magnitude closer to the cathode to be approximately 10 higher than at the anode. The higher field suppresses electron-ion recombination near the cathode. In addition, due to spatial distortions, tracks reconstructed closer to the cathode are bent and squeezed making reconstructed values smaller. Thus we observe a higher collected charge per unit track length closer to the cathode.
2.1.3 Electron attachment to impurities
When a cloud of ionization electrons drifts to the anode, electronegative contaminants such as H2O and O2 can capture some of the drifting electrons, reducing the measured at the wire planes. The capture rate is inversely dependent on the strength of the applied electric field, as at higher electric field magnitude, ionization electrons have a greater chance of making it to the anode plane before attaching to electronegative impurities.
Equation 2.1 describes the depletion of a cloud of ionization electrons due to capture by electronegative contaminants as the electrons drift towards the anode:
[TABLE]
where is the start time, is the initial number of electrons at time , is the number of electrons collected by anode plane wires after a time and is the electron lifetime.
The electron lifetime depends on the amount of electronegative contaminants present in the medium, where higher electron lifetime corresponds to lower argon contamination levels. Figure 5 shows the effect of electron attachment to impurities on values under different purity conditions. In high argon purity conditions space charge effects become dominant, causing to increase from anode to cathode. In low argon purity conditions value drops from the anode to the cathode due to electron attachment to impurities.
2.1.4 Diffusion
Both longitudinal and transverse diffusion can be studied. The cloud of ionization electrons tends to get smeared out in the direction of the drift because of longitudinal diffusion. This widens and lowers the pulse height of the signal at longer drift distance, which can lead to loss of signal if the pulse-height is below reconstruction threshold. The effect of longitudinal diffusion is small in practice because the electronics shaping time is selected to be comparable to the diffusion smearing. In addition, transverse diffusion charge can allow charge to be detected on multiple wires, which can smear the detected signal and reduce the resolution of charge reconstruction. This smearing is not corrected in the signal processing procedure used for this analysis.
2.1.5 Temporal variations
The detector response can change over time because of effects such as drift of the electronics gains, changes in temperature, different running conditions, etc. Note the time referred to here is the calendar time, not to be confused with drift time in the TPC. The most significant time-dependent change affecting MicroBooNE calibrations are changes in argon purity.
2.1.6 dQ/dx equalization strategy
The general strategy of the calibration is to separate detector nonuniformities into yz plane, x (drift) direction, and calendar time variations, and calibrate them in sequence using CR. More details will be discussed in section 2.4.
CR are the standard candle for uniform energy deposition throughout the detector. These CR have typical momenta in the range of 4 - 5 GeV, which results in a peak of 1.7 MeV/cm .
It should be noted that the calibration scheme described here is an approximation because contributions such as space charge effects cannot be completely factorized in separate yz plane and drift directions: for a given value, the variations in the yz plane are different. Ideally we should carry out this calibration by voxelizing the detector into small three-dimensional cubes and derive a calibration constant for each cube. But limited statistics make the voxelation approach impractical. Moreover, spatial distortions introduced by space charge effects that impact track reconstruction would not be completely addressed by such a voxel-based calibration scheme.
2.2 Data sample
The calibration of the detector is carried out using both data and Monte Carlo (MC) simulation. For MC simulation we use samples of simulated CORSIKA CR events [17] overlaid with neutrino interactions simulated with the genie generator [18]. The simulation of particle propagation in the MicroBooNE detector is based on Geant4 [19]. The drift of ionization electrons to the wire planes and propagation of scintillation light to the PMTs are modeled in LArSoft [20], which induces effects such as recombination, diffusion, space charge effects and electronics readout including cross-connected TPC channels. The electron lifetime is set to infinity in the simulation so there is no electron attachment to impurities. For data, we use data collected in the MicroBooNE detector with a trigger coincident with the beam and with a random trigger anti-coincident with the beam from February to October of 2016, and a second period from September 2017 to March 2018. In both MC and data we use the Pandora [21] pattern recognition program, combined with a Kalman filter implemented in the LArSoft framework to fit the CR tracks.
2.3 Event selection
For the calibration of the detector, we choose to use the anode-cathode crossing CR. The main reason is that the anode-cathode crossing CR span the entire drift distance, which make them a very valuable sample to study any effects that depend on the drift distance. Each CR in the selection is “tagged” with an initial time , corresponding to the time that the electrons produced nearest the anode are detected. Crossing CR have a wide spatial but limited angular coverage. Figure 6 shows an example of anode-cathode crossing CR in MicroBooNE data. Figure 7 shows the x projected track length distribution of all CR. It has been estimated that 0.13 of the tracks are anode-cathode crossing CR. Out of a total of 69429 tracks from a single run 5709 taken on April 1, 2016, 87 are crossing CR and 52 tracks survived after angular cuts.
The tracks used in the calibration satisfy the following selection criteria:
- •
250 cm Track projected x length 270 cm: Any track that satisfies this selection cut is considered to be an anode-cathode crossing CR. A 20 cm wide window is selected in order to account for imperfect reconstruction of track start and end positions.
- •
The absolute value of track angle (see figure 1) should not be in the range of 75*∘* to 105*∘*: This selection ensures that we are excluding tracks which are nearly orthogonal to the wire planes, which can be mis-reconstructed [3]. See figure 8.
- •
The absolute value of track angle (see figure 1) should not be in the range of 80*∘* to 100*∘*: This selection helps to remove tracks that are nearly parallel to the collection plane wires. See figure 8. The difference between MC and data in these figures is due to the fact that the MC simulation of detector effects such as space charge effects and the wire field response is not in a perfect agreement with data.
2.4 Analysis method
The calibration of the collection plane response is carried out in three separate steps.
Detector calibration in the yz plane. This step aims to remove the effects of space charge, cross-connected TPC channels, and transverse diffusion. This step is performed using all crossing tracks occurring over a period of several months. 2. 2.
Detector calibration in the drift (x) direction, which may be time-dependent. This step removes effects of electron attachment to impurities, space charge, and longitudinal diffusion. A separate correction is derived for each day of data. 3. 3.
Detector calibration in time. This step removes any temporal variations in the overall detector response, and is only performed for data. There is currently no time dependence in the MC simulation.
2.4.1 Method for detector calibration in the yz plane
We segment the yz plane into 5 cm by 5 cm cells. Each three-dimensional (3D) hit of each crossing CR track is assigned to a cell based on the and coordinates of the hit as reconstructed by Pandora, without correction for SCE-induced distortion. For each cell containing more than five hits, we calculate a median charge deposit per unit length, , where and are the central coordinates of each cell. Figure 9 shows the distribution of number of hits for each yz plane cell in the collection plane for both MC and data respectively. The statistics is low in the peripheral cells because of the spatial distortion caused by space charge. However, this is not a problem since 1) yz correction factors are still quite uniform near the TPC edges as shown in figures 11 and 12; 2) all the analyses will define a fiducial volume to exclude activities in the peripheral region. The use of the median mitigates effects of delta ray contamination, misreconstructed calorimetric information, and long tails of the distribution. We also calculate a global median charge deposit per unit length, using all 3D hits of all crossing CR tracks. The correction factor for each cell is defined as
[TABLE]
If a cell has five or fewer space points, the correction factor for that cell is calculated from the values of nearby cells.
2.4.2 Method for detector calibration in the drift direction
We segment the drift direction into bins of 10 cm in data and bins of 5 cm in MC. These bin sizes are chosen to minimize statistical fluctuations in the correction factors. Each 3D hit of each crossing CR track on a given day is assigned to a bin based on the reconstructed coordinate of the hit, with position corrected only for the of the CR as described in section 2.3, and charge deposit corrected only by the factor. For each bin containing more than five hits, we calculate a median charge deposit per unit length, , where is the central position of the bin, and the prime on denotes that the yz correction has been applied. The correction factor for each bin in the drift direction is defined as
[TABLE]
where is the global median value after applying the yz correction to each 3D hit.
In data, we derive separate drift direction correction factors for each day as space charge effects and electron lifetime can change over time. To get a reliable set of correction factors, we need a data sample which contains enough statistics. For this purpose, we derive correction factors in the drift direction only for the days where there are more than 40 crossing tracks after all the angular cuts. In the case of lack of statistics for a given day, we use the drift direction correction factors derived for the immediate neighboring day. For MC we derive only a single set of drift direction correction factors for the collection plane as there is no time variation of space charge effects and the electron lifetime is set to infinity.
2.4.3 Time dependent calibration of the detector
After applying the yz and drift direction factors to each 3D hit, the daily global median value of of all hits on crossing CR tracks varies over time, as shown in figure 10. Dips lasting several days reflect times when the argon purity was low, e.g., during recirculation pump maintenance. We choose a reference value for the global median in the collection plane, since all global median values calculated for each day fluctuate around this value after yz- and drift-dependent correction. We then define the overall time dependent correction as
[TABLE]
where is the reference value in the collection plane, and is the global median value in the collection plane after correcting for yz plane and drift direction irregularities for each day.
Values of fully calibrated to the reference value are obtained by multiplying the reconstructed values by the factor . For MC simulated data without temporal variation we define .
2.5 Results
2.5.1 dQ/dx calibration (Simulation)
The calibration of the detector in MC is performed using a single large dataset described in section 2.2. Figure 11 shows the extracted yz correction factors and drift direction x correction factors . The variation of the correction factors with respect to the drift distance could be attributed to electron attachment to impurities as well as space charge effects as shown in Figure 5. The bump in the last bin of the drift direction correction distribution is due to low statistics in that bin.
2.5.2 dQ/dx calibration (data)
The yz and drift direction x correction factors are derived for two separate datasets as described in section 2.2. In the first dataset, we have data combined from February 2016 to October 2016; and in the second dataset, we have data combined from September 2017 to March 2018. See figure 12 for the variation of the yz correction factors for data taken from February to October in 2016 and drift direction x correction factors for data taken on February 25, 2016 in the collection plane. The distributions for data taken from September 2017 to March 2018 are similar.
The drift direction corrections factors and time corrections are derived for the time period from February 2016 to October 2016 and the time period from September 2017 to March 2018. is calculated for each day. Due to intermittent periods of detector downtime, some days are skipped in this measurement. See Figure 13 for the variation of the time-dependent correction factor over time. Figure 14 shows the impact of the calibration on the distribution in both MC and data. In both data and MC, we see that the width of the central peak of the distribution has narrowed by 2%, and the significant excess of low values in the data before calibration has been corrected. This represents a significant improvement in the measurement of , particularly in regions requiring the largest yz corrections, which is important for the measurement of total deposited energy and particle identification based on energy loss per unit length as a function of residual range.
2.6 Angular dependence study
The calibration procedure described in this paper does not make any angular-dependent corrections because of the limited angular coverage of anode-cathode crossing CR. The angular distributions of the anode-cathode crossing CR tracks are shown in figure 8. In order to more closely study the angular dependence of the calibration, we divide the crossing CR into four angular subsamples:
- •
,
- •
,
- •
and
- •
.
and derive the yz and x correction factors for each subsample following the same procedure described above. All of the other selection criteria remain the same.
For each yz cell in each subsample, we compute the relative difference of the yz correction factor for the subsample from the correction factor found using all crossing CR . Figure 15 shows the distribution of the relative differences of the yz correction factors relative to the complete sample, , for each angular subsample. The shift in the mean of the distribution away from zero is indicative of the amount of bias observed in a given angular bin. The peaks of the four distributions all agree within 2%.
Figure 16 shows the relative differences in the x correction factors between the different angular subsamples and the combined sample. The maximum variation is .
Based on the above studies, we assign a systematic error on the yz correction and a systematic error on the x correction for the collection plane. This study shows very little angular dependence for the sample selected for the calibration. However, because of the limited angular coverage of this sample, more studies are underway to further refine the angular dependence of the detector response.
3 dE/dx Calibration
3.1 Introduction
After the correction is complete we can determine the absolute energy loss per unit length scale. To move from the calibrated to , we use the modified box model [13] for recombination:
[TABLE]
with
is a calibration constant used to convert ADC values to number of electrons,
[TABLE]
The last two parameters were measured by the ArgoNeuT experiment [13] at an electric field of 0.481 kV/cm. The modified box model is applied at MicroBooNE’s electric field of 0.273 kV/cm.
According to the above equation, precise determination of the calibration constant which translates \sayADC/cm to \say(number of electrons)/cm, is important in determining the absolute energy scale. The charge is measured by the integral of the pulse of the deconvoluted anode response signal where the simulated electronics and field responses are removed. is normalized so that the unit (\sayADC) corresponds to 200 electrons. In the case where the detector response is perfectly modeled (e.g. in the simulation), the calibration constant should be exactly 1/200. Here the goal of the calibration is to determine the calibration constant using stopping muons as the standard candle, because they have a well-understood energy loss profile. A sample of stopping muons can either be identified from cosmic data or neutrino interactions. The method we use for calibration is discussed in section 3.4.
3.2 Data sample
The calibration of the detector is performed on both MC and data samples. For MC we use a CORSIKA CR simulation overlaid with neutrino interactions to obtain stopping muons induced by the neutrino interactions. In data we use datasets collected from February 2016 to October 2016 and from September 2017 to March 2018, with the final states of the inclusive charged current interactions [22], to obtain candidate stopping muons from charged current interactions. In both MC and data we use the Pandora [21] pattern recognition program to reconstruct the neutrino and cosmic interactions.
3.3 Event selection
In the calibration of the detector, we employ stopping muons induced by neutrino interactions. The motivation is to get a sample of muons from neutrino interactions with known muon energy loss per unit length so that a portion of the track can be selected that corresponds closely to a minimum ionizing particle (MIP), for which the is very well understood theoretically to better than 1%. This sample contains the particles with the energy regime and the angular profile well representing the signatures of neutrino interactions. Further, most events have forward-going muons, providing us with plenty hits in the collection plane. The well known starting time of the stopping muons, determined by the arrival of the neutrinos produced in BNB, allows us to apply the dQ/dx calibration corrections.
We select the candidate muon tracks with the reconstructed start and end points 10 cm away from the TPC border in each dimension. The tracks with the reconstructed angles, and , are discarded to exclude tracks pointing into the wire planes and tracks nearly parallel to the collection plane wires, respectively. To further ensure the quality of the reconstruction and the feature of minimum ionization, the tracks are required to have a minimum length of 150 cm. In the MC sample, the candidate muon track has to match the true stopping muon coming from a neutrino interaction, and the distance between the start (end) point of the reconstructed track and that of the true muon has to be shorter than 5 cm. In the data sample, the ratio of the median value in the last 5 cm segment to that of the first 5 cm segment of the candidate track is required to be greater than 1.5, ensuring the selected tracks to have the profile of a stopping muon. The selection criterion based on the profile is determined from a MC simulation sample, as shown in figure 17. The non-muon contamination in the selected data sample is estimated to be less than 1% using MC.
3.4 Analysis method
The following explains the procedure to perform the calibration both in MC and data:
- (i)
Start with a set of tracks that satisfy the selection in section 3.3. 2. (ii)
Segment the last 200 cm residual ranges (residual range is defined as the path of the particle before its stopping point) of tracks into 5 cm bins, which leads to a total of 40 residual range bins. If the track length is less than 200 cm, the entire track is used, otherwise, the last 200 cm segment with respect to the stopping point is used. 3. (iii)
Loop over all the 3D hits of each selected track and fill the residual range bins with values derived using equation 3.1 by setting the calibration constant to an arbitrary value. Moreover, we consider only the 3D hits which are separated by 0.3 to 0.4 cm; as the minimum spacing between 3D hits is set by the wire spacing of 0.3 cm, this selects forward going muons. 4. (iv)
After looping over all the tracks, fit each distribution to a Landau-convoluted Gaussian function [24] to extract the most probable value (MPV) representing that particular residual range bin. 5. (v)
Plot the MPV values against the kinetic energy of the particle; for each residual range bin, we take the middle bin value as the representative residual range value of that bin and transform that to kinetic energy. In this transformation we use a cubic spline fit to the tabulated values of CSDA (continuous slowing down approximation) residual range vs. kinetic energy for stopping muons in liquid argon [25]. 6. (vi)
Compare the curve generated in the previous step with the prediction made by the Landau-Vavilov function [26] in the kinetic energy range of the muons from 250 MeV to 450 MeV, which is in the region expected for MIPs, and a 2 value is calculated using equation 3.2. The Landau-Vavilov function describes the energy loss probability distribution for a particle in a given medium. The most probable energy loss of a particle is dependent on the thickness of the energy absorber. To get the predicted MPV , we set the absorber thickness to be 0.35 cm, the average value of .
[TABLE]
where we sum all the data points in the kinetic energy region of 250 MeV to 450 MeV.
For MC, , where is the uncertainty associated with the MPV extracted by fitting a Landau-convoluted Gaussian function to the energy distribution.
For data, , where is the systematic error associated with recombination model uncertainties (here we take the uncertainty to be 1.5 of the measured . This uncertainty is coming from the 1 level uncertainties of the two free parameters and in the modified box model for recombination in equation 3.1).
Note the MC sample was simulated and reconstructed with the same recombination model. Therefore, there is no systematic uncertainty associated with the recombination model for MC. 7. (vii)
Iterate through the steps iii to vi described above several times to generate tabulated set of data between a given calibration constant () and 2 value (here we change the parameter in each iteration and calculate the 2 value. The total number of iterations in this procedure is the total number of data points in the plots shown in figure 18). 8. (viii)
In the final step, plot the 2 values generated against calibration constants and fit that distribution with a second order polynomial to get the calibration constant which corresponds to the lowest 2 value. 9. (ix)
With the newly derived calibration constant (), calculate the values using equation 3.1 with calibrated values as the input.
After the calibration, we compare the newly derived values with uncorrected values to see the effects of applying the algorithm. (See section 3.5).
3.5 Results
The absolute calibration of the detector is performed on both MC and data samples. Figure 18 shows vs. the calibration constant for the collection plane. The best fit value is the one that gives the minimal . The uncertainty on the extracted value of is given by .
The calibration constants derived for the collection plane are shown in table 1. The MC calibration constant uncertainty is statistical only. The data calibration constant uncertainty includes both the statistical uncertainty and the recombination uncertainty. The difference between data and MC calibration constants is due to the fact that the MC simulation of detector response is not in a perfect agreement with data.
Figure 19 shows the comparison between the predicted most probable energy loss [26] with the fitted most probable energy loss using the calibration constants shown in table 1 for stopping muons both in MC and data. Here the disagreement between the fitted distribution and theory at lower kinetic energies in data is mainly due to the recombination model uncertainty. Results can be applied to all time ranges, since temporal variations are taken into account by the relative corrections. The first fitted point in the distributions corresponds to the end of the tracks. Most of the time the end of the track can happen in between two wires. Therefore calorimetric information for the end points of tracks is not properly reconstructed. The significant deviation between prediction and our fitted distribution for MC in the first bin is due to this reason.
The absolute energy-scale calibration can be validated in a data-driven way by comparing the range-based energy to that obtained via calorimetry for selected stopping muon candidates in data and MC. For each selected stopping muon track, the kinetic energy calculated by range and by calorimetry is computed, and the relative difference between these methods is shown in figure 20. Here, in calculating the kinetic energies, the entire track is used. The agreement between the two, approximately 2% for data (1% in MC), gives confidence in the proper absolute energy-scale calibration to within this level of agreement, and provides a closure test for this calibration procedure. The tails on the positive side are caused by contributions of other particles to the muon track near the neutrino interaction vertex. A MC study shows that after the calibration procedure, the bias in reduces from -3.2% to -1.2% and the resolution improves from 8.1% to 7.7%.
4 Energy correction with protons
4.1 Introduction
After the nonuniformities in detector response have been corrected for and the absolute energy scale has been determined as described in the previous sections, we perform studies using a proton sample to investigate the effect of electron-ion recombination.
The calibration of and presented in the previous sections sets the correct scale for energy loss per unit length for MIPs, for which remains nearly a constant. As particles approach the end of their range, they become increasingly ionizing. Electron-ion recombination introduces a non-linear relationship between and , which has been extensively studied [27, 28]. In this study, we use stopping protons to measure the non-linear relationship between and . The variation of with the proton energy allows us to measure the recombination effect for a wide range of stopping powers up to about 20 MeV/cm.
The detector simulation used by MicroBooNE employs the modified box model by default to simulate the recombination effect, with the parameters obtained by the ArgoNeuT experiment [13]. The charge deposition at each Geant4 step is calculated using the energy loss following:
[TABLE]
with the parameters , , , \beta\textsuperscript{\prime} and , already described in the inverse of this equation, eq. 3.1.
In this analysis, we compare the proton vs. measurements with two recombination models: the modified box model as described in eq. 4.1 and Birks’ law, following eq. 4.2 with and , as measured by the ICARUS collaboration [14]:
[TABLE]
The two models are parameterized by ArgoNeuT and ICARUS at different electric fields from the operating electric field at MicroBooNE. Therefore, it is interesting to compare both models with the MicroBooNE measurement to study the dependence of recombination on electric field.
To perform these studies, we use the calibrated information and calculate using an empirical fit to Bethe-Bloch energy loss using the range information, . The values for protons are calculated using the residual range information per hit, as described in Ref. [13] according to eq. 4.3
[TABLE]
where is the residual range and and are empirical constants (the constant in eq. 4.3 should not be confused with from Birks’ law in eq. 4.2). We use the values obtained by ArgoNeuT, and in the calculation. The values calculated in the way are independent of the charge measurement, and therefore can be compared to the calibrated values to derive recombination model parameters.
4.2 Event selection
In this study we use protons emerging from neutrino interactions with argon. We choose a sample of events with one muon and two protons. This sample is referred to as . Such events are produced almost exclusively by the neutrino interactions, rather than the cosmic ray interactions, as verified using data taken when the BNB beam was off. Having three tracks sharing the same vertex improves the resolution of vertex and track reconstruction. This sample provides a highly pure proton sample with well reconstructed track trajectories, which is ideal for the recombination measurement.
The event selection is based on a set of cuts designed to select charged-current interactions producing final state in the TPC fiducial volume (FV). We take advantage of the very specific number of charged particles and the absence of in the final state, which yields a sample of well reconstructed tracks.
For the events, the signature is three tracks originating from the same vertex, one consistent with a muon and the other two consistent with protons. The selection criteria are described as follows:
- •
Beam window: 32 PMTs are used in MicroBooNE to record scintillation light information from particles traveling inside the TPC. In order to select neutrino interactions, we require there should be a scintillation light signal detected in the PMTs coincident with the neutrino beam spill.
- •
Vertex: The reconstructed neutrino vertex is required to be in the FV. The FV is defined as a rectangular box shaped as follows: the boundary from the anode plane and the cathode plane is 10 cm, the boundary from the upstream and downstream ends is 10 cm, and the boundary from the top and bottom of the TPC is 20 cm. It is the same FV as used by the MicroBooNE physics analysis reported in [23]. The more stringent requirement in the vertical direction ensures a higher CR background removal.
- •
Three tracks cut: Only three tracks are connected to the vertex within a 5 cm radius. If more than one neutrino candidate is present in an event, the one where the sum of the distance between the start position of each track to each other is smallest is selected. This suppresses cosmic ray background mimicking neutrino interactions.
- •
Proton containment: The leading particle, defined as the longest track, is taken as the muon candidate; the remaining two tracks are taken as proton candidates. These two proton candidates are required to be contained in the TPC FV to aid in particle identification (PID).
- •
Minimum number of hits: For each track we require a minimum number of 10 hits in total, summing all three planes. In order to have the most reliable calorimetry information, we also require at least 5 hits in the collection plane. This requirement is applied to all tracks in a event.
- •
PID cut: We use the test with respect to a proton hypothesis in the vs. residual range curves, from Bethe-Bloch predictions, to discriminate MIP with respect to non-MIP particles. The predicted proton vs. residual range curve is calculated using the Geant4 simulation, where the residual range is sampled with a step size of 0.08 cm.
For each selected track (either in MC or data), we reconstruct per hit in the collection plane as described in section 3.1 and calculate the between the reconstructed and the Geant4-determined mean at each given residual range bin. These values are then summed for all hits on the selected track.
We exclude the first and last hits on the track from this calculation because of the uncertainty in the calculation. This is due to the fact that the calculation for the first and last hits may be wrong since the exact start and end positions of the track between the wires are unknown. It also mitigates the complications caused by energy deposition overlap at the interaction vertex from other particles (e.g. short protons) produced by the neutrino interaction.
The value obtained is then normalized by the number of degrees of freedom (d.o.f.), which corresponds to the number of hits in the collection plane. This calculation can be determined for several particle assumptions, but in this analysis we exclusively use the proton hypothesis.
[TABLE]
where is the estimated resolution in measuring a fixed value of , the subscript measured and theory corresponds to the measured at the wires using the charge deposition information and the predicted by Geant4 respectively. Values used for resolution in are taken from studies performed by the ArgoNeuT Collaboration. This does not take into account the mm wire spacing in MicroBooNE (compared to the mm wire spacing in ArgoNeuT), which leads to an overestimated resolution in when compared to the resolution we obtain from the MicroBooNE MC studies. We prefer to use the values from ArgoNeuT as a conservative approach since the recombination model is not tuned based on MicroBooNE data at this stage yet. In order to select a sample of highly ionizing particles, an initial guess of recombination correction is needed, for which we employ the modified box model with the parameters measured by ArgoNeuT. This model is sufficiently close to the final measurement to ensure a good separation of protons and muons.
Taking into account both the particle discrimination (muon vs. proton) and the data/MC shape agreement, we apply a PID requirement that identifies protons as the tracks with , and muons as tracks with . The values are optimized by maximizing the efficiency and purity for the proton identification.
Table 2 shows the purity and efficiency of the muon and proton selections, estimated from the MC sample. Purity is defined as the quantity of particles from an specific species which has been correctly identified, divided by the total number of particles of that species in the sample. For comparison, the same values are shown before applying the equalization procedure described in section 2. Note that in the MC simulation, we reconstruct using the recombination parameters obtained by ArgoNeuT, as they are the values used in the simulation. The equalization calibration improves both the purity and efficiency of the muon and proton selections.
- •
Shower veto: We remove events with a shower reconstructed within 15 cm of the vertex, both to reduce contamination and to reduce background caused by tracks misreconstructed as showers.
In this analysis, a total of 226 proton candidates are selected from the neutrino data.
4.3 Analysis method
With the proton sample described we study two different recombination models: the modified box model, eq. 4.1, and Birks’ law, eq. 4.2. The comparison of these two models with the MicroBooNE proton dataset (described in section 4.2) is shown in figure 21. In this figure, the measured , after calibration, is shown with respect to the theoretically predicted energy deposit , computed from the proton residual range following eq. 4.3, for both data (left) and MC (right). Both model predictions are calculated at the nominal electric field of 0.273 kV/cm for MicroBooNE.
We observe a good agreement of data with Birks’ law and the modified box model at low in figure LABEL:sub@fig:protons2D_defaultBirksBox_data, as expected. Low corresponds to the MIP region, which has been calibrated with muons previously. At higher , data shows a lower recombination effect than the one predicted by the modified box model. On the other hand, Birks’ law agrees much better with data at higher ionization. This is not surprising since Birks’ law was tuned with ICARUS data taken at drift fields of 0.2, 0.35 and 0.5 kV/cm while the modified box model was tuned with ArgoNeuT data taken at a single drift field of 0.481 kV/cm. As expected, the MC distribution agrees well with the modified box model using the ArgoNeuT parameters, shown as the black curve in figure LABEL:sub@fig:protons2D_defaultBirksBox_mc, since it is the same model used for the simulation.
Because the parameters in the modified box model and in Birks’ law are coupled to the electric field in eqs 4.1 and 4.2, it is the quantity or that we are able to measure. The accumulation of positive ions produced by CR in the liquid argon leads to a nonuniform electric field. Even though the nonuniformities are largely reduced by the equalization procedure described in section 2, there could still be residual nonuniformities in the electric field distribution. The discrepancies shown in figure LABEL:sub@fig:protons2D_defaultBirksBox_data could be due to the fact that the recombination model parameters derived at other experiments are not applicable at the MicroBooNE electric field, or that the average electric field at MicroBooNE is different from the nominal electric field of 0.273 kV/cm. For the recombination measurements reported in this paper, we adopt an effective approach by fixing the electric field at 0.273 kV/cm and tune the other model parameters to improve the agreement with the MicroBooNE proton data.
We obtain the effective recombination parameters through a fit. We construct a function based on the mean value per each bin in data and MC. We then minimize the function to obtain the best fit parameters ( and for the modified box model, and and for Birks’ law). For data, in is calculated by dividing the calibrated in , as described in section 2, by the calibration constant , as described in section 3.5; i.e.
[TABLE]
and is calculated using the residual range following eq. 4.3. For MC, is also calculated using residual range and is converted from using the recombination model (the modified box model eq. 4.1 or Birks’ law eq. 4.2). The is defined as:
[TABLE]
where and are the statistical errors from data and MC respectively.
4.4 Results
We perform the fit using both the modified box model and Birks’ law. The best fit results are summarized in table 3, including the statistical uncertainties from the fit. To obtain these results, we assume an electric field of , which is the nominal electric field in MicroBooNE. Under this assumption, the parameters that are not coupled to the electric field, and , are consistent with the measurements from ArgoNeuT and ICARUS, while the parameters that are coupled to the electric field, and , are in a disagreement with the measurements from ArgoNeuT and ICARUS. The effect of the obtained result can be clearly observed in figure 22, which is the same as figure LABEL:sub@fig:protons2D_defaultBirksBox_data now showing also the model curves using effective recombination parameters. We observe a better agreement of the data to the two models with the new parameters.
The effective recombination parameters are validated using the MicroBooNE CC inclusive data sample. The calculated values using the effective recombination parameters are compared with the Bethe-Bloch prediction with respect to residual range in figure LABEL:sub@fig:allprotons_dEdxvsrr_OnBeam_ub. As a comparison, we show in figure LABEL:sub@fig:allprotons_dEdxvsrr_OnBeam_argo the reconstructed using the modified box model with the ArgoNeuT parametrization. Improvement between the calculated distribution and the Bethe-Bloch prediction for protons can be seen when using the effective recombination parameters (right), with respect to the same model using the parameters obtained by ArgoNeuT (left).
The impact of the full calibration procedure on proton reconstruction is shown in figure 24. For figures LABEL:sub@fig:comparison_protondEdx_a, LABEL:sub@fig:comparison_protondEdx_b and LABEL:sub@fig:comparison_protondEdx_c, the distributions show the per hit of the proton candidates in data compared to those from the simulation. Simulated is calculated using calibrated and the modified box model with ArgoNeuT parameters. Data in figures LABEL:sub@fig:comparison_protondEdx_a is calculated using uncalibrated and the modified box model with ArgoNeuT parameters. Data in figures LABEL:sub@fig:comparison_protondEdx_b is calculated using calibrated and the modified box model with ArgoNeuT parameters. The equalization calibration clearly improves the reconstruction, but there is a discrepancy at high values. Data in figures LABEL:sub@fig:comparison_protondEdx_c plot is calculated using calibrated and the modified box model with the MicroBooNE effective recombination parameters as reported in table 3. The agreement is improved for high values. The agreement for low values gets slightly worse. This will be improved by retuning the scale using stopping muons, as described in section 3, after applying the effective recombination parameters in the future.
We also show the per hit distributions for selected protons and muons from data in figures LABEL:sub@fig:comparison_protondEdx_d. In this plot, the equalization calibration and the effective recombination correction are applied. We can clearly see from the figure that protons and muons can be well identified and have different widths in as expected.
5 Conclusions
In this paper, we describe for the first time a method that calibrates the calorimetric response of the ionization signal in a LArTPC using muons. We use crossing muons to remove spatial and temporal variations in the detector response. We use stopping muons to determine the absolute energy loss per unit length scale for minimum ionizing particles. We use stopping protons to further refine the relation between the measured charge and energy loss for highly-ionizing particles. The most probable value of calibrated energy loss agrees with expectation. The calibrated energy loss per unit length is used in particle identification and energy reconstruction. This method can be followed by other LArTPC experiments to calibrate their detectors.
Acknowledgements
This document was prepared by the MicroBooNE collaboration using the resources of the Fermi National Accelerator Laboratory (Fermilab), a U.S. Department of Energy, Office of Science, HEP User Facility. Fermilab is managed by Fermi Research Alliance, LLC (FRA), acting under Contract No. DE-AC02-07CH11359. MicroBooNE is supported by the following: the U.S. Department of Energy, Office of Science, Offices of High Energy Physics and Nuclear Physics; the U.S. National Science Foundation; the Swiss National Science Foundation; the Science and Technology Facilities Council (STFC), part of the United Kingdom Research and Innovation; and The Royal Society (United Kingdom). Additional support for the laser calibration system and cosmic ray tagger was provided by the Albert Einstein Center for Fundamental Physics, Bern, Switzerland.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1[1] R. Acciarri et al. (Micro Boo NE Collaboration), “Design and construction of the Micro Boo NE detector”, JINST 12 (2017), P 02017 [ar Xiv:1612.05824].
- 2[2] J. Conrad et al. [Micro Boo NE Collaboration], “The Photomultiplier Tube Calibration System of the Micro Boo NE Experiment,” JINST 10 , no. 06, T 06001 (2015) doi:10.1088/1748-0221/10/06/T 06001 [ar Xiv:1502.04159 [physics.ins-det]].
- 3[3] R. Acciarri et al. (Micro Boo NE Collaboration), “Ionization Electron Signal Processing in Single Phase L Ar TP Cs I: Algorithm Description and Quantitative Evaluation with Micro Boo NE Simulation”, JINST 13 (2018), P 07006 [ar Xiv:1802.08709].
- 4[4] R. Acciarri et al. (Micro Boo NE Collaboration), “Ionization Electron Signal Processing in Single Phase L Ar TP Cs II: Data/Simulation Comparison and Performance in Micro Boo NE”, JINST 13 (2018), P 07007 [ar Xiv:1804.02583].
- 5[5] R. Acciarri et al. (Micro Boo NE Collaboration), “Noise Characterization and Filtering in the Micro Boo NE Liquid Argon TPC”, JINST 12 (2017) P 08003 [ar Xiv:1705.07341].
- 6[6] M. Mooney, “The Micro Boo NE Experiment and the impact of Space Charge Effects”, Proceeding of DPF 2015 Meeting of APS, [ar Xiv:1511.01563].
- 7[7] E. Buckley et al., “A study of ionization electrons drifting over large distances in liquid argon”, Nucl. Instrum. and Methods in Physics Research A 275, 364-372 (1989).
- 8[8] A. Bettini et al., “A study of the factors affecting the electron lifetime in ultra-pure liquid argon”, Nucl. Instrum. and Methods in Physics Research A 305, 177-186 (1991).