Increasing signal amplitude in electrical impedance tomography of neural activity using a parallel resistor inductor capacitor (RLC) circuit
James Hope, Zaid Aqrawe, Marshall Lim, Frederique Vanholsbeeck, Andrew, McDaid

TL;DR
This study introduces a parallel RLC circuit to enhance impedance signal amplitude in neural EIT, demonstrating significant increases in signal strength in vitro, which could improve spatial resolution and real-time imaging capabilities.
Contribution
The paper presents a novel application of a parallel RLC circuit in neural impedance tomography to significantly boost signal amplitude, aiding real-time imaging.
Findings
Impedance signal increased by up to 44% with RLC circuit.
Optimal frequency for RLC implementation was around 17 kHz.
Enhanced signal amplitude can improve spatial resolution in EIT.
Abstract
Objective: To increase the impedance signal amplitude produced during neural activity using a novel approach of implementing a parallel resistor inductor capacitor (RLC) circuit across the current source used in electrical impedance tomography (EIT) of peripheral nerve. Approach: Experiments were performed in vitro on sciatic nerve of Sprague-Dawley rats. Design of the RLC circuit was performed in electrical circuit modelling software, aided by in vitro impedance measurements on nerve and nerve cuff in the range 5 Hz to 50 kHz. Main results: The frequency range 17 +/- 1 kHz was selected for the RLC experiment. The RLC experiment was performed on four subjects using an RLC circuit designed to produce a resonant frequency of 17 kHz with a bandwidth of 3.6 kHz, and containing a 22 mH inductive element and a 3.45 nF capacitive element. With the RLC circuit connected, relative increases in…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.