Production and optical properties of liquid scintillator for the JSNS$^{2}$ experiment
J. S. Park, S. Y. Kim, C. Rott, D. H. Lee, D. Jung, F. Suekane, H., Furuta, H. I. Jang, H. K. Jeon, I. Yu, J. H. Choi, J. S. Jang, K. K. Joo, K., W. Ju, M. Pac, P. J. Gwak, S. B. Kim, S. Hasegawa, S. H. Jeon, T. Maruyama,, R. Ujiie, Y. Hino, Y. S. Park

TL;DR
This paper details the production, optical characterization, and stability assessment of gadolinium-loaded liquid scintillator for the JSNS$^{2}$ neutrino experiment, demonstrating its suitability for long-term use in neutrino detection.
Contribution
It presents the production process and optical property measurements of the LS, confirming its stability when stored in ISO tanks for the JSNS$^{2}$ experiment.
Findings
LS produced at a refurbished facility meets optical requirements
Storing LS in ISO tanks does not degrade optical properties
Long-term stability of LS confirmed for experimental use
Abstract
The JSNS (J-PARC Sterile Neutrino Search at J-PARC Spallation Neutron Source) experiment will search for neutrino oscillations over a 24 m short baseline at J-PARC. The JSNS inner detector will be filled with 17 tons of gadolinium-loaded liquid scintillator (LS) with an additional 31 tons of unloaded LS in the intermediate -catcher and outer veto volumes. JSNS has chosen Linear Alkyl Benzene (LAB) as an organic solvent because of its chemical properties. The unloaded LS was produced at a refurbished facility, originally used for scintillator production by the RENO experiment. JSNS plans to use ISO tanks for the storage and transportation of the LS. In this paper, we describe the LS production, and present measurements of its optical properties and long term stability. Our measurements show that storing the LS in ISO tanks does not result in degradation of…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
11footnotetext: e-mail : [email protected]
Production and optical properties of liquid scintillator for the JSNS2 experiment
J. S. Park
S. Y. Kim
C. Rott
D. H. Lee
D. Jung
F. Suekane
H. Furuta
H. I. Jang
H. K. Jeon
I. Yu
J. H. Choi
J. S. Jang
K. K. Joo
K. W. Ju
M. Pac
P. J. Gwak
S. B. Kim
S. Hasegawa
S. H. Jeon
T. Maruyama
R. Ujiie
Y. Hino
Y. S. Park
Abstract
The JSNS2 (J-PARC Sterile Neutrino Search at J-PARC Spallation Neutron Source) experiment will search for neutrino oscillations over a 24 m short baseline at J-PARC. The JSNS2 inner detector will be filled with 17 tons of gadolinium-loaded liquid scintillator (LS) with an additional 31 tons of unloaded LS in the intermediate -catcher and outer veto volumes. JSNS2 has chosen Linear Alkyl Benzene (LAB) as an organic solvent because of its chemical properties. The unloaded LS was produced at a refurbished facility, originally used for scintillator production by the RENO experiment. JSNS2 plans to use ISO tanks for the storage and transportation of the LS. In this paper, we describe the LS production, and present measurements of its optical properties and long term stability. Our measurements show that storing the LS in ISO tanks does not result in degradation of its optical properties.
1 Introduction
The JSNS2 (J-PARC Sterile Neutrino Search at J-PARC Spallation Neutron Source) experiment aims to search for sterile neutrinos at a 24 m baseline with m2 near 1 eV2 [1]. JSNS2 will use an intense source of muon antineutrinos from muon decay-at-rest; the muons are produced by the interactions of 3 GeV protons with a mercury target at the J-PARC Material and Life Science Experimental Facility. JSNS2 will search for oscillations, which can be detected via inverse beta decay, . The detector consists of an inner target volume, an intermediate -catcher region, and an outer veto [2, 3]. The target region will be filled with 17 tons of gadolinium loaded liquid scintillator (LS), which will be donated by the Daya Bay experiment [4]. The optically separated -catcher and veto will be filled with 31 tons of LS. Linear alkyl benzene (LAB) was chosen as the organic solvent due to safety considerations and because of its high flash point, large attenuation length, and sufficient light yield. We produced the LS using a mass production facility from the RENO experiment. JSNS2 plans to use ISO (International Organization for Standardization) tanks to store LS and transport it from the RENO site to Japan. The ISO tanks are made according to the ISO international standard, and the surface in contact with the LS is stainless steel (SUS316). Therefore, we do not expect any material compatibility issue.
2 Mass production of liquid scintillator
2.1 Refurbishment of the RENO mass production facility
The RENO LS mass production facility consists of two of 10,000 L stainless steel LAB storage tanks, two acrylic production vessels (2000 L and 250 L), pipelines, stirring motors, impellers, magnetic pumps, and a nitrogen purging system. Two 24 kL ISO tanks were used to transfer the LS to Japan. The LS production facility was unused for 7 years and the pipelines were optimized for RENO, so the facility required cleaning and design modifications before it could be employed to make the JSNS2 LS. A 3-step cleaning procedure was used for each tank and vessel. The initial cleaning was done with normal soap to remove organic contaminants and dust. Next, we used Alconox, which is a special powder detergent to remove deposits, chemicals, and radioactive contamination [5]. Finally, everything was cleaned several times with high-pressure water to remove any remaining soap or Alconox. To enable safe operation of the facility, we purchased new stirring motors, impellers, and magnetic pumps. Due to size limitations, the ISO tanks were placed outside of the RENO entrance tunnel, and as a result, the pipelines were extended 300 m to reach the tanks. Additionally, we installed a strong diaphragm pump with an air compressor to transfer the LS from the production facility to the ISO tank. Figure 1 shows conceptual drawings of the refurbished facility.
2.2 Mass production of LS
The LS components and mass production method are similar to those used in RENO [6]. We chose 3 g/L of PPO (2,5-diphenyloxazole) as the primary fluor and 30 mg/L of bis-MSB (1,4-bis(2-methylstyryl) benzene) as the secondary wavelength shifter. We produced 1800 L of LS in each batch. For simplicity, the LS is produced beginning with a master batch with 10 times the concentration of bis-MSB and PPO. To produce the master, we added 100 L of LAB to the 250 L acrylic production vessel, turned on a stirring motor, and added 54 g of bis-MSB. PPO is not added at this stage because the solubility of bis-MSB in LAB is lower than PPO. It took approximately 1 hour for the bis-MSB to dissolve into the LAB. Next, we added 80 L of LAB and 5.4 kg of PPO and stirred 4 additional hours. After stirring the mixture, 30 psig of nitrogen gas were injected to purge master solution. Finally, the master solution was transferred to the 2000 L acrylic vessel, and an additional 1800 L of LAB was added using a magnetic pump. The combined mixture was stirred for an additional 5 minutes in the 2000 L acrylic vessel to ensure uniform mixing of LS master and LAB. We took a 100 mL sample of each batch to measure the optical properties before pumping the LS into the ISO tank. A total of 38000 L of LS was produced in 21 batches including a small amount of extra scintillator (a few percent of the total).
3 Optical properties of LS
3.1 Water content
Measuring the LS water content is an indirect method to assess the purging efficiency as shown by RENO [6]. Purging the LS with nitrogen serves two main purposes: removing oxygen which may quench the scintillation and removing tiny bubbles which form in the scintillator. These tiny bubbles appear after several hours of powerful stirring during LS production and make the LS appear cloudy. It may be the case that these bubbles originate from small amounts of water vapor or oil and would dissipate over time. However, we decided to remove them forcibly by purging the LS. We measured the water content of each sample at Chonnam National University using a Mettler-toledo CH/C20 Coulometric Karl-Fischer Titrator, which can precisely measure the water content of the LS in the range 1 - 50000 ppm. The measured water content of each sample was 50 - 60 ppm, which is a similar to the RENO experiment meaning the amount of nitrogen purging was sufficient.
3.2 Relative Transmittance
The transmittance of each LS sample was also measured at Chonnam National University. The measurements were done a few days after production with a 10 cm cuvette and UV-1800 spectrometer by Shimadzu JP. The quality of the LAB used for the LS production was confirmed by the RENO experiment and we can expect an attenuation length longer than a few meters, which is good enough for the JSNS2 experiment. However, because the LS was stored inside of ISO tanks during transit, additional measurements in Japan were needed to determine if any degradation occurred. The LS was sampled twice in Japan after it was transported from Korea. The first sample was taken immediately after the scintillator arrived in Japan, and the second sample was taken two months later. Both samples were measured at Tohoku University with a 10 cm cuvette. Figure 2 shows the results of the transmittance measurements made in Korea and Japan. The black solid line is the measurement made in Korea, and the red and blue solid lines are the measurements made in Japan. All three measurements show identical shapes meaning that there was no significant degradation during transportation from Korea to Japan, or during two months of storage inside of an ISO tank. Note that data is normalized to 100 at 440 nm to eliminate the effect of baseline fluctuations of the UV-VIS spectrometers.
3.3 Relative Light yield
With a small vial (100 mL) of LS and a radioactive source, it is difficult to extract an absolute measurement of the light yield because events are not usually completely contained (i.e. no clear energy peak is visible) and there is uncertainty in the optical photon collection efficiency of the photomultiplier tube (PMT). However, the LS production method described above is well-established and our main concern was the possibility of relative degradation due to storage in the ISO tanks. To reduce systematic uncertainty, we used a coincidence setup with LS and a NaI crystal to measure the LS light yield. Figure 3(a) shows a simplified diagram of the measurement. The measurement was done inside a dark box with two 2-inch PMTs: one attached to the LS and one attached to the NaI crystal. The distance between the LS and the NaI crystal was 40 cm. We placed a 137Cs source, which emits monoenergetic 0.662 MeV gamma rays, between the LS and the NaI crystal. When a gamma from 137Cs enters the LS, it can backscatter and the backscattered gamma can be detected by the NaI crystal. In this case, 0.48 MeV is deposited into the LS sample and 0.18 MeV is deposited into the NaI crystal (i.e. the energy loss in the LS and the NaI is Gaussian with peaks at these energies). Waveforms from both the LS and the NaI crystal were collected with a 14 bit Flash Analog-to-Digital Converter with 500 MS/s and a 1 s event window. For the LS, the total charge was computed by integrating the waveform from 50 ns before the peak to 250 ns after the peak. For the NaI crystal, the whole waveform was used. After computing the integrated charge, we selected backscattering events from the charge distribution of the NaI crystal. Figure 3(b) shows the LS charge distribution with a single Gaussian fit after the backscattering event selection. The red and blue histograms are two different samples from the same ISO tank taken 2 months apart. The fitted mean values, 3578 18 (red) and 3563 20 (blue, two months later), are the same within statistical error. The similarity in the measured charge distributions indicate that storing the LS in an ISO tank does not result in any significant degradation of the light yield over several months of exposure to the ISO tanks.
3.4 Temperature monitoring
The ISO tanks were leased from and stored by a company located in Kawasaki, Japan. A covered storage area was not available, so the ISO tanks stored outside. The ISO tanks are insulated, and therefore it is expected that the temperature of the LS will slowly follow the external temperature. We measured the temperature of the LS inside of each ISO tank every month. Figure 4 shows the temperature variation of the LS. The green squares (blue circles) indicate the temperature of the LS in ISO tank 1 (2). We compared the temperature of the LS with the average temperature in Yokohama (red line), which is the closest meteorological observatory with data available from the Japanese government’s climate database. These measurements show that the LS temperature is consistent with the average temperature in Yokohama, and we can expect that the LS will not get hotter than 30 degrees during the summer months. A study of LS aging by JUNO [7] showed that LS stored at 40 degrees did not show notable degradation. Therefore we do not expect the LS storage in Kawasaki to create any issues for JSNS2.
4 Conclusion
We produced 38000 L of LS with 3g/L of PPO and 30mg/L of bis-MSB at the RENO LS production site. The site was refurbished and thoroughly cleaned before the production of the LS. The measured water content in the LS was similar to the RENO experiment. Both the relative transmittance and light yield of the LS were measured and showed no significant degradation after transportation from Korea to Japan. The long term stability of the LS properties during storage inside of the ISO tanks was confirmed with measurements of the transmittance and light yield in Japan over a period of several months.
Acknowledgements
This work is supported by: the National Research Foundation of Korea (NRF) Grant No. 2009-0083526, NRF-2017K1A3A7A09016427, and 2018R1D1A1B07050425, funded by the Korea Ministry of Science and ICT. Some of us have been supported by a fund from the BK21 of the NRF; the JSPS grants-in-aid (Grant Number 16H06344, 16H03967), the Ministry of Education, Culture, Sports, Science and Technology (MEXT), Japan.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1[1] JSNS 2 Proposal, M. Harada et al., https://arxiv.org/abs/1310.1437
- 2[2] JSNS 2 TDR, S. Ajimura et al., https://arxiv.org/abs/1705.08629
- 3[3] Y. Hino et al., https://ar Xiv.org/abs/1904.08674
- 4[4] M. Yeh et al., Nuclear Instruments and Methods in Physics Research A 584 (2008) 238 - 243
- 5[5] https://alconox.com/resources/standarddocuments/tb/techbull_alconox.pdf
- 6[6] J. S. Park et al., Nuclear Instruments and Methods in Physics Research A 707 (2013) 45 - 53
- 7[7] Hai-tao Chen et al., Chin. Phys. C 39 (2015) 066002