Detection of low-conductivity objects using eddy current measurements with an optical magnetometer
Kasper Jensen, Michael Zugenmaier, Jens Arnbak, Hans St{\ae}rkind,, Mikhail V. Balabas, and Eugene S. Polzik

TL;DR
This paper demonstrates that an optical magnetometer can detect low-conductivity objects via eddy current measurements, significantly improving sensitivity and opening possibilities for non-invasive medical diagnostics.
Contribution
The study introduces a noise-canceling optical magnetometer capable of detecting low-conductivity objects with high sensitivity, surpassing previous methods.
Findings
Detected small salt-water containers with conductivity 4-24 S/m
Achieved over three orders of magnitude increase in SNR
Potential to detect objects with conductivity below 1 S/m
Abstract
Detection and imaging of an electrically conductive object at a distance can be achieved by inducing eddy currents in it and measuring the associated magnetic field. We have detected low-conductivity objects with an optical magnetometer based on room-temperature cesium atomic vapor and a noise-canceling differential technique which increased the signal-to-noise ratio (SNR) by more than three orders of magnitude. We detected small containers with a few mL of salt-water with conductivity ranging from 4-24 S/m with a good SNR. This demonstrates that our optical magnetometer should be capable of detecting objects with conductivity < 1 S/m with a SNR > 1 and opens up new avenues for using optical magnetometers to image low-conductivity biological tissue including the human heart which would enable non-invasive diagnostics of heart diseases.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
††thanks: corresponding author
email: [email protected]
Detection of low-conductivity objects using eddy current measurements with an optical magnetometer
Kasper Jensen
School of Physics and Astronomy, University of Nottingham, University Park, Nottingham NG7 2RD, England, United Kingdom
Niels Bohr Institute, University of Copenhagen, Blegdamsvej 17, 2100 Copenhagen, Denmark
Michael Zugenmaier
Niels Bohr Institute, University of Copenhagen, Blegdamsvej 17, 2100 Copenhagen, Denmark
Jens Arnbak
Niels Bohr Institute, University of Copenhagen, Blegdamsvej 17, 2100 Copenhagen, Denmark
Hans Stærkind
Niels Bohr Institute, University of Copenhagen, Blegdamsvej 17, 2100 Copenhagen, Denmark
Mikhail V. Balabas
Niels Bohr Institute, University of Copenhagen, Blegdamsvej 17, 2100 Copenhagen, Denmark
Department of Physics, St Petersburg State University, Universitetskii pr. 28, 198504 Staryi Peterhof, Russia
Eugene S. Polzik
Niels Bohr Institute, University of Copenhagen, Blegdamsvej 17, 2100 Copenhagen, Denmark
Abstract
Detection and imaging of an electrically conductive object at a distance can be achieved by inducing eddy currents in it and measuring the associated magnetic field. We have detected low-conductivity objects with an optical magnetometer based on room-temperature cesium atomic vapor and a noise-canceling differential technique which increased the signal-to-noise ratio (SNR) by more than three orders of magnitude. We detected small containers with a few mL of salt-water with conductivity ranging from 4–24 S/m with a good SNR. This demonstrates that our optical magnetometer should be capable of detecting objects with conductivity S/m with a SNR and opens up new avenues for using optical magnetometers to image low-conductivity biological tissue including the human heart which would enable non-invasive diagnostics of heart diseases.
Optical magnetometers [1] based on laser-interrogation of cesium or rubidium vapor can detect magnetic fields with sub-fT/ sensitivity [2, 3, 4, 5]. This high sensitivity is particularly useful for biomedical applications where tiny magnetic fields from the human body are detected. For example, optical magnetometers have detected brain activity [6, 7, 8], the heartbeat from adults [9] and fetuses [10, 11], and nerve impulses [12]. Optical magnetometers can potentially also be used to non-invasively image the electrical conductivity of the heart [13] using a technique called magnetic induction tomography (MIT) [14, 15]. In MIT of the heart, one or more coils are used to induce eddy currents in the heart and an image of the heart is constructed from measurements of the associated induced magnetic field. This is a challenging task for several reasons, with the main one being the low conductivity S/m of the heart [13].
Imaging of low-conductivity objects has previously been done using coils for inducing and detecting the eddy currents. Large containers ( mL) with salt-water with conductivity as low as 0.7 S/m has been imaged [14, 16, 17], and more recently, the spinal column has been imaged with a single scanning coil [18]. Optical magnetometers have several advantages compared to induction coils, in particular, they are widely tunable and can achieve high sensitivity which is fundamentally independent of the operating frequency. This is in contrast to induction coils which are sensitive to the change in magnetic flux and therefore have worse sensitivity the lower the frequency. So far, optical magnetometers have been used to image highly conductive metallic samples ( S/m) [19, 20, 21] and also recently semiconductor materials ( S/m) [22].
In this work, we introduce a differential technique which improves the signal-to-noise ratio by more than three orders of magnitude and then demonstrate detection of small containers with 8 mL of salt-water with conductivity as low as 4 S/m. This represents an improvement by two orders of magnitude compared to previous results with optical magnetometers [22] and is a big step towards magnetic induction tomography of biological tissue with optical magnetometers.
We first discuss the standard approach for detecting and imaging a conductive object, in our case a container with salt-water. Later we will discuss the differential technique. Consider a conductive object, a magnetometer, and a coil [denoted coil 1 in Fig. 1(a)] that generates a primary magnetic field oscillating at the frequency . The primary field induces eddy currents in the object which in turn generate a secondary magnetic field . One can measure the total field and by scanning the magnetometer or the object around it is possible to construct an image of the conductivity [19, 20, 21, 22]. Varying the frequency can be useful for 3D imaging [13] and for material characterization [20, 22].
It is instructive to note that the primary field is attenuated while penetrating into the object due to the skin effect. The skin depth is , where is the vacuum permeability and we assumed that the object is non-magnetic. When the thickness of the object is much smaller than the skin depth , the secondary field is out of phase with the primary field and the ratio of the amplitude of the secondary field to the amplitude of the primary field at the magnetometer position is [14, 23, 24]
[TABLE]
where is a geometrical factor with dimensions of length squared. For a container with salt-water with conductivity S/m we calculate cm and estimate [24] when the frequency is 2 MHz. We demonstrate that it is possible to detect such a small change in signal with an optical magnetometer when using a differential technique.
The key component of our magnetometer is a paraffin-coated cesium vapor cell with a inner volume [25]. The cesium atoms are spin-polarized in the -direction using circularly polarized pump and repump light and are detected using linearly polarized probe light [see Fig. 1(b)–(c)]. We denote the total angular momentum in the hyperfine ground state manifold and full polarization corresponds to , where is the number of cesium atoms. The atoms are placed in a static magnetic field and we are interested in detecting an oscillating magnetic field . The time evolution of the atomic spins is modelled using the differential equation [26]
[TABLE]
where is the cesium gyromagnetic ratio, , is the rate of optical pumping, is the decay rate in the absence of light, and is the decay rate due to the probe light. We solve the differential equation in the frame rotating around the -axis at the frequency . Denoting the spin-vector in the rotating frame and assuming a steady state , we find the spin-components
[TABLE]
Here , is the detuning of the applied frequency from the Larmor frequency , and .
If we only consider (i.e. ), we see that and have dispersive and lorentzian lineshapes, respectively, as a function of detuning. The total width of the resonance is where . This means that the resonance is power-broadened by the oscillating magnetic field . If the magnetic field is on resonance () we have which means that is only linear with the magnetic field for small fields .
The atoms are probed with linearly polarized light which due to the Faraday effect is rotated by an amount proportional to the spin-component along the probe propagation direction. The light polarization rotation is measured with a balanced detection scheme leading to the magnetometer signal
[TABLE]
The rotating spin-components and are extracted from the magnetometer signal using lock-in detection at the frequency . The lock-in provides an in-phase output and an out-of-phase output .
We characterize the magnetometer (without any conductive object) by applying the magnetic field . Figure 2(a) shows the lock-in outputs as a function of frequency. The - and -outputs have lorentzian and dispersive lineshapes centered around the Larmor frequency kHz as expected from Eqs. (4) and (5) when and . The small side resonances (towards lower frequencies) are due to the non-linear Zeeman effect [27, 28]. The dataset labeled “” in Fig. 2(b) shows the lock-in outputs when the oscillating magnetic field is on resonance (). We see that the mean values are V and and that there is a significant amount of noise in the -output. In order to characterize the noise we calculate the Allan deviation [29] of the -output which is roughly independent of averaging time with the value mV [see Fig. 2(c)]. The noise is mainly due to temporal fluctuations in the -field. A change in the -field shifts the Larmor frequency which then changes the -output. Close to the Larmor frequency where V/kHz [see Fig. 2(a)] which means that a small change of the -output of 22 mV corresponds to a shift in the Larmor frequency of 2.8 Hz and a relative change in the -field of . This small number illustrates that an optical magnetometer requires a very stable -field in order to precisely measure an oscillating magnetic field.
When detecting conductive objects, the amplitude of the secondary field is often much smaller than the amplitude of the primary field. This is the case if the object is much thinner than the skin depth or if the object is far away from the coil or magnetometer. For a thin sample, the secondary field is out of phase with the primary field such that with where . When detecting the total field , the field from the eddy currents gives a signal in the -output and the primary field gives a signal in the -output [see Eqs. (4) and (5) with , and ]. It is problematic to detect the total field for several reasons. First of all, one would like to increase the amplitude of the primary field as . However, when there is significant power broadening which leads to reduced signal size and non-linearities. Even if such that the magnetometer signal is linear and the lock-in outputs are and , it is still problematic to measure the total field as in most cases both the signal and the noise in the magnetometer are proportional to the amplitude of the total signal. In particular, if the dominant source of noise is the instability in the -field, then both signal and noise are proportional to . If we detect the total field, the smallest detectable field from the eddy currents is with . This is clearly not sufficient to detect low conductivity objects such as biological tissue or salt-water phantoms.
In order to mitigate the above mentioned problems, we introduce a differential technique where we use a second coil [denoted coil 2 in Fig. 1(a)] that generates a magnetic field such that in the absence of the conductive object, the total magnetic field at the position of the vapor cell. Coil 2 is placed further away from the conductive object than coil 1 such that eddy currents are mainly generated by coil 1 only. With this technique, the magnetometer signal is zero in the absence of the conductive object and the magnetometer should not be affected by power broadening or non-linearities as long as the field from the eddy currents is smaller than [see Eqs. (4) and (5) with , and ]. Furthermore, the signal-to-noise ratio of the measurement will improve by a factor if the noise in the magnetometer is proportional to the total signal. This is a dramatic improvement as for our measurements on salt-water.
Figure 2(b) shows three data sets. We see the noisy signal when is applied. When the opposite magnetic field is applied, the lock-in outputs change sign. Applying both magnetic fields at the same time (and increasing the amplitudes by a factor of 10) gives lock-in outputs close to zero with significantly reduced noise. Coil 1 and 2 are connected to two outputs of the same function generator and the amplitude and phase of the two outputs can be precisely set in order to zero the lock-in outputs. In Fig. 2(c) we see that the Allan deviation is times smaller for integration times s when applying both magnetic fields compared to only applying . Taking the factor of 10 into account, we find an improvement in signal-to-noise ratio of 1300 if detecting a low-conductivity object and a smallest detectable relative signal of when using the differential technique.
We now continue with detecting salt-water inside a small container. The conductivity of the water can be conveniently varied between 0–24 S/m by changing the concentration of salt. Using a motorized translation stage, we scan the container 50 mm in the -direction a few mm above coil 1. Real-time traces of the -output when the container is either empty or filled with salt-water with varying conductivity are shown in Fig. 3(a)–(d). With salt-water present, we clearly see a change in the -output when the container is on top of coil 1 (at the time around 10 s). The -outputs (not shown) are close to zero and do not change during the scan (within the statistical uncertainties). In order to reduce noise, the container is scanned times over coil 1 and the recorded traces are averaged. Figure 3(e) shows the relative change in signal for the averaged traces as a function of position. In order to guide the eye and to extract the maximum change in signal, we fit the data with salt-water to a Gaussian function. For the S/m data we have the maximum change which agrees reasonably well with the expected value [24]. The maximum change in signal is plotted in Fig. 4(a) as a function of conductivity and we observe a linear dependence as expected from Eq. (1) confirming that the small observed signals are due to the salt-water. We also vary the applied frequency (while at the same time adjusting the bias field to fulfill the resonance condition ), and as shown in Fig. 4(b) we again observe the expected linear behavior. Finally we vary the amplitude of the applied field. The signal starts out growing linearly but then some saturation occurs for higher amplitudes [see Fig. 4(c)]. The data are fitted to the function and we extract the saturation parameter V. This saturation is not expected when using the differential technique. We note that when only is applied, saturation happens at 10 times lower amplitudes. To avoid issues related to saturation, we used the amplitude V for all other differential measurements (and V for all measurements with one coil only).
We emphasize that we detect the salt-water with good signal-to-noise ratio (SNR). We calculate the SNR as the maximum change in signal divided by the standard deviation (found from the data recorded with an empty container). For the traces in Fig 3(b)–(d) we have the SNR of 0.8, 2.5 and 6.1 for the conductivities 3.8, 10.7 and 24.1 S/m. For these traces, the integration time was only 40 ms. For the average traces in Fig 3(e) we have the SNR of 6.4, 20, and 46. This demonstrates that our setup should be capable of detecting objects with conductivity S/m with a SNR and that it should be possible to detect and image biological tissue which has conductivity S/m with our optical magnetometer.
In conclusion, we have demonstrated detection of small containers with salt-water with conductivity ranging from 4–24 S/m using an optical magnetometer and a differential technique which improved the signal-to-noise by more than three orders of magnitude. Our measurements were performed inside a magnetic shield, however, we expect that the differential technique will yield a larger improvement in unshielded conditions [30] as there will be more magnetic field noise which can be canceled. The technique also gives a large improvement when detecting objects with high conductivity (such as metal objects) as long as the detected field from the eddy currents is small compared to the primary field which for instance is the case when the object is far away. We note that during the preparation of this manuscript similar techniques have been reported and used for imaging of structural defects in metal samples [31]. By further optimizing the sensitivity and long term stability of our magnetometer we expect that high-resolution imaging of biological tissue will be possible. This will make optical magnetometers promising candidates for localizing conduction disturbances in the heart allowing for non-invasive diagnostics of heart diseases such as, for example, atrial fibrillation[13].
Acknowledgements.
We would like to thank M. A. Skarsfeldt for preparing the salt-water solutions and B. H. Bentzen and S.-P. Olesen for motivating discussions. This work was supported by the Danish Quantum Innovation Center (QUBIZ)/Innovation Fund Denmark, the ERC AdG QUANTUM-N and the EU project MACQSIMAL.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1[1] Dmitry Budker and Michael Romalis. Optical magnetometry. Nature Physics 3 , 227 (2007).
- 2[2] I. K. Kominis, T. W. Kornack, J. C. Allred, and M. V. Romalis. A subfemtotesla multichannel atomic magnetometer. Nature 422 , 596 (2003).
- 3[3] I. M. Savukov, S. J. Seltzer, M. V. Romalis, and K. L. Sauer. Tunable atomic magnetometer for detection of radio-frequency magnetic fields. Phys. Rev. Lett. 95 , 063004 (2005).
- 4[4] W. Wasilewski, K. Jensen, H. Krauter, J. J. Renema, M. V. Balabas, and E. S. Polzik Quantum noise limited and entanglement-assisted magnetometry. Phys. Rev. Lett. 104 , 133601 (2010).
- 5[5] W. Chalupczak, R. M. Godun, S. Pustelny, and W. Gawlik. Room temperature femtotesla radio-frequency atomic magnetometer. Appl. Phys. Lett. 100 , 242401 (2012).
- 6[6] H. Xia, A. Ben-Amar Baranga, D. Hoffman, and M. V. Romalis. Magnetoencephalography with an atomic magnetometer. Appl. Phys. Lett. 89 , 211104 (2006).
- 7[7] E. Boto et al. A new generation of magnetoencephalography: Room temperature measurements using optically-pumped magnetometers. Neuro Image 149 , 404 (2017).
- 8[8] E. Boto et al. Moving magnetoencephalography towards real-world applications with a wearable system. Nature 555 , 657 (2018).