Development of an "$\alpha$-ToF" detector for correlated measurement of atomic masses and decay properties
T. Niwase, M. Wada, P. Schury, H. Haba, S. Ishizawa, Y. Ito, D. Kaji,, S. Kimura, H. Miyatake, K. Morimoto, K. Morita, M. Rosenbusch, H. Wollnik, T., Shanley, Y. Benari

TL;DR
The paper introduces an innovative $\alpha$-ToF detector that enables precise, correlated measurements of atomic masses and decay properties of short-lived isotopes, significantly improving background suppression and measurement accuracy.
Contribution
It presents a novel $\alpha$-ToF detector design that enhances correlated measurement capabilities for radioactive isotopes in mass spectrometry.
Findings
Time-of-flight detection efficiency >90% for 5.48 MeV $\alpha$-rays
Time resolution of 251.5(68) ps achieved
Effective resolution of mixed $\alpha$ sources in 2D spectrum
Abstract
We have developed a novel detector, referred to as an "-ToF detector", for correlated measurements of atomic masses and decay properties of low-yield, short-lived radioactive isotopes using a multi-reflection time-of-flight mass spectrograph. By correlating measured time-of-flight signals with decay events, it will be possible to suppress background events and obtain accurate, high-precision mass values even in cases of very low event rates. An offline test of the -ToF detector has shown that the time-of-flight detection efficiency for 5.48~MeV -rays is more than 90\% and yields a time resolution of 251.5(68)~ps and an energy resolution of 141.1(9)~keV. Using a two-dimensional spectrum of the correlated -ray energy and time-of-flight, the -rays from mixed sources could be fairly well resolved.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Development of an “-ToF” detector for correlated measurement of atomic masses
and decay properties
T. Niwase
M. Wada
P. Schury
H. Haba
S. Ishizawa
Y. Ito
D. Kaji
S. Kimura
H. Miyatake
K. Morimoto
K. Morita
M. Rosenbusch
H. Wollnik
T. Shanley
Y. Benari
Department of Physics, Kyushu University, Fukuoka City, Fukuoka, Japan
RIKEN, Nishina Center for Accelerator-Based Science, Wako City, Saitama, Japan
Wako Nuclear Science Center(WNSC), Institute of Particle and Nuclear Studies(IPNS),
High Energy Accelerator Research Organization(KEK), Wako City, Saitama, Japan
Japan Atomic Energy Agency(JAEA), Tokai, Ibaraki, Japan
Graduate School of Science and Engineering, Yamagata University, Yamagata, Japan
Department of Chemistry and Biochemistry, New Mexico State University, Las Cruces, NM, USA
ETP Ion Detect, 8 Martha Street, Clyde, New South Wales, Australia
Abstract
We have developed a novel detector, referred to as an “-ToF detector”, for correlated measurements of atomic masses and decay properties of low-yield, short-lived radioactive isotopes using a multi-reflection time-of-flight mass spectrograph. By correlating measured time-of-flight signals with decay events, it will be possible to suppress background events and obtain accurate, high-precision mass values even in cases of very low event rates. An offline test of the -ToF detector has shown that the time-of-flight detection efficiency for 5.48 MeV -rays is more than 90% and yields a time resolution of 251.5(68) ps and an energy resolution of 141.1(9) keV. Using a two-dimensional spectrum of the correlated -ray energy and time-of-flight, the -rays from mixed sources could be fairly well resolved.
keywords:
MRTOF-MS, Mass spectrometry, Time-of-flight detector, Superheavy elements, decay
1 Introduction
The multi-reflection time-of-flight mass spectrograph (MRTOF-MS) was invented nearly 30 years ago for general purpose mass measurements [1]. In 2013, such devices were used to measure the masses of radioactive nuclei at three different laboratories for the first time [2, 3, 4]. Since then, the device has been acknowledged to be useful for mass measurement of short-lived rare isotopes. We performed mass measurements of fusion-evaporation products [5, 6, 7, 10] provided at the RIKEN gas-filled recoil ion separator GARIS-II [8] and showed that an MRTOF-MS is applicable to short-lived nuclei having a half-life of 10 ms by achieving a relative mass precision of \delta$$m/m$$\sim10*-7* [9], and to very low-yield isotopes by mass analyzing 249Md which had a detection rate of one ion per 1000 seconds [10].
We plan to measure even lower-yield isotopes, such as superheavy nuclei which will have yield of one ion per day or less. In such a measurement, reliable discrimination of background events is essential. Decay properties – such as emitted particles and their energies, along with decay half-lives – can be used as fingerprints of radioactive isotopes. If we can correlate such decay properties with the time-of-flight (ToF) signal, we will be able to distinguish a true event from background events which can originate from stable, possibly molecular, contaminant ions as well as dark counts in the detector.
For this purpose, we have developed a new detector, referred to as an “-ToF detector”, which can measure an ion’s time-of-flight as well as its decay energy and absolute time of subsequent decays from implanted radioactive ions. In addition to background reduction in mass measurements, this detector will be used for nuclear spectroscopy.
2 Apparatus
The -ToF detector is a modified commercial MagneToF detector (ETP 14DM572), as shown in Fig. 1. It consists of an impact plate, an internal circuit for generating a suitable potentials, permanent magnets, and an electron multiplier. When an ion hits the impact plate of the MagneToF detector, secondary electrons are emitted and those electrons are isochronously guided by crossed electric and magnetic fields to the electron multiplier amplifying section of the detector to provide a fast timing signal for the ions. We replaced the impact plate with a silicon surface barrier detector (SSD, Hamamatsu S-3590) whose surface was coated for efficient secondary electron emission, where both Au+Al2O3 and Au+MgO were tested.
In order to maintain the potential of the impact plate, the gold coated surface area was wire-bonded to the terminal of the front side of the SSD, which is electrically connected to the impact plate. MgO and Al2O3 are materials known to have high secondary electron emission probabilities [11, 12, 13] and no deliquescence. The gold and Al2O3 coatings were made by a vacuum evaporation method, while the MgO coating was made by a sputtering method. The thickness of each layer is roughly ten nanometers. Photographs of the impact plate and the -ToF detector assembly are shown in Fig.2.
In normal operation, the -ToF detector’s impact plate must be placed at a negative high potential, approximately kV. Since the SSD is also placed on the impact plate potential, an optically isolated circuit was developed to provide a bias voltage to the SSD, and to amplify the decay signal. A sketch of the design for such an optical isolation system is shown in Fig. 3. The signal from the Si detector is amplified by two charge sensitive pre-amplifiers having different gains: a high-gain one for -ray signals and a low-gain one for spontaneous fission signals. The amplified analog signals are sent to shaping amplifiers on ground potential through optical transceivers. The data from the Si detector and the time-of-flight data are separately recorded event by event with absolute time stamps.
3 Offline performance tests
The new detector was thoroughly tested offline. In particular, the detection efficiency and timing resolution were confirmed to be sufficient to perform atomic mass measurements. In addition, the new detector’s fitness for particle identification was confirmed by -energy correlated time-of-flight (-ToF) measurements.
3.1 Detection efficiency
We measured the detection efficiency of the -ToF detector by using collimated 5.48 MeV -rays from a 241Am source. As the SSD should have near unity detection efficiency for 5.48 MeV -rays, the efficiency was defined as the ratio of coincident ToF signals to the -singles counts. The counting rate was set to be as low as 0.1 cps by using a collimator. A ToF signal was considered to be coincident when the timing signal were detected within 500 ns of each other. The -singles signals were limited to those signals with an SSD signal amplitude commensurate with 241Am -rays. Fig. 4 shows the detection efficiency measured for both MgO and Al2O3 surface coatings as functions of the high-voltage potential applied to the impact plate. At the nominal operation potential of V, the efficiency was more than 90% for both surface coatings.
3.2 Time resolution
The reference timing signal for the time-of-flight was provided by a Busch type triangle-roof time-of-flight detector [14, 15] located 27 cm upstream from the -ToF detector as shown in Fig. 5(a) The time resolution was determined from the time-of-flight spectrum. In the following discussion, the time resolution of the triangle-roof detector and the -ToF detector are indicated as and , respectively. The measured time resolution, being a convolution of these two, is written .
As the start signal was given by the triangle-roof detector and the stop signal by the -ToF detector, the measured time resolution, ps. However, as this includes both the start timing of the triangle-roof detector and the stop timing in the -ToF, a comparison setup has been used as shown in Fig. 5(b). Two triangle-roof detectors of the same type have been placed opposite to each other, where one has been used as the start and the other as the stop signal. An SSD was placed behind the triangle-roof detector to produce a trigger signal as in the detection efficiency measurement. The standard deviation of the time differences was found to be 465.5(40) ps. In that way, the resolution of a single triangle-roof detector was deduced to be 328.5(28) ps. Using this value, we then determined the intrinsic time resolution of the -ToF detector to be 251.5(68) ps.
3.3 Correlation measurement
In order to make an offline determination of the -ToF detector performance in a correlated -ToF measurement, a mixed -source comprising 244Cm (5804 keV), 241Am (5485 keV) and 237Np (4788 keV) was utilized.
Due to its unity detection efficiency, the signal from the SSD was used to trigger the data acquisition system (Fig. 6), as the common start signal of the time-to-digital converter (TDC), and to provide -ray energies. Because the SSD response is slow, the timing signal of the triangle-roof detector and of the -ToF detector were each connected to stop inputs of the TDC through analog delay lines. The true ToF was determined from the time difference between the two stop signals.
A VME based data acquisition system, with an analog to digital converter module (MADC-32) for energy signals and a TDC module (MTDC-32) for timing signals, was used. The TDC had a time resolution of 80 ps. Figure 7 shows the -ToF correlated data measured using the mixed -source. The projection of the two-dimensional spectrum (Fig. 7 (c)) onto the horizontal axis yields the singles time-of-flight spectrum. In the singles ToF spectrum, the three different -ray signals are not resolved. However, they can be resolved in a correlated measurement with the energy spectrum.
The energy resolution for the 5.48 MeV -rays was determined from the energy spectrum (Fig. 7 (b)) to be 141.1(9) keV, which is about twice the typical resolution of such an SSD. The degradation of the resolution is presumed to be due to noise from the MagneToF bias voltage supply.
4 Anticipated application for nuclear spectroscopy
It would be desirable for the -ToF detector to be used for nuclear decay spectroscopy. Table 1 shows candidate nuclides in the superheavy element region to be studied with a combination of an MRTOF-MS and the -ToF detector. The -ToF detector has the capability of correlated measurements for mass and decay properties. Using all the information – namely time-of-flight, decay energies, and decay times – obtained in the -ToF detector, we will be able to separate the ground state from the isomeric state and perform accurate mass measurements of each. As such, we have used the above-measured performance data to perform a numerical simulation to produce a three-dimensional parameter space figure like Fig. 8. Using the decay properties, we performed a numerical simulation that demonstrates the possibility to resolve an isomeric state in time-of-flight spectra for 263Sg and 267Hs.
Let us consider the isotope 263Sg for which two -decay energies, 9.06 MeV and 9.25 MeV, have been reported in the literature [16, 17]. However, the isomeric state and the ground state are not assigned. Lacking contrary evidence, for our simulation we assume the excitation energy of the isomeric state to be 200 keV – similar to the difference in reported -decay energies – and the population of the two states to be equal. Figure 9 shows the simulated results for 100 events. For an MRTOF-MS operating with a mass resolving power of =200,000 [10], the ToF of 263a,bSg2+ peaks in the spectrum are fully unresolved (Fig. 9 (c) - total). However, the -decay energy spectrum is sufficiently resolved so as to permit two dimensional fitting of the -energy and the ToF to 2 two-dimensional Gaussians as shown in Fig. 9. In this way, it is possible to separately produce resolved time-of-flight spectra for each state and to assign the states without ambiguity.
Let us further consider the isotope 267Hs. In this case, the energies of the -rays emitted by the two states would not be resolvable by our -ToF detector. However, it is reported [18] that the ground state and the isomeric state have greatly differing half-lives of 52 ms and 0.80 s, respectively. Due to the large half-life difference, we could use the decay time rather than the -energy to distinguish the states. Figure 10 (a) shows the results of a simulated measurement of 267Hs2+ in terms of a ToF versus decay time scatter plot, where the decay time is the time interval between the ToF signal and the -decay signal. Within 200 ms of striking the -ToF detector, 94 % of the 267gHs will have decayed. In the same time, less than 20 % of the 267mHs will decay. Thus, by gating on -decay events occurring more than 200 ms after the ToF signal is observed it is possible to produce a ToF spectrum which is almost purified from 267mHs ion, as shown in Fig. 10 (b).
5 Summary and outlook
We developed an -ToF detector for correlated measurements of the heavy-ion time-of-flight and the decay energy. We also developed a front-end signal processing circuit for the -ToF detector which is floated by a high voltage potential. The performance test of the -ToF detector showed an efficiency of more than 90% for the timing signal, an energy resolution of 141.1(9) keV for 5.48 MeV -rays, and a time resolution of 251.5(68) ps.
Although the energy resolution is somewhat degraded, it should still be sufficient to resolve many isomeric states. Efforts to reduce the noise, in collaboration with ETP Ion Detect, are ongoing. The measured off-line performance is sufficient to meet the initial requirements for an on-line detection system. As such, the newly developed -ToF detector has been installed as the detection system in an already-commissioned MRTOF-MS for online mass measurements at the RIKENs RI-Beam Factory. In the near future, first on-line testing will be performed using medium-yield (0.1 s*-1*) -emitting nuclei produced in fusion-evaporation reactions. Following successful on-line commissioning, it is planned that first measurements with very low-yield superheavy elements will commence.
Acknowledgments
This work was financially supported by the Japan Society for the Promotion of Science KAKENHI (Grant No. 17H06090). One of the author (T.N.) thanks the RIKEN Junior Research Associate Program.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1[1] H. Wollnik, M. Przewioka, Time-of-flight mass spectrometers with multiply reflected ion trajectories, Int. J. Mass Spectrom. Ion Process 96 (1990) 267. doi:10.1016/0168-1176(90)85127-N . · doi ↗
- 2[2] Y. Ito, P. Schury, M. Wada, S. Naimi, T. Sonoda, H. Mita, F. Arai, A. Takamine, K. Okada, A. Ozawa, H. Wollnik, Single-reference high-precision mass measurement with a multireflection time-of-flight mass spectrograph, Phys. Rev. C 88 (2013) 011306(R). doi:10.1103/Phys Rev C.88.011306 . · doi ↗
- 3[3] R. N. Wolf, F. Wienholtz, D. Atanasov, D. Beck, K. Blaum, C. Borgmann, F. Herfurth, M. Kowalska, S. Kreim, Y. A. Litvinov, D. Lunney, V. Manea, D. Neidherr, M. Rosenbusch, L. Schweikhard, J. Stanja, K. Zuber, Isoltrap’s multi-reflection time-of-flight mass separator/spectrometer, Int. J. Mass spectrum. 349-350 (2013) 123–133. doi:10.1016/j.ijms.2013.03.020 . · doi ↗
- 4[4] W. R. Plass, T. Dickel, C. Scheidenberger, Mutiple-reflection time-of-flight mass spectrometry, Int. J. Mass spectrum. 349-350 (2013) 134–144. doi:10.1016/j.ijms.2013.06.005 . · doi ↗
- 5[5] P. Schury, M. Wada, Y. Ito, D. Kaji, F. Arai, M. Mac Cormick, I. Murray, H. Haba, S. Jeong, S. Kimura, H. Koura, H. Miyatake, K. Morimoto, K. Morita, A. Ozawa, M. Rosenbusch, M. Reponen, P. A. Söderström, A. Takamine, T. Tanaka, H. Wollnik, First online multireflection time-of-flight measurements of isobar chains produced by fusion-evaporation reactions: toward identification of superheavy elements via mass spectroscopy, Phys. Rev. C 95 (2017) 011305. doi:10.1103/Phys Rev C.95.011305 . · doi ↗
- 6[6] S. Kimura, Y. Ito, D. Kaji, P. Schury, M. Wada, H. Haba, T. Hashimoto, Y. Hirayama, M. Mac Cormick, H. Miyatake, J. Y. MOON, K. Morimoto, M. Mukai, I. Murray, A. Ozawa, M. Rosenbusch, H. Schatz, A. Takamine, T. Tanaka, Y. X. Watanabe, H. Wollnik, Atomic masses of intermediate-mass neutron-deficient nuclei with relative uncertainty down to 35-ppb via multireflection time-of-flight mass spectrograph, Int. J. Mass spectrum. 430 (2018) 134–142.
- 7[7] M. Rosenbusch, Y. Ito, P. Schury, M. Wada, D. Kaji, K. Morimoto, H. Haba, S. Kimura, H. Koura, M. Mac Cormick, H. Miyatake, J. Y. MOON, K. Morita, I. Murray, T. Niwase, A. Ozawa, M. Ropenen, A. Takamine, T. Tanaka, H. Wollnik, New mass anchor points for neutron-deficient heavy nuclei from direct mass measurements of radium and actinium isotopes, Phys. Rev. C 97 (2018) 064306. doi:10.1103/Phys Rev C.97.064306 . · doi ↗