A Coincidence Velocity Map Imaging Spectrometer for Ions and High-Energy Electrons to Study Inner-Shell Photoionization of Gas-Phase Molecules
Utuq Ablikim, C\'edric Bomme, Timur Osipov, Hui Xiong, Razib Obaid,, Ren\'e C. Bilodeau, Nora G. Kling, Ileana Dumitriu, Sven Augustin, Shashank, Pathak, Kirsten Schnorr, David Kilcoyne, Nora Berrah, Daniel Rolles

TL;DR
This paper introduces a specialized coincidence velocity map imaging spectrometer designed for studying inner-shell photoionization in gas-phase molecules, enabling detailed electron-ion and ion-ion correlation measurements with high energy resolution.
Contribution
The paper presents a novel double-sided spectrometer setup optimized for coincidence detection of electrons and ions in inner-shell photoionization studies, demonstrating its capabilities with multiple molecular examples.
Findings
Successful measurement of valence photoelectron and ion spectra of neon and nitrogen.
Detailed channel-resolved photoelectron and Auger spectra for chlorine 2p ionization.
Observation of fragment-ion momentum correlations in dichloroethene.
Abstract
We report on the design and performance of a double-sided coincidence velocity map imaging spectrometer optimized for electron-ion and ion-ion coincidence experiments studying inner-shell photoionization of gas-phase molecules with soft X-ray synchrotron radiation. The apparatus employs two microchannel plate detectors equipped with delay-line anodes for coincident, time- and position-resolved detection of photo- and Auger electrons with kinetic energies up to 300\,eV on one side of the spectrometer and photoions up to 25\,eV per unit charge on the opposite side. We demonstrate its capabilities by measuring valence photoelectron and ion spectra of neon and nitrogen, and by studying channel-resolved photoelectron and Auger spectra along with fragment-ion momentum correlations for chlorine inner-shell ionization of \textit{cis}- and \textit{trans}-1,2-dichloroethene.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
A Coincidence Velocity Map Imaging Spectrometer for Ions and High-Energy Electrons to Study Inner-Shell Photoionization of Gas-Phase Molecules
Utuq Ablikim
J.R. Macdonald Laboratory, Department of Physics, Kansas State University, Manhattan, KS 66506, USA
Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, CA 94720, USA
Cédric Bomme
Deutsches Elektronen-Synchrotron (DESY), 22607 Hamburg, Germany
Timur Osipov
SLAC National Accelerator Laboratory, Menlo Park, CA 94025, USA
Hui Xiong
Department of Physics, University of Connecticut, Storrs, CT 06269, USA
Razib Obaid
Department of Physics, University of Connecticut, Storrs, CT 06269, USA
René C. Bilodeau
Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, CA 94720, USA
Department of Physics, University of Connecticut, Storrs, CT 06269, USA
Nora G. Kling
Department of Physics, University of Connecticut, Storrs, CT 06269, USA
Ileana Dumitriu
Hobart and William Smith Colleges, Geneva, NY 14456, USA
Sven Augustin
J.R. Macdonald Laboratory, Department of Physics, Kansas State University, Manhattan, KS 66506, USA
Max Planck Institute for Nuclear Physics, 69117 Heidelberg, Germany
Shashank Pathak
J.R. Macdonald Laboratory, Department of Physics, Kansas State University, Manhattan, KS 66506, USA
Kirsten Schnorr
Max Planck Institute for Nuclear Physics, 69117 Heidelberg, Germany
Department of Chemistry, University of California, Berkeley, CA 94720, USA
David Kilcoyne
Advanced Light Source, Lawrence Berkeley National Laboratory, Berkeley, CA 94720, USA
Nora Berrah
Department of Physics, University of Connecticut, Storrs, CT 06269, USA
Daniel Rolles
J.R. Macdonald Laboratory, Department of Physics, Kansas State University, Manhattan, KS 66506, USA
Abstract
We report on the design and performance of a double-sided coincidence velocity map imaging spectrometer optimized for electron-ion and ion-ion coincidence experiments studying inner-shell photoionization of gas-phase molecules with soft X-ray synchrotron radiation. The apparatus employs two microchannel plate detectors equipped with delay-line anodes for coincident, time- and position-resolved detection of photo- and Auger electrons with kinetic energies up to 300 eV on one side of the spectrometer and photoions up to 25 eV per unit charge on the opposite side. We demonstrate its capabilities by measuring valence photoelectron and ion spectra of neon and nitrogen, and by studying channel-resolved photoelectron and Auger spectra along with fragment-ion momentum correlations for chlorine inner-shell ionization of cis- and trans-1,2-dichloroethene.
velocity map imaging, photoionization, synchrotron radiation, electron-ion coincidences, photoelectron spectroscopy, ion momentum imaging
.
I INTRODUCTION
Since the pioneering work of Houston and Chandler Chandler and Houston (1987) and Eppink and Parker Eppink and Parker (1997), charged-particle imaging and, in particular, velocity map imaging (VMI) have become a wide-spread technique to study photoionization, dissociation, and gas-phase chemical reactions dynamics in a wide variety of targets and with a number of different light sources. Chandler, Houston, and Parker (2017); Suits and Continetti (2001); Reid (2012) Common applications of the VMI technique range from experiments with laboratory sources such as gas-discharge lamps, picosecond and femtosecond laser systems, and high harmonic generation sources to studies performed on large-scale facilities such as synchrotron radiation sources and free-electron lasers. Velocity map imaging is particularly important for measurements with extreme ultraviolet (XUV) and X-ray photons since the radiation can ionize along the entire beam path, not just in a high-intensity focal region, as is it the case, e.g., for femtosecond near-infrared lasers. In order to obtain good energy resolution, it is therefore crucial to focus charged particles with the same velocity but originating from different starting points in the interaction region to the same spot on the detector.
In parallel to the emergence of the VMI, coincidence momentum imaging methods, such as COLTRIMS Dörner et al. (2000); Ullrich et al. (2003, 1997) and related techniques Davies et al. (1999); Lafosse et al. (2000); Dowek et al. (2002), also developed into powerful tools for imaging gas-phase photoionization and photofragmentation dynamics. In recent years, several ”hybrid” spectrometers have been developed that use elements of both traditional VMI and COLTRIMS-style spectrometers, thereby overcoming some of the limitations of each of the techniques. Examples of such double-sided hybrid spectrometers for coincident electron and ion detection include combinations of a VMI spectrometer with a traditional time-of-flight or electrostatic cylindrical analyzer opposite to the VMI side Garcia, Nahon, and Powis (2004); Garcia et al. (2005); Rolles et al. (2007); Pešić et al. (2007); Red et al. (2010); Bodi et al. (2009); O’Keeffe et al. (2011), as well as double-sided VMI spectrometers with two position-sensitive detectors for coincident momentum-resolved imaging of both, electrons and ions Strüder et al. (2010); Rolles et al. (2014); Takahashi, Cave, and Eland (2000); Vredenborg, Roeterdink, and Janssen (2008a), many of which were specifically designed for photoionization studies using synchrotron radiation Garcia et al. (2013); Bomme et al. (2013); Sztáray et al. (2017); Hosaka et al. (2006); Tang et al. (2009); Bodi et al. (2012). With few exceptions Hosaka et al. (2006), the latter are mostly optimized for and limited to rather low-kinetic energy electrons. Here, we present the design of a double-sided VMI spectrometer capable of detecting high-energy electrons up to 300 eV in order to allow for electron-ion and ion-ion coincidence measurements Ablikim et al. (2016, 2017); Xiong et al. (2017) and for the angle-resolved detection of Auger electrons produced by inner-shell photoionization with soft X-ray synchrotron radiation. As described in detail in section III, this enables, e.g., detection of Auger electrons in coincidence with momentum-resolved fragment ions in order to measure channel-resolved Auger spectra, thereby disentangling congested molecular Auger spectra into individual contributions from specific ionic final states.
Instead of a double-sided design as presented here, an alternative way of realizing electron-ion coincidence experiments with a VMI spectrometer is to use a single-sided spectrometer where the extraction fields are switched after the electron detection in order to detect the ions on the same detector Lehmann, Ram, and Janssen (2012); Zhao, Sándor, and Weinacht (2017); Fan et al. (2017). While the latter allows for a fully independent choice of extraction fields for electrons and ions and can reduce the cost of the apparatus since only one position-sensitive detector is needed, the high-voltage switching typically generates a large amount of ringing on the detector that can make it difficult or impossible to detect ions with short flight times, such as H+. In general, it also reduces the momentum resolution of the ions, although the actual amount depends on the quality of the high-voltage pulse and other details of the experiment.
II APPARATUS
II.1 Vacuum system and molecular beam setup
The ultrahigh vacuum (UHV) system of the coincidence end-station consists of three differentially pumped chambers: (i) The interaction chamber containing the double-sided VMI spectrometer; (ii) the molecular beam source; and (iii) a differential pumping section to decouple the experimental chamber from the beamline vacuum. The interaction chamber is pumped by two turbomolecular pumps (600 l/s and 360 l/s) and typically reaches a base pressure in the low 10*-8* mbar without bake-out. It has an internal double -metal shielding to minimize the influence of the earth’s magnetic field and other stray magnetic fields on the electron trajectories. The continuous molecular beam is mounted on an xyz-manipulator and consists of a stainless steel tube terminated by 30 micron flat aperture that can be cooled by liquid nitrogen or (moderately) heated by a resistive heater. The molecular beam expansion chamber is pumped by two 1000 l/s turbomolecular pumps. Typically, a single skimmer with 500 m diameter separates the expansion chamber from the interaction chamber, but a second pumping stage with an additional skimmer can be added, depending on the experimental requirements.
II.2 Spectrometer design and performance
Although the VMI apparatus can, in principle, be used at different light sources and different beamlines, it was designed with the parameters of beamline 10.0.1.3 of the Advanced Light Source at the Lawrence Berkeley National Laboratory in mind since it is most frequently being used at this beamline. Beamline 10.0.1.3 delivers VUV and soft X-ray radiation in the photon energy range between 17 eV and 350 eV with a resolving power of 35,000 at 60 eV Kaindl et al. (1995); Warwick et al. (1995), covering, for example, the first inner-shell ionization thresholds of many halogens such as chlorine, bromine, and iodine, the L-shell ionization threshold of sulfur, and the K-shell ionization threshold of carbon. In order to be able to detect all photo- and Auger electrons over the entire photon energy range, it is therefore necessary that the spectrometer can detect electrons up to 300 eV with a full-solid-angle acceptance. Furthermore, since the focus of our research lies on the photoionization and fragmentation dynamics of molecules after inner-shell photoionization, typical fragment ion kinetic energies can reach up to 25 eV, and up to three or even more ionic fragments need to be detected in coincidence and with good time and position resolution in order to be able to reconstruct their three-dimensional momentum vectors.
Since the electron and the ion side in a double-sided VMI spectrometer share a common extraction field, their performance is interconnected. Therefore, the lengths of the drift tubes on both sides need to be designed according to the required kinetic energy ranges mentioned above. Fig. 1 shows the dimensions of the spectrometer along with typical voltage settings and the corresponding SIMION simulations. Electrons and ions created in the interaction region located in the middle between the electron and ion extractor electrodes are accelerated by static electric fields towards two opposite MCP detectors with 80 mm diameter. In addition to the two extractor electrodes, two additional electrodes (”electron focus” and ”ion focus”) provide tunable electrostatic lenses that help focusing the extended interaction region Rolles et al. (2007) and help compensate for field distortions due to the open extractor/repeller geometry Vredenborg, Roeterdink, and Janssen (2008b). Varying the voltage on these focusing electrodes will shift which kinetic energy range is focused best, and thus has the best kinetic energy resolution.
For the simulation shown here, the electrode voltages of the VMI spectrometer were chosen to allow for the collection of electrons up to 290 eV and ions up to 25 eV per unit charge over the full solid angle. This was achieved by applying +180 V and -180 V to the two inner-most extractor electrodes, +1080 and -1260 V to the two focusing electrodes, and +5580 V/-4770 V to the two drift tubes.
In order to detect electrons and ions in coincidence and to record both position and time information for the charged fragments, which is necessary to determine their three-dimensional momentum vectors, the double-sided VMI is equipped with microchannel plate (MCP) detectors with multi-hit delay-line anodes (RoentDek DLD80 for the electrons and RoentDek HEX80 for the ions). The analog MCP and delay line signals are amplified, processed by a constant fraction discriminator (CFD), and then recorded by the data acquisition system consisting of two RoentDek TDC8HP 8-channel multi-hit time-to-digital converters (TDC). The TDCs have a resolution of 100 ps and a multi-hit dead-time of 10 ns. They are triggered by the detection of an electron (which could be either a photoelectron or an Auger electron), which typically arrives at the detector after a flight time of approximately 5 ns. The ion time of flight is then measured with respect to the arrival time of the first detected electron. This allows the experiment to be performed during the standard ALS multi-bunch top-off mode of operation, which has an electron bunch spacing in the storage ring of 2 ns. In the ALS’s two-bunch timing mode of operation, the spacing between the soft X-ray pulses is large enough to unambiguously attribute the detected electrons to a specific light pulse, thus making it possible to also determine the electron flight times by using the ALS bunch marker signal as a timing reference.
II.3 Two-dimensional electron imaging
Since the time-of-flight spread of the electrons due to their initial momentum along the spectrometer axis is very small and since the spacing between electron bunches in the storage ring in the multi-bunch operation mode is not sufficient to link the detected photo- or Auger electrons to a specific soft X-ray pulse, we typically do not attempt to retrieve any momentum information from the electron time of flight. Instead, the electron images are treated as two-dimensional projections of the three-dimensional momentum sphere, and an image inversion procedure is employed to retrieve electron kinetic energy spectra and angular distributions as typically done for ”conventional” VMI spectrometers. To demonstrate the performance of the electron side of the double-sided spectrometer and to characterize its kinetic energy resolution, photoelectron images for the valence photoionization of neon atoms were recorded over a range of photon energies between 50 and 150 eV.
A raw electron image recorded at a photon energy of 125 eV is shown on the left hand side of Fig. 2(a), together with a central slice through the three-dimensional momentum sphere, obtained by applying an inverse Abel transformation Hickstein et al. , shown on the right hand side of the same image. The corresponding angle-integrated photoelectron spectrum obtained from the Abel inversion is shown in Fig. 2(b). By determining the width of the Ne(2) photoelectron peak as a function of photon energy and thus of photoelectron kinetic energy, the resolution of the electron side of the spectrometer can be characterized, as shown in Fig. 2(c). We neglect the line broadening due to the photon bandwidth, which was less than 10 meV for the monochromator exit slit settings chosen for this measurement. Note that in order to correct for distortions of the electron images, which severely affect the achievable kinetic energy resolution, a circularization procedure has been applied to the raw images before the inversion. Hickstein et al. ; Gascooke, Gibson, and Lawrance (2017)
Given the known Ne() and Ne() ionization potentials of 21.6 eV and 48.5 eV, respectively, the neon measurement can also be used to determine the calibration function for converting the radius of the inverted electron images into electron kinetic energy, as shown in Fig. 3. Experimental data and SIMION trajectory simulations are in good agreement, verifying that the latter can safely be used for calibrating the electron kinetic energies.
II.4 Three-dimensional ion momentum imaging
In order to test and characterize the momentum imaging capabilities of the ion side, we have performed an electron-ion-ion coincidence experiment on the dissociative valence double photoionization of N2 and O2. N2 and O2 molecules were introduced into the interaction chamber as a supersonic jet by directly opening the gas line to ambient air. The spectrometer voltages for this measurement were +100 V and -100 V on the two inner-most extractor electrodes, +600 and -700 V on the two additional focusing lenses, and +3100 V and -2650 V on the two drift tubes.
Fig. 4(a) shows the resulting photoion-photoion coincidence (PIPICO) plot for photoionization at 160 eV photon energy. The elongated tails towards longer flight times that are visible for each coincidence channel are due to fragmentation of ”warm” N2 and O2 molecules that have either leaked through the skimmer from the expansion chamber into the main chamber or that were present as residual gas. They are ionized along the whole light path. The resulting fragment ions are not focused properly and, as confirmed by SIMION simulations, have longer flight times since they are not produced in the center of the spectrometer. However, their contribution can be significantly reduced by selecting a narrow gate in the PIPICO spectrum.
Using the time-of-flight and the detector hit position, the three-dimensional momentum of each ion can be reconstructed and the total kinetic energy release (KER) spectrum of the N++N+ channel can be extracted, as shown as a black line in Fig. 4(b). To demonstrate the good momentum resolution that can be achieved by our spectrometer, the KER spectrum is compared to a KER spectrum measured for N2 inner-shell ionization obtained using a COLTRIMS apparatus Weber et al. (2001). The width of the two sharp peaks at 8 and 10 eV are comparable in both measurements (the FWHM of the peak at 10 eV is 500 meV (FWHM) in the COLTRIMS measurement and 580 meV (FWHM) in the present experiment). Note that the COLTRIMS data was recorded after inner-shell ionization, where the doubly-charged intermediate state is reached via Auger decay, while the present data was recorded after valence ionization, where the doubly-charged intermediate state is reached via a shake-off process, which may lead to different branching ratios. In addition, for the data analysis shown here, only two layers of the HEX detector were used, which leads to some dead time for fragments with identical mass-to-charge ratio. The dead time depends on both, kinetic energies and angular distributions of the fragments and may therefore affect certain parts of the KER spectrum, such as the two sharp peaks, more than other parts.
Since the electric field in a VMI spectrometer is not homogeneous, one cannot derive analytical formulas to reconstruct the ion momenta from the measured time of flight and hit positions of each ion. Instead, we use the SIMION 8.1 software package to simulate the expected time of flight and hit positions for ions starting in the interaction region with different kinetic energies and emission angles.
By varying the starting position of the ions in the SIMION simulation within a sphere of 1 mm diameter (which does not reflect the assumed size of the interaction region but rather the uncertainty with which the exact position is known), the uncertainty of the ion kinetic energy calibration is estimated to be on the order of 10% for ions with a kinetic energy of 1 eV and less than 5% for 10-eV ions. To verify the energy calibration, the KER spectrum of N2 molecules was measured (see Fig. 4(b)). The measured kinetic energies of several sharp features in this spectrum agree with literature values Lundqvist et al. (1996); Weber et al. (2001) to within 4%, confirming the above estimate of the energy uncertainty. This uncertainty will increase, and the overall performance of the spectrometer will suffer if the interaction region is not well centered between the two extractor electrodes. Careful and precise alignment of the spectrometer with respect to the synchrotron beam is therefore critical.
III Inner-shell ionization of cis- and trans-1,2-dichloroethene
In order to demonstrate the capabilities of our double-sided, high-energy coincidence VMI, this section presents results of an electron-ion-ion coincidence experiment studying the inner-shell ionization of - and -1,2-C2H2Cl2 molecules in the gas phase. The commercially available, isomerically pure samples (96% purity, Sigma Aldrich) are liquid at room temperature and were first outgassed in a freeze/pump/thaw cycle before introducing them into the gas phase via expansion through a 30 micron aperture without additional carrier gas. The molecular beam was crossed by a beam of linearly polarized soft X-ray radiation from the ALS (240 eV photon energy with 150 meV bandwidth) in the interaction region of the VMI spectrometer. The lens voltages of the VMI spectrometer were chosen as shown in Fig. 1 to allow for the collection of electrons up to 290 eV, singly charged ions up to 25 eV, and doubly charged ions up to 50 eV over the full solid angle.
A sketch of the molecular geometry of the and isomers along with the ion time-of-flight spectra recorded for each isomer at a photon energy of 240 eV (40 eV above the Cl() inner-shell ionization threshold) is shown in Fig. 5. The two spectra are normalized to have an equal yield of 35Cl+ ions and clearly show isomer-dependent differences in the branching ratios, most notably in the yield of the Cl fragments, which are created by bond recombination and almost exclusively occur after photoionization of -1,2-C2H2Cl2.
Further information about the fragmentation of 1,2-C2H2Cl2 can be obtained from the PIPICO and the photoion-photoion-photoion coincidence (PIPIPICO) maps shown in Fig. 6(a) and (b), respectively, for the isomer. In the PIPICO map, the time of flight of the second detected ion (time-of-flight2) is plotted against the time of flight of the first detected ion (time-of-flight1). Bright orange and yellow islands mark ionic fragments that arrive in coincidence, i.e., they originate from the same molecule that broke up as a consequence of the ionization event. In particular, sharp diagonal stripes, as shown, e.g., in the zoomed inset in the lower right of Fig. 6(a), indicate that the molecule broke up into two charged fragments that carry equal momentum, while broader features indicate breakup into more than two fragments with significant momentum. These can be further differentiated in the triple coincidence PIPIPICO map, where the time-of-flight sum of the second and third detected ion (time-of-flight2 + time-of-flight3) is plotted as a function of the time of flight of the first detected ion (time-of-flight1). Note that chlorine has two naturally occurring isotopes, 35Cl and 37Cl, which lead to several coincidence channels consisting of all possible combinations of these two isotopes. Additional channels result from the possible loss of one or both hydrogen atoms, all of which can be clearly resolved in the PIPICO and PIPIPICO maps. The strongest complete two-body fragmentation channel, C2H235Cl++35Cl+, and strongest complete three-body fragmentation channel, C2H+35Cl++35Cl+, are marked by white arrows in Fig. 6.
By selecting only those events where the ions appear in a specific ionic coincidence channel in the PIPICO or PIPIPICO map, it is possible to distinguish photo- and Auger electrons corresponding to different ionic final states and to produce channel-resolved photo- and Auger electron spectra, as shown in Fig. 7. The width of the Auger feature in the channel-resolved spectrum (red) is considerably narrower than in the non-coincident electron spectrum (green) and the spectrum where only one of the ionic fragments is determined (blue) since the latter two contain a multitude of Auger transitions, while only one or a few specific transitions lead to a given ionic final state. Measurement of electron-ion-ion and electron-ion-ion-ion coincidences can thus help with disentangling convoluted molecular Auger spectra, as will be discussed in more detail in a forthcoming publication.
Finally, as we have shown previously Ablikim et al. (2016), it is possible to distinguish between different geometric isomers by means of momentum-resolved ion-ion coincidence momentum imaging. For this purpose, we determine the angle between the momentum vectors of two Cl+ fragments detected in coincidence with a C2H2+ fragment. The corresponding yield as a function of the cosine of the angle for both and isomers is shown in Fig. 8. For the isomers, the distribution peaks at , while it peaks at for the isomers. Note that the latter value is quite different from our finding in -C2H2Br2 Ablikim et al. (2016), where the yield for the isomers peaked at . This shift to a smaller angle is also reproduced by numerical Coulomb explosion simulation, which assume purely Coulombic repulsion between point charges located at the center of mass of each fragment and instantaneous creation of the charges followed by explosion from the equilibrium geometry of the neutral molecule, as explained in Refs. Ablikim et al. (2016, 2017). The simulations yield value of for the isomer of C2H2Cl2 as opposed to a value of for -C2H2Br2. Note that the simulation overestimates the angle, similar to our findings in the case of C2H2Br2 Ablikim et al. (2016).
Nevertheless, with the help of these simulations, we can determine that a significant contribution to the difference in the momentum correlations for the isomers of C2H2Cl2 and C2H2Br2 is the mass difference between Cl and Br, which changes the energy and momentum sharing with the C2H2+ fragment, while the slightly different equilibrium geometry of C2H2Cl2 and C2H2Br2 affects the simulated angle between the fragments only very little.
Another common way of investigating both the kinematics and the dynamics of a three-body fragmentation process, is by representing the fragment ion momentum correlations in a Newton plot, as shown in Fig. 9. Here, the momenta of two of the fragments (C2H and Cl*+) are plotted with respect to the momentum of the third fragment (Cl+). For the isomer, the Newton plot in Fig. 9(b) shows that the two Cl+* fragments are emitted close to back to back, while the C2H fragment remains almost at rest, as one would expect from an instantaneous fragmentation in the undistorted equilibrium geometry, where the center of mass of the C2H2 moiety is exactly in the middle between the two Cl atoms. The isomer, shown in Fig. 9(a), has a distinctly different fragmentation pattern with the Cl*+* fragments being emitted at a much steeper angle with respect to each other and the C2H fragment also carrying a significant momentum. In both cases, the shape of the features in the Newton plot suggests that the two bonds that are broken in this particular fragmentation process are broken in a concerted manner, since a sequential breakup with a time interval close to (or larger than) the rotational period of the intermediate would lead to a pronounced circular shape of the features in the Newton plot Ablikim et al. (2017), which is clearly absent in the present case.
The difference in the Coulomb explosion kinematics for - and -C2H2Cl2 is also reflected in the kinetic energy distributions and the KER, which are shown in Fig. 10 for the same C2H+Cl++Cl+ coincidence channel as discussed above. For the isomer, the C2H fragment has significantly higher kinetic energy than each of the Cl+ fragments, while for the isomer, the kinetic energy distribution of the C2H fragment has a maximum at 0 eV and the Cl+ fragments are more energetic than for the isomer. Similar to the case of C2H2Br2 Ablikim et al. (2016), this can be explained by an “obstructed instantaneous explosion” Eland (1987); Hsieh and Eland (1997) in the isomer, where the C2H-fragment is trapped between the two Cl+ fragments. This simple picture is corroborated by the results of our Coulomb explosion model calculations shown as dashed lines in Fig. 10, which reproduce the difference between the fragment kinetic energy distributions in the two isomers.
IV CONCLUSIONS
We have presented the design and performance of a double-sided VMI spectrometer optimized for the coincident detection of high-energy photo- or Auger electrons with energetic ions resulting from the Coulomb explosion of inner-shell ionized molecules. With typical lens voltages up to +/- 6 keV, the spectrometer is capable of time- and position-resolved detection of photo- and Auger electrons with kinetic energies up to 300 eV and of photoions with up to 25 eV per unit charge. The kinetic energy acceptance can be increased if higher lens voltages are used. For the typical experimental conditions investigated here, we find a kinetic energy resolution of for electrons with 100 eV kinetic energy and for ions with 10 eV kinetic energy. We have illustrated the coincidence capabilities and, in particular, the possibility to record ion coincidence-channel-resolved Auger electron spectra on the example of an Auger electron-ion-ion-ion coincidence experiment performed on cis- and trans-1,2-dichloroethene at the Advanced Light Source synchrotron radiation storage ring facility.
By combining some of the advantages of traditional VMIs with those of COLTRIMS-type coincidence momentum imaging spectrometers, our double-sided coincidence VMI design is also well suited for high-repetition-rate XUV high harmonic generation sources and for upcoming high-repetition-rate free-electron lasers, where coincident detection of high-energy electron with fragment ions will be an important prerequisit for many gas-phase experiments in atomic and molecular physics and ultrafast photochemistry Young et al. (2018).
Acknowledgements.
This work is supported by the Chemical Sciences, Geosciences, and Biosciences Division, Office of Basic Energy Sciences, Office of Science, U.S. Department of Energy, Grant No. DE-FG02-86ER13491 (Kansas State group) and DE-SC0012376 (UConn group). The research used resources of the Advanced Light Source, which is a DOE Office of Science User Facility under contract no. DE-AC02-05CH11231. We thank the staff of the Advanced Light Source for their hospitality and their support during the beamtimes.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1Chandler and Houston (1987) D. W. Chandler and P. L. Houston, The Journal of Chemical Physics 87 , 1445 (1987) . · doi ↗
- 2Eppink and Parker (1997) A. T. J. B. Eppink and D. H. Parker, Review of Scientific Instruments 68 , 3477 (1997) . · doi ↗
- 3Chandler, Houston, and Parker (2017) D. W. Chandler, P. L. Houston, and D. H. Parker, The Journal of Chemical Physics 147 , 013601 (2017) . · doi ↗
- 4Suits and Continetti (2001) A. G. Suits and R. E. Continetti, Imaging in Chemical Dynamics , ACS Symposium Volume 770 (American Chemical Society, 2001).
- 5Reid (2012) K. L. Reid, Molecular Physics 110 , 131 (2012).
- 6Dörner et al. (2000) R. Dörner, V. Mergel, O. Jagutzki, L. Spielberger, J. Ullrich, R. Moshammer, and H. Schmidt-Böcking, Physics Reports 330 , 95 (2000).
- 7Ullrich et al. (2003) J. Ullrich, R. Moshammer, A. Dorn, R. Dörner, L. P. H. Schmidt, and H. Schmidt-Böcking, Reports on Progress in Physics 66 , 1463 (2003) .
- 8Ullrich et al. (1997) J. Ullrich, R. Moshammer, R. Dörner, O. Jagutzki, V. Mergel, H. Schmidt-Böcking, and L. Spielberger, Journal of Physics B: Atomic, Molecular and Optical Physics 30 , 2917 (1997) . · doi ↗