Analysis of ring laser gyroscopes including laser dynamics
Angela D.V. Di Virgilio, Nicol\`o Beverini, Giorgio Carelli, Donatella, Ciampini, Francesco Fuso, and Enrico Maccioni

TL;DR
This paper presents a novel data analysis technique for ring laser gyroscopes that effectively mitigates laser dynamics non-linearities, enhancing sensitivity and accuracy for both large and small scale inertial sensors.
Contribution
A new analysis method is introduced to study and remove laser dynamics non-linearities, improving the performance of ring laser gyroscopes.
Findings
The technique reduces back-scatter effects, improving measurement accuracy.
Small scale gyroscopes with nanoradian per second sensitivity are feasible.
Performance improvements are demonstrated on prototypes GP2 and GINGERINO.
Abstract
Inertial sensors stimulate very large interest, not only for their application but also for fundamental physics tests. Ring laser gyros, which measure angular rotation rate, are certainly among the most sensitive inertial sensors, with excellent dynamic range and bandwidth. Large area ring laser gyros are routinely able to measure fractions of prad/s, with high duty cycle and bandwidth, providing fast, direct and local measurement of relevant geodetic and geophysical signals. Improvements of a factor would open the windows for general relativity tests, as the GINGER project, an Earth based experiment aiming at the Lense-Thirring test at level. However, it is well known that the dynamics of the laser induces non-linearities, and those effects are more evident in small scale instruments. Sensitivity and accuracy improvements are always worthwhile, and in general there is…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Analysis of ring laser gyroscopes including laser dynamics
Angela D.V. Di Virgilio
INFN Sez. di Pisa,
Polo Fibonacci, Largo B Pontecorvo 3, I-56127 Pisa, Italy
&Nicolò Beverini, Giorgio Carelli, Donatella Ciampini, Francesco Fuso, and Enrico Maccioni
Università di Pisa,
Dipartimento di Fisica "E. Fermi", Largo B Pontecorvo 3, I-56127 Pisa, Italy
Abstract
Inertial sensors stimulate very large interest, not only for their application but also for fundamental physics tests. Ring laser gyros, which measure angular rotation rate, are certainly among the most sensitive inertial sensors, with excellent dynamic range and bandwidth. Large area ring laser gyros are routinely able to measure fractions of prad/s, with high duty cycle and bandwidth, providing fast, direct and local measurement of relevant geodetic and geophysical signals. Improvements of a factor would open the windows for general relativity tests, as the GINGER project, an Earth based experiment aiming at the Lense-Thirring test at level. However, it is well known that the dynamics of the laser induces non-linearities, and those effects are more evident in small scale instruments. Sensitivity and accuracy improvements are always worthwhile, and in general there is demand for high sensitivity environmental study and development of inertial platforms, where small scale transportable instruments should be used. We discuss a novel technique to analyse the data, aiming at studying and removing those non-linearity. The analysis is applied to the two ring laser prototypes GP2 and GINGERINO, and angular rotation rate evaluated with the new and standard methods are compared. The improvement is evident, it shows that the back-scatter problem of the ring laser gyros is negligible with a proper analysis of the data, improving the performances of large scale ring laser gyros, but also indicating that small scale instruments with sensitivity of nrad/s are feasible.
1 Introduction
Ring laser gyroscopes (RLGs) are inertial sensors based on the Sagnac effect [1, 2, 3]. They are largely utilised for inertial navigation, and applications in geodesy, geophysics and even for General Relativity tests are foreseen [4]. Since 2011 we are studying the feasibility of the Lense Thirring test at the level of with an array of large frame RLGs [5, 6, 7]. For that purpose it is necessary to push the relative accuracy of the Earth rotation rate measurement in the range from part in up to part in . RLG consists of a laser with a cavity comprising of three or four mirrors, depending if the cavity is triangular or square, rigidly attached to a frame; large frame RLGs are utilised to measure the Earth rotation rate, being attached to the Earth crust. Because of the Sagnac effect, the two counter-propagating cavity modes have slightly different frequency, and the beat note of the two beams is proportional to the angular rotation rate felt by the ring cavity. Large frame RLGs are the most sensitive instruments for inertial angular rotation measurements. The Sagnac frequency of a RLG is in fact proportional to the angular rotation rate which affects the apparatus:
[TABLE]
where is the area of the ring cavity, is its perimeter, the wavelength of the light, and is angle between the area versor of the ring and the orientation of . For RLGs horizontally aligned (area versor vertical) is the colatitude angle, while for RLGs aligned at the maximum Sagnac signal . Eq. 1 connects with the scale factor , which depends on the geometry, and , quantities than can be measured. Further to sensitivity, other advantages of such instruments rely on the broad bandwidth, which can span from kHz down to DC, and the very large dynamical range. In fact the same device can record microseismic events and high magnitude nearby earthquakes [8], owing to the fact that the signal is based on the measurement of the beat note. It has been proven that large size RLGs with state of the art mirrors can reach the relative precision of parts in in one day integration time, for the Earth rotation rate measurement[1]. If shot noise limited, the sensitivity scales with the second power of the size of the ring cavity.
The main limitation of RLG performances is given by the coupling between the two counter propagating laser modes. On each cavity mirror a little fraction of the two traveling waves is backscattered in the opposite direction. As a result of the interference between the reflected waves from each mirror, we have effective backscattering amplitude reflectivity and of the beam over the beam and of the beam over the beam , respectively. We outline that the interference of the back reflected waves, and consequently the values of are very sensitive to any perturbation of the optical cavity geometry. This coupling produces for small rotational rate a pulling of the Sagnac frequency from the value given by eq. 1, and eventually a locking of the two laser frequencies when the value become lower than [9]. As a rough estimation, it is possible to evaluate for a square cavity , where is the velocity of light, the diameter of the beam and the total scattered fractional amplitude at each reflection.
In order to ensure the functionality of small scale RLGs[10], mechanical dithering is usually implied to increase the bias between the two modes and avoid locking. Large frame rings utilise the Earth rotation rate as bias. Any improvement in the accurate evaluation of the backscatter noise, and in general of the systematics induced by the non linear dynamics of the lasing process, is advantageous for increasing the performance of both large and small frame RLGs. Presently there is large interest in this kind of device, large scale apparatus should further improve their sensitivity and accuracy for geodetic and fundamental physics application, and small scale and transportable devices at nrad/s sensitivity are required to improve the low frequency response of gravitational wave antennas, for the development of inertial platforms, and for seismology [11, 12, 13, 14, 15].
The problem of the reconstruction of signals is a general one, and sophisticated filters can be developed to this aim. In the past we have addressed this problem utilizing Kalman filters, with whom we have obtained good results, but which were rather time consuming. At present, we have the necessity to analyze a very large set of data and to set up mathematical tools for the analysis with the aim not only to evaluate the sensitivity, but also to precisely identify specific issues in the setup which are limiting the sensitivity.
This paper presents a mathematical approach to measure the Sagnac frequency taking into account the laser dynamics. This issue has been addressed several times [9, 16, 17, 18, 19], but no general solution exists. Analytical solutions can be derived in the case the backscattered light is equal in the two counter propagating modes, or the ratio between the intensities of the two modes is constant, conditions which are not fulfilled in the actual generation of RLGs [17]. The discussion is composed of two main parts. The first one, after a short description of the RLG and the standard analysis approach, describes the general RLG dynamics and reconstructs the Sagnac frequency taking into account the laser dynamics in the general case through a single analytical formula containing the laser coefficients (Lamb coefficients), which can be separately evaluated based on experimental measurements. This formula can be further divided as linear sum of six contributions. One of the contributions, called , is the dominant one, the others being small corrections. In the second part of the paper the implementation of is discussed and applied to data of the RLG prototypes GP2 and GINGERINO. The other additional terms are not considered in the implementation, they will be subject of future work based on the data of GINGERINO. The appendix reports a short discussion about noise, and practical methods to identify portions of data which have to be discarded.
2 Typical RLG and standard analysis method
Fig. 1 shows the typical lay-out of a square cavity RLG. The four mirrors are placed at the corner of the square ring, they are contained inside small vacuum chambers, which are connected by pipes.
The whole setup is vacuum tight and filled with Helium and an isotopic mixture of 20Ne and 22Ne. In one of the side the laser discharge is located to generate the plasma required for laser operation (top side). Piezoelectric translators are utilised to translate the mirrors, allowing a control of the RLG perimeter length. The Sagnac beat note signal is observed at one corner (bottom-left) by superimposing the two output beams on a photodiode. Other two photodiodes monitor at one of the output corners (top-left) the laser output power of the two beams (, ). We will indicate them in the following as the mono-beam signals. Another photodiode (discharge monitor) records the fluorescence from the discharge, filtered around nm by a narrow width interferometer filter, which provides a rough indication of the density of excited atoms. In the usual operations, the plasma discharge is electronically controlled in order to keep constant the output power of one of the two mono-beam signal. All these signals are acquired by an ADC with a frequency rate of a few kHz, suitable to well reconstruct them from DC up to the Sagnac frequency.
In the standard analysis, the Sagnac angular frequency is assumed to be equal to the instantaneous frequency , reconstructed from the interferogram by means of the Hilbert transform or of the standard AR2 recursive algorithms based on autocorrelation. The backscatter noise is usually subtracted by fitting the quantity , where is the backscatter phase and and are the amplitudes of DC and spectral components of mono-beams 1 and 2 , respectively [16].
3 Ring Laser dynamics, approximations and the stationary solution
The present analysis is dedicated to the evaluation of the Sagnac signal taking into account the dynamics of the ring laser. The most general description of the RLG is based on the model developed by Aronowitz following the more general Lamb theory [20, 21]. The general equations are
[TABLE]
where , are the intra-cavity laser intensities expressed in the dimensionless "Lamb units"; and are the instantaneous phase difference and its time derivative; index 1 and 2 refers to the clockwise and counter-clockwise laser beam respectively. It is important to remind that all terms of eq. 2,3, and 4 are time depending. Here , , , , are the Lamb parameters. In particular, , are the threshold gain minus losses, is the self saturation, , describe cross-(mutual-)saturation. The Lamb theory involves a large number of parameters, however, the special mixture of two isotopes of Neon and the working point close to the laser threshold allow adoption of a simplified model [17, 18, 22]. In our present analysis, we assume , and . This assumption is justified by the fact that our RLGs operate close to threshold in mono-mode regime (for operation near multi-mode regime, a further approximation is feasible). In the following will be neglected, owing to the mono-mode operation.
Without loss of generality we can define , which is usually referred to as null shift; it is generally accepted that is a small quantity to be neglected [9, 17, 20]. In the present analysis it will not be neglected: will be considered a perturbation of , defining a new variable ( being the frequency effectively measured by the interferogram, called also ).
Assuming that RLG is at the steady state [17], the solutions are the following:
[TABLE]
The validity of the above solutions has been previously tested with a Monte Carlo simulation and with the experimental data of the RLG G-Pisa, which was a m perimeter RLG [18, 23]. Here the validity of previous results is taken for granted and is analytically expressed.
Assuming that the parameters are constant in the time interval between and , we have . From the above relation it is straightforward to deduce that, at the first order in :
[TABLE]
In the above equation is the Sagnac angular frequency, the quantity we are looking for, and we have conveniently defined :
[TABLE]
contains oscillatory terms at the Sagnac frequency . Considering that is almost constant, for frequency much below the Sagnac frequency, it is possible to look for approximated solutions. Eq. 6 can be written as:
[TABLE]
where we dropped the time dependence in . The occurrence of the oscillations of at the Sagnac frequency makes the evaluation of non trivial. The average value of is very small for frequencies much below Sagnac frequency, since the average value of sinus and co-sinus oscillating at the Sagnac frequency goes to zero for frequency much below . can be found with eq.8, provided that , , , , and the average value of are available, which is in principle feasible utilising the mono-beam signals and the measured losses of the cavity and employing numerical recursive methods to evaluate .
In the following eq. 8 will be decomposed in several pieces, which can be separately evaluated. In any case, when and , eq. 8 can be expanded in and at first and second order, obtaining:
[TABLE]
Eq. 9 is composed of terms, which can be independently evaluated, and analysed. Careful evaluation is necessary for , , since the determination of the parameters , , , , and , which are function of the beam area , the output power, the mirrors transmission and the total losses , is required. The mathematical relationships to evaluate those terms can be found in previous papers [16, 17, 18].
The reconstruction of those terms will be addressed in future work, and applied to the analysis of the data of GINGERINO.
In the following the implementation of the first term will be specialised for data acquired with large frame RLGs and compared with the standard analysis method. Backscatter noise is accounted for, and it has been checked that the standard method to subtract the backscatter noise can be derived from eq. 10 assuming and expanding at first order.
3.1 Application to the actual data
The analysis described in the following will take into account data streams at normal operation and far from transients of the laser as mode jumps and split modes. Appendix A describes methods to identify and eliminate those portions of data. As already said eq. 9 is valid for ; referring to our smaller prototype G-Pisa (perimeter m), and utilising published parameters [17], we obtain rad m/s, to be compared with rad m/s: consequently eq. 9 is valid. We underline that quoted values are conservative ones, since depends on the mirror quality and the size of the ring. The prototype G-Pisa was smaller than GP2 and GINGERINO and equipped with less performing mirrors. Determining requires in turn to evaluate and . Following previous works [17, 18], it is possible to link such quantities with available measured data:
[TABLE]
with all symbols already defined. Similarly, the relative phase is found comparing the mono-beams signal at the Sagnac frequency. All above quantities are commonly used in the standard analysis [16, 22, 24]. Substituting and simplifying, it is straightforward to show that:
[TABLE]
The term () has been added in order to take into account inaccuracies on the mono-beams signals. It is important to remark that the quantities , , and and refer to the laser power inside the optical cavity, while measured ones are obtained utilising the power transmitted outside the cavity. Since Eq. 13 exploits the ratios, in principle it is not affected by the measurement scheme, and the voltage output of the photodiodes can be used.
However, it is necessary to consider the presence of noise in the mono-beams signals, which can be due to the inherent noise of the photodiodes or by the discharge fluorescence, which cannot be completely removed. The related noise affects the evaluation of done with eq. 13. Therefore, in order to have the possibility to correct it with common statistical methods, the term has been added. Expanding at first order in , we obtain:
[TABLE]
It is straightforward to evaluate the term , while the corrective one, , must be evaluated by fitting the parameter . Remarkably, the above relation does not contain any Lamb parameter of the laser and can therefore be determined without knowledge of such parameters.
4 Reconstruction of s0 for GP2 and GINGERINO
Data acquired by our prototypes GINGERINO and GP2 are utilised. GP2 is an apparatus with comparatively low quality mirrors and located in a noisy environment [25, 26], while GINGERINO is located in a very quiet place [24, 27], and is presently equipped with state of the art mirrors. We remark that GINGERINO is free running, the geometry is not controlled and long time operation and high duty cycle () are possible since it is located in the underground Gran Sasso laboratory, which exhibits high natural thermal stability (typically oC in one day).
4.1 Comparison of standard and new analysis
Fig. 2 shows the comparison between the Sagnac frequency from GINGERINO data reconstructed with the standard method (referred to as ) and the one presented here. It is interesting to observe that the average value of the frequency is higher, this is what we expect when the noise is dominated by backscatter. In such conditions, frequency is shifted upward (’pull’) [9]. The average values are different for the two analysis methods; as far as GINGERINO is concerned, the difference is quite small, for example the analysis of days in November gives a relative difference of , with evaluated by the method presented here a bit larger than . Since the absolute orientation of the RLG is unknown, in both cases the measured Sagnac frequency is compatible with the expected one assuming an inclination of mrad northwards with respect to the horizontal plane.
Large frame RLGs are instruments dedicated to the study of phenomena with typical frequency below Hz; we have checked that the two methods are equivalent in the high frequency band of interest. Fig. 3, showing the power spectral density (PSD) as a function of frequency, demonstrates that, for GINGERINO and above mHz, the difference between the two methods is less than nrad/s in 1 second measurement. This comparison shows that the new analysis is not introducing extra noise above mHz at this level of sensitivity. It has been also checked that the old method, which estimates and subtracts the backscattering effect through a linear fitting procedure, provides results distributed with width similar to , and, as already said, slightly different mean value. We outline that systematics of the laser dynamics include non linear terms, which in principle cannot be eliminated with linear methods. Then, the standard method, being linear, cannot guarantee a full correction.
The systematics of RLG depends on the size and the mirror quality, large frame RLGs are usually closer to behave in an ideal manner. For reduced size RLG and when the mirrors are not top quality, deviations from the ideal case are more relevant. This is the case of our prototype GP2. Fig. 4 shows the histogram of the Sagnac frequency data of GP2 analysed with the two methods. The standard analysis leads to a broader distribution and the mean value is compatible with a mean rotation frequency Hz higher than expected. GP2 is oriented at the maximum signal, so its response should be close to (and never higher than) the Earth rotation rate. The new method gives an averaged rotation rate rad/s, in agreement with the Earth rotation rate rad/s. With the new analysis the average rotation rate is evaluated with a relative systematic error of part in , while with part in with the standard analysis: a factor improvement in accuracy has been achieved. The present result is very similar to the one obtained in previous analysis based on the Kalman filter in term of accuracy and sensitivity. In both cases the best sensitivity was of the order of a few nrad/s with tens of seconds integration time [18].
It is puzzling to note that with the standard analysis method GP2 is showing higher than expected Sagnac frequency. A possible explanation is given within our mathematical approach to measure the Sagnac frequency. GP2 has been designed to test the geometry control developed for GINGER and based on diagonal length measurements; data with and without geometry control have been compared and it has been checked that, with adequate analysis, sensitivities are comparable [26]. Fig. 5 compares the fringe contrast (TOP) and the relative phase (BOTTOM) for GP2 data taken during the geometry control test. This was achieved by keeping constant the length of the two diagonals within 80 nm [26]. Fig. 5 shows that mode jumps occur in order to keep close to , this in support of the fact that GP2 has quite large backscatter light, and stable operation is favourite when is such that the two modes have an extra shift of Hz, also called "dissipative coupling regime" [9]. It is not perfectly clear from the theory why this regime takes place, but it is a matter of fact that the coupling between the two modes decreases increasing the bias frequency, accordingly stable operation is favourite. Another consequence is that mode jumps occur also to keep the relative phase close to a certain range, not only to compensate changes of the perimeter. It has been checked that GINGERINO exhibits all values of .
5 Conclusion
Systematics induced by the non linear dynamics of the laser, mainly due to backscatter light, induces non linear terms in the output of high sensitivity RLGs, severely limiting the development of RLGs with sensitivity of the order of nrad/s level, which in principle should have a large range of applications. An analytical method, suitable to reconstruct the Sagnac frequency taking into account the laser dynamics, has been developed in the general case in which the two backscattered beams are not equal, and the ratio between the power of the two counter-propagating modes is not constant. The application of this formula requires the knowledge of the fractions of backscattered waves, and the laser parameters , , and , all quantities which can be evaluated. In the present theory the term is not considered, this term takes into account the multimode operation, and can be neglected in the description of high sensitivity RLGs which operates mono-mode close to threshold.
Expanding in series at first and second order it is possible to divide the general formula as the sum of six terms which can be separately evaluated. The analytical expansion for the whole set of terms is reported.
The term called , which is the dominant one and does not contain any laser parameter, is evaluated in details and expressed as a function of the available measurements; this term has been evaluated for the two RLG prototypes GINGERINO and GP2, and compared with the standard analysis method. The advantage of the new approach is evident: not only the width of the distribution is reduced, but the reconstructed Sagnac frequency is more accurate and in better agreement with the expected value. In short, eliminates the so called backscatter noise, which is the dominant systematics especially for small and medium size RLG.
The GP2 prototype has more backscatter light, because it is smaller and has lower quality mirrors with respect to GINGERINO. In this case the standard method evaluates the Earth rotation rate with a relative systematic error of part in , while in the new way part in is obtained, a factor improvement in accuracy, with a sensitivity in the range of nrad/s with tens of seconds integration time. This work opens up the window for the development of high sensitivity transportable RLGs, for applications in seismology, as environmental monitors to improve the low frequency performance of the test mass suspension of the gravitational wave antennas, and for the development of inertial platforms in general. Further efforts will be devoted to analyse the data of GINGERINO using the full set of terms.
Appendix A Discussion about the noise
Since we deal with high sensitivity measurements, it is important to estimate the noise injected in the evaluation of . and () are utilised to evaluate , and their noise will contribute to the total noise budget. In general all measurements of these quantities are limited by shot noise of the power collected by the photodiode , and the total noise is the incoherent sum of the photodiode noise. The contribution of each term, and gives:
[TABLE]
( and are obtained changing 1 with 2 in a symmetric way).
Usually for top quality mirrors losses are minimised, but there are no requirements for the transmission. In order to minimise the contribution of the mono-beams to the total noise the optimal choice would be to have transmission of the same order of the losses, at least for one of the mirrors (one output only is enough since in order to evaluate it is necessary to observe the two mono-beams transmitted by the same mirror). Care is also necessary in order to avoid small spurious reflections from one of the windows of the vacuum chamber, and narrow band filters are necessary in order to reduce the spurious signal from the discharge fluorescence. In any case, especially the measurement of the two terms could be a real limitation for the very low frequency measurements, since they are DC quantities affected by the well known noise of any electronic device.
Appendix B The fringe contrast: a suitable tool to remove bad portions of data
In large frame RLG attached to the Earth crust the Sagnac frequency is usually above Hz, and it is determined by the Earth rotation rate, which is almost constant in time. The relative phase is slowly changing, since the cavity is rigid. In general, unless the geometry is electronically controlled, it happens that the RLG changes its operational points; accordingly the wavelength changes separately for both modes, or for one only, and mode jumps or split mode operations occur. Split mode operation occurs from time to time; in principle the split mode regime provides good measurement of , but in this case (FSR, Free Spectral Range), and data acquisition at high rate and accurate knowledge of the perimeter are necessary. In the present analysis data affected by split mode operation have been disregarded. Mode jumps are very fast transients, affecting only few seconds of data acquisition. During these discontinuities the RLG is not at the stationary condition, so portions of data have to be discarded. The observation of the fringe contrast provides a very efficient tool to identify and eliminate those imperfections. Fig. 6 shows corresponding split mode and mode jumps. Sometime some instabilities in the operation are visible before the mode jump takes place. Fig. 7 shows the typical behaviour of the mode jump.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1[1] Karl Ulrich Schreiber and Jon-Paul R. Wells. Invited review article: Large ring lasers for rotation sensing. Review of Scientific Instruments , 84(4):041101, 2013.
- 2[2] Dossier sommaire. Comptes Rendus Physique , 15(10):iii – iv, 2014. The Sagnac effect: 100 years later.
- 3[3] Angela Di Virgilio, Maria Allegrini, Alessandro Beghi, Jacopo Belfi, Nicolo Beverini, Filippo Bosi, Bachir Bouhadef, Massimo Calamai, Giorgio Carelli, Davide Cuccato, Enrico Maccioni, Antonello Ortolan, Giuseppe Passeggio, Alberto Porzio, Matteo Luca Ruggiero, Rosa Santagata, and Angelo Tartaglia. A ring lasers array for fundamental physics. Comptes Rendus Physique , 15(10):866 – 874, 2014. The Sagnac effect: 100 years later / L’effet Sagnac : 100 ans après.
- 4[4] Monika Tercjak and Aleksander Brzeziński. On the influence of known diurnal and subdiurnal signals in polar motion and ut 1 on ring laser gyroscope observations. Pure and Applied Geophysics , 174(7):2719–2731, Jul 2017.
- 5[5] F. Bosi, G. Cella, A. Di Virgilio, A. Ortolan, A. Porzio, S. Solimeno, M. Cerdonio, J. P. Zendri, M. Allegrini, J. Belfi, N. Beverini, B. Bouhadef, G. Carelli, I. Ferrante, E. Maccioni, R. Passaquieti, F. Stefani, M. L. Ruggiero, A. Tartaglia, K. U. Schreiber, A. Gebauer, and J-P. R. Wells. Measuring gravitomagnetic effects by a multi-ring-laser gyroscope. Phys. Rev. D , 84:122002, Dec 2011.
- 6[6] Angelo Tartaglia, Angela Di Virgilio, Jacopo Belfi, Nicolò Beverini, and Matteo Luca Ruggiero. Testing general relativity by means of ring lasers. The European Physical Journal Plus , 132(2):73, Feb 2017.
- 7[7] Angela D. V. Di Virgilio, Jacopo Belfi, Wei-Tou Ni, Nicolo Beverini, Giorgio Carelli, Enrico Maccioni, and Alberto Porzio. Ginger: A feasibility study. The European Physical Journal Plus , 132(4):157, Apr 2017.
- 8[8] A Simonelli, H Igel, J Wassermann, J Belfi, A Di Virgilio, N Beverini, G De Luca, and G Saccorotti. Rotational motions from the 2016, central italy seismic sequence, as observed by an underground ring laser gyroscope. Geophysical Journal International , 214(1):705–715, 2018.