# Characterization of the Si(Li) detector for Monte Carlo calculations of   beta spectra

**Authors:** Pavel Novotny, Pavel Dryak, Jaroslav Solc, Petr Kovar, Zdenek, Vykydal

arXiv: 1904.01294 · 2019-04-03

## TL;DR

This paper develops a detailed Monte Carlo model of a Si(Li) detector for accurate beta spectrum calculations, validated through efficiency measurements and comparison with experimental spectra.

## Contribution

The paper presents a novel, precise Monte Carlo model of a Si(Li) detector validated against experimental data for beta and photon spectra.

## Key findings

- Model accurately predicts detector efficiency from 5 to 136 keV.
- Good agreement between calculated and measured beta spectra for Pm-147.
- Validated model enhances beta spectrum analysis accuracy.

## Abstract

A precise model of a Si(Li) detector ORTEC model SLP-06165P-OPT-0.5 was created for beta spectra calculations using the Monte Carlo (MC) code MCNPX. Detector parameters were determined from X-ray radiograms obtained with a film and a Timepix detector. The MC model of the detector was validated by comparison of calculated and experimental full-energy peak efficiencies in the energy range from 5 to 136 keV using a range of point-like photon emitting radionuclide standards. A comparison of measured and calculated beta spectrum of a radionuclide Pm-147 is presented.

---
Source: https://tomesphere.com/paper/1904.01294