A fast field-cycling MRI relaxometer for physical contrasts design and pre-clinical studies in small animals
Javier A. Romero, Gonzalo G. Rodriguez, Esteban Anoardo

TL;DR
This paper introduces a rapid field-cycling MRI relaxometer capable of producing images and contrast variations for small animal studies, combining specific hardware design with novel contrast techniques.
Contribution
A novel fast field-cycling relaxometer with integrated imaging capabilities and new contrast methods for pre-clinical small animal research.
Findings
Produced images of reasonable quality at 5 MHz
Demonstrated T1-dispersion weighted imaging and temperature mapping
Introduced positive and negative contrast techniques
Abstract
We present a fast field-cycling NMR relaxometer with added magnetic resonance imaging capabilities. The instrument operates at a maximum proton Larmor frequency of 5 MHz for a sample volume of 35 mL. The magnetic field homogeneity across the sample is 1400 ppm. The main field is generated with a notch-coil electromagnet of own design, fed with a current whose stability is 220 ppm. We show that images of reasonable quality can still be produced under such conditions. The machine is being designed for concept testing of the involved instrumentation and specific contrast agents aimed for field-cycling magnetic resonance imaging applications. The general performance of the prototype was tested through localized relaxometry experiments, T1-dispersion weighted images, temperature maps and T1-weighted images at different magnetic field intensities. We introduce the concept of positive and…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.