Photon flux determination of a liquid-metal jet x-ray source by means of photon scattering
Malte Wansleben, Claudia Zech, Cornelia Streeck, Jan Weser, Christoph, Genzel, Burkhard Beckhoff, Roland Mainz

TL;DR
This paper presents a novel experimental method to accurately determine the photon flux of a liquid-metal jet X-ray source using photon scattering, enabling precise flux measurements crucial for quantitative X-ray analysis.
Contribution
The study introduces an indirect photon flux measurement technique based on photon scattering, reducing uncertainties and avoiding detector damage in high-flux X-ray sources.
Findings
Photon flux for Ga Kα line: 6.0(5)×10^12 s⁻¹sr⁻¹
Photon flux for In Kα line: 3.8(4)×10^11 s⁻¹sr⁻¹
Method provides detailed uncertainty budget and reference to polarized synchrotron radiation.
Abstract
Liquid-metal jet X-ray sources promise to deliver high photon fluxes, which are unprecedented for laboratory based X-ray sources, because the regenerating liquid-metal anode is less sensitive to damage caused by an increased electron beam power density. For some quantitative X-ray analysis techniques, knowledge of the absolute photon flux is needed. However, a precise experimental determination of the photon flux of high-performance X-ray sources is challenging, because a direct measurement results in significant dead time losses in the detector or could even harm the detector. Indirect determinations rely on data base values of attenuation or scattering cross sections leading to large uncertainties. In this study we present an experimental determination of the photon flux of a liquid-metal jet X-ray source by means of elastic and inelastic photon scattering. Our approach allows for…
| line | / keV | power / W | / s-1sr |
|---|---|---|---|
| Ga K | 9.25111according to ref.Deslattes et al. (2003) | 200.0222nominal voltage 70 kV, current 2.857 mA | |
| 457.1333nominal voltage 160 kV, current 2.857 mA. We note that these settings are above the specifications of the manufacturer and may damage the source. | |||
| In K | 24.21 | 200.0 | |
| 457.1 |
| parameter | rel. uncertainty / % | comment | |
| Ga K | In K | ||
| 3 | 8 | detected count rate LMJ | |
| 2 | 2 | detected count rate BAMline | |
| 1 | 1 | determined flux BAMline | |
| d4-d2 | 0.7 | 0.05 | distance of air absorption |
| 0.4 | 0.03 | density of air | |
| 4 | 0.3 | attenuation coefficient of air | |
| 5 | 5 | solid angle of the LMJ | |
| 8 | 10 | determined flux per solid angle | |
| of the LMJ | |||
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Photon flux determination of a liquid-metal jet x-ray source by means of photon scattering
Malte Wansleben
Claudia Zech
Cornelia Streeck
Jan Weser
Physikalisch-Technische Bundesanstalt (PTB), Abbestraße 2-12, 10587 Berlin, Germany
Christoph Genzel
Helmholtz-Zentrum Berlin für Materialien und Energie GmbH (HZB), Hahn-Meitner-Platz 1, 14109 Berlin, Germany
Burkhard Beckhoff
Physikalisch-Technische Bundesanstalt (PTB), Abbestraße 2-12, 10587 Berlin, Germany
Roland Mainz
Helmholtz-Zentrum Berlin für Materialien und Energie GmbH (HZB), Hahn-Meitner-Platz 1, 14109 Berlin, Germany
Abstract
Liquid-metal jet X-ray sources promise to deliver high photon fluxes, which are unprecedented for laboratory based X-ray sources, because the regenerating liquid-metal anode is less sensitive to damage caused by an increased electron beam power density. For some quantitative X-ray analysis techniques, knowledge of the absolute photon flux is needed. However, a precise experimental determination of the photon flux of high-performance X-ray sources is challenging, because a direct measurement results in significant dead time losses in the detector or could even harm the detector. Indirect determinations rely on data base values of attenuation or scattering cross sections leading to large uncertainties. In this study we present an experimental determination of the photon flux of a liquid-metal jet X-ray source by means of elastic and inelastic photon scattering. Our approach allows for referencing the unpolarized output radiation of the liquid-metal jet X-ray source to polarized synchrotron radiation in a simple setup. Absolute photon fluxes per solid angle are presented with a detailed uncertainty budget for the characteristic emission lines of Ga K and In K for two different acceleration voltages of the X-ray source. For a nominal setting of 200 W (70 kV, 2.857 mA) the determined values for Ga K and In K are 6.0(5) s*-1sr-1* and 3.8(4) s*-1sr-1*, respectively.
††preprint: AIP/123-QED
Electron-impact X-ray sources are compact solutions for laboratory based research and applications. As the brilliance scales with the electron-beam power density at the anode, the physical requirements of the anode material become more challenging since thermal stress and damage at the anode surface degrade the output power. The implementation of a rotating anode overcomes this quandary partly, however, intrinsic thermal limitations are still present. A more recent concept is the use of a liquid-metal jet anode materialHemberg, Otendal, and Hertz (2003). The fluid anode jet is capable of taking a higher thermal load. Additionally, the increased velocities of the jet (up to 100 m/s) compared to a rotating anode provide a faster regeneration and perpetual smooth anode surfaceHemberg, Otendal, and Hertz (2003); Hallstedt et al. (2018). The X-ray flux produced by the liquid-metal jet X-ray source (LMJ) is expected to come close to bending-magnet synchrotron sourcesOtendal et al. (2008), enabling numerous applications such as high-resolution imagingLarsson et al. (2011), X-ray microscopyThüring et al. (2013), X-ray phase-contrast tomography Larsson et al. (2016), confocal micro X-ray fluorescence spectroscopyBauer et al. (2018), or real-time X-ray diffraction in the laboratory.
The knowledge of the incident X-ray flux is needed for quantitative analysesBeckhoff (2008), however, a direct experimental determination of high X-ray fluxes is troublesome because conventional X-ray spectrometers are not capable of dealing with count rates of the above-mentioned magnitude resulting in significant dead time losses or even radiation damage. Hence, there are various indirect experimental or semi-empirical approaches to estimate the output spectrum of X-ray tubes, such as Compton scattering Yaffe, Taylor, and Johns (1976); Matscheko and Ribberfors (1987); Maeda, Matsumoto, and Taniguchi (2005), induced fluorescence emission in targets Moran et al. (1985), combined attenuation and scattering of the primary beam Mainardi and Barrea (1989), combined scattering experiments and Monte Carlo simulation Padilla et al. (2005), or attenuation of the scattered beam Mainardi and Bonzi (2008). All of these approaches rely on data base values describing the respective physical processes involved in the photon matter interaction, such as the attenuation coefficient or scattering cross sections. Hence, experimental results strongly depend on these parameters and their uncertainties. Although the PTB has access to calibrated silicon drift detectors (SDD)Beckhoff et al. (2000), a direct determination with the detector directed towards the LMJ is not feasible. The main reason is the expected high photon flux which would require appropriate absorbers to prevent the detector from exposure to high photon fluxes. Due to the strong energy-dependence of the attenuation coefficients, it is not possible to sufficiently attenuate the high-energy range of the radiation without completely suppressing the low-energy range.
In this work we present an indirect approach that allows to determine the absolute photon flux with low uncertainty based on referencing detected elastically and inelastically scattered photons measured at the LMJ to scattered photons measured at a synchrotron radiation beamline. This method is independent of tabulated scattering cross sections. Only the attenuation coefficient for air is needed. We apply this method to determine the fluxes of the characteristic K emission lines of gallium and indium which are the main constituents of the LMJ anode.
A schematic view of the two experimental setups is shown in Fig. 1. In both setups the same experimental end-station, the so-called SDD scanner, is used to relate the detected count rate of the scattered characteristic emission lines of the LMJ to the scattered monochromatic synchrotron radiation. The SDD scanner allows for the recording of an X-ray spectrum in almost any combination of incident and exit angle. A schematic of the SDD scanner is depicted in Fig. 2. The SDD scanner comprises a sample manipulator and a detector arm and allows for an independent alignment of the surface normal of the sample and the detector front-end with any unit vector of the spherical coordinate system, respectively. While the sample rests in the center of the common spherical coordinate system, the SDD on the detector arm is allowed to move on a spherical surface around this center. The rotational degrees of freedom are fully motorized. The angular resolution is approximately 3*∘*. The SDD-scanner was designed for the experimental determination of scattering cross section and spectroscopic analysis of light elements in the hard X-ray regime.
The anode material (ExAlloy I1) of the used LMJ from Excillium (model D2) contains nominal weight fractions of 68% gallium (Ga), 22% indium (In) and 10% tin (Sn). The dimension of the electron beam was nominally set to 50 m high and 80 m wide resulting in a projected size of 50 m 20 m according to the manufacturer. We note that the projected spot size can be further reduced to 20 m 20 m with the same power settings. The electron impact ionizes the atoms of the jet which results in the isotropic emission of characteristic X-ray fluorescence. An exit window defines the emission direction and separates the vacuum chamber from the ambient condition in the laboratory. Two apertures behind the exit window collimate the emitted X-rays towards the experiment. In order to determine the solid angle of emission defined by the two apertures, an image of the circular excitation beam profile is recorded by means of an X-ray area detector at distance from the source point (Fig. 1(a)). The beam radius at the detector position is determined by considering the pixel size of the detector leading to a solid angle of . Thus, at the position of the scattering target, the diameter of the beam corresponds to 0.74(2) mm.
The scattering experiment with the SDD scanner was repeated with monochromatic synchrotron radiation provided by the 7-T wavelength shifter (WLS) beamline (BAMline)Görner et al. (2001); Riesemeier et al. (2005) at the electron storage ring BESSY II. The BAMline is equipped with a double-crystal monochromator (DCM) and a double multilayer monochromator (DMM) which was operated in series. While the DCM provides high spectral resolving power, the DMM efficiently suppresses higher-order radiation below . Energy calibration of the monochromators is realized with an energy scan across the K-absorption edge of a 12.5-m-thick palladium foil for the excitation energy of 24.21 keV and 10-m-thick copper foil for the excitation energy of 9.25 keV. An experimental hutch enables experiments under ambient conditions. An ionization chamber is installed behind the exit window of the beamline for beam monitoring. A calibrated photodiode at distance from the exit window is used to determine the incident radiant power which can then be converted into photon flux using the experimentally determined responsivityGerlach et al. (2008) after the sample has been removed from the setup (Fig. 1 (b)). A set of apertures within the beamline was used to define a rectangular beam spot of 11 mm2.
A 1.84-mm-thick soda-lime glass sample is used as scattering target. Depending on the incident energy and angle of detection, metal foils may exhibit interfering diffraction peaks as reported in Gerlach et al.Gerlach et al. (2009). Glass is an amorphous material which prevents diffraction peaks. It mainly consists of silicone dioxide. Silicon and oxygen are light elements which do not exhibit any characteristic fluorescence lines in the examined spectral region of the characteristic emission lines of the LMJ. A disadvantage of light elements is the low scattering cross sectionsChantler (1995) requiring a sufficiently thick scattering target.
Equal to any other electron impact based X-ray source, the emitted radiation of the LMJ is unpolarized with characteristic emission lines of the anode material accompanied by continuous Bremsstrahlung. This is in contrast to the BAMline providing highly monochromatized and polarized radiation within the storage ring plane of the synchrotron radiation facility BESSY II. Hence, the detected photons of the polarized excitation source has to be transformed such that it is comparable to the photons detected at the unpolarized LMJ. The definition of the differential scattering cross sections for elastic and inelastic scattering of unpolarized (superscript and polarized radiation (superscirpt ) can be found elsewhereCesareo et al. (1992); Schoonjans et al. (2011). Conversion of the differential scattering cross section from polarized to unpolarized radiation demands an averaging over all possible azimuthal angles , which describes the angle out of the polarization plane of the synchrotron radiation (Fig. 2). This can also be expressed in terms of the differential scattering cross section in the polarization plane (subscript ) and out of the polarization plane (subscript )Cesareo et al. (1992); Roy, Pratt, and Kissel (1993):
[TABLE]
where subscript corresponds to and subscript corresponds to . For all measurements the scattering angle was fixed. Using a relatively thick scatterer, self-absorption is not neglectable. Thus, the surface normal of scattering target is set to and , ensuring the same exit angle for and . The incident angle is in both cases.
Figure 3 (a) displays one of the two SDD spectra () measured at the LMJ. Various fluorescence lines are observed including argon (Ar) of the surrounding air and 3d metals (Ti, Cr, Fe, Ni) of the SDD scanner’s steel construction caused by secondary excitation. The traces of rubidium (Rb), strontium (Sr) and zirconium (Zr) as well as large amounts of silicon (Si) and calcium (Ca) are found in the glass scattering target. The intensity of the Si peak is highly attenuated by the air distance between target and detector. The two highlighted areas mark the elastic and inelastic scattering signals of the characteristic emission lines of the liquid-metal anode. A more detailed view of these spectral regions is given in Fig.3 (b) and (c). The corresponding features of In and Sn are superimposed as indicated by the arrows in Fig.3 (c). The energy of the inelastically scattered photons is calculated according to ComptonCompton (1923). Only the elastic count rate of Sn K and the inelastic count rate of In K can be well separated. Hence, in order to calculate the photon flux of the characteristic In K line of the LMJ, only the inelastic In K is considered, whereas for Ga K line the sum of the inelastic and elastic peak is considered. The respective count rates are obtained by summing over the background subtracted region of interest as indicated by the grey shaded areas in Figs. 3 (b) and (c).
Figures 4 (a) and (b) show the SDD spectra recorded for excitation energies 9.25 keV and 24.21 keV for the two different azimuthal angles at the BAMline. The count rates from both, elastic and inelastic scattering, are obtained from deconvolution of the spectra with a known spectrometer responseScholze and Procop (2001). The count rates associated with elastic and inelastic scattering increase with increasing azimuthal angle as expected due to the linearly polarized synchrotron radiation. The inelastic scattering cross section and the inelastic energy shift increase with increasing energy. Since elastic and inelastic peaks are hardly separated in the case of Ga, the count rate of both peaks are combined in a common region of interest. This is reasonable because the elastic and inelastic peaks are well separated from other fluorescence peaks in the spectra.
For the unpolarized radiation of the LMJ the detected count rate of the scattered photons with energy can be described by the following equation derived from Henke et al.Henke, Gullikson, and Davis (1993)
[TABLE]
is the unpolarized photon flux of energy leaving the LMJ in the solid angle defined by the aperture. expresses the attenuation of the flux from the source window to the scattering target (Fig. 1(a)). is the fraction of the incidence flux that is scattered in the direction of the detector, where comprises the radiation path through the sample, its density, and self-attenuation effects. The factor summarizes all detector related quantities, such as detection efficiency, solid angle of detection, and air absorption between target and detector. The factors and are exactly the same for all measurements at the LMJ and the synchrotron radiation source so they need not to be specified. For elastic scattering .
For the polarized radiation () of the synchrotron radiaition beamline we can express the count rates (scattering in the polarization plane) and (scattering out of the polarization plane) in a similar way:
[TABLE]
is the polarized flux at the window of the beamline. For the calibration measurement by the photodiode (Fig. 1(b)), this flux is also attenuated by air absorption and hence the derived photon flux is
By inserting Eq. (1) and Eqs. (3) into Eq. (2) we obtain the flux of the LMJ
[TABLE]
The count rates , , and are extracted from the spectra in Figs. 3 and 4 as described above. Table 1 presents a summary of the determined fluxes of the Ga K and In K lines of the LMJ for two different power settings. To be able to compare the flux with that of other sources we calculate the brilliance for the LMJ for Ga K and In K with 200 W electron beam power. We obtain an estimated brilliance of for Ga K and for In K, assuming a projected circular source of diameter, a K to K intensity ratio of 2:1, and by scaling the natural line widthKrause and Oliver (1979) of 2.6 eV (Ga K) or 10.6 eV (In K).
Because target and detector specific parameters with potentially large uncertainties cancel out, the uncertainties of the determined fluxes are strongly reduced. Eq. (4) allows for an analytic determination of the photon flux. Thus, the resulting uncertainties are derived according to the variance formula considering the propagation of uncertainties of the input parameters. An overview of the relative uncertainties of all used parameters is given in Table 2. Largest contributions to the overall uncertainties are from the determination of the count rates at the LMJ due to background subtraction and the definition of a region of interest (Fig. 3 (b) and (c)). Especially the overlapping peaks in the spectral region of In K lead to an estimated uncertainty of 8% of the inelastic count rate. Measurement times were chosen in all experiments such that the relative statistical error is below 0.5%.
In summary, we experimentally determined the photon flux per solid angle of two characteristic emission lines of a LMJ by referencing the unpolarized radiation of the LMJ to a well-known polarized synchrotron radiation beamline. High accuracy is achieved since knowledge of any type of X-ray interaction cross section is not needed as compared to other indirect methods. A calibrated SDD regarding its detection efficiency or the solid angle of detection is not needed since these quantities cancel given that the same setup is used at the synchrotron radiation source and the LMJ. The key prerequisites of our method are access to monochromatic X-rays in the desired energy range and a calibrated photo diode. Once a suitable scattering target has been measured and the scattering count rates have been related to the incoming photon flux, the compact setup allows for an easy transfer to numerous laboratory or medical X-ray sources.
Acknowledgements
The collaboration of PTB and HZB leading to this work was supported by the EMPIR project HyMet. The financial support of the EMPIR program is gratefully acknowledged. It is jointly funded by the EMPIR and participating countries within the European Association of National Metrology Institutes (EURAMET) and the European Union. We thank R. Häfner, PTB, for the construction of the experimental end-station and G. Wagener, HZB, for technical support in the LIMAX laboratory.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1Hemberg, Otendal, and Hertz (2003) O. Hemberg, M. Otendal, and H. M. Hertz, “Liquid-metal-jet anode electron-impact x-ray source,” Appl. Phys. Lett. 83 , 1483–1485 (2003) . · doi ↗
- 2Hallstedt et al. (2018) J. Hallstedt, E. Espes, U. Lundstrom, and B. Hansson, “Liquid-metal-jet X-ray technology for nanoelectronics characterization and metrology,” in 2018 29th Annu. SEMI Adv. Semicond. Manuf. Conf. ASMC 2018 (2018) pp. 151–154. · doi ↗
- 3Otendal et al. (2008) M. Otendal, T. Tuohimaa, U. Vogt, and H. M. Hertz, “A 9 ke V electron-impact liquid-gallium-jet x-ray source,” Rev. Sci. Instrum. 79 , 016102 (2008) . · doi ↗
- 4Larsson et al. (2011) D. H. Larsson, P. A. Takman, U. Lundström, A. Burvall, and H. M. Hertz, “A 24 ke V liquid-metal-jet x-ray source for biomedical applications,” Rev. Sci. Instrum. 82 , 123701 (2011) . · doi ↗
- 5Thüring et al. (2013) T. Thüring, T. Zhou, U. Lundström, A. Burvall, S. Rutishauser, C. David, H. M. Hertz, and M. Stampanoni, “X-ray grating interferometry with a liquid-metal jet source,” Appl. Phys. Lett. 103 , 091105 (2013) . · doi ↗
- 6Larsson et al. (2016) D. H. Larsson, W. Vågberg, A. Yaroshenko, A. Ö. Yildirim, and H. M. Hertz, “High-resolution short-exposure small-animal laboratory x-ray phase-contrast tomography,” Sci. Rep. 6 , 1–8 (2016) . · doi ↗
- 7Bauer et al. (2018) L. Bauer, M. Lindqvist, F. Förste, U. Lundström, B. Hansson, M. Thiel, S. Bjeoumikhova, D. Grötzsch, W. Malzer, B. Kanngießer, and I. Mantouvalou, “Confocal micro-X-ray fluorescence spectroscopy with a liquid metal jet source,” J. Anal. At. Spectrom. 33 , 1552–1558 (2018) . · doi ↗
- 8Beckhoff (2008) B. Beckhoff, “Reference-free X-ray spectrometry based on metrology using synchrotron radiation,” J. Anal. At. Spectrom. 23 , 845–853 (2008) . · doi ↗