# Natively Fat-Suppressed 5D Whole-Heart MRI with a Radial Free-Running   Fast-Interrupted Steady-State (FISS) Sequence at 1.5T and 3T

**Authors:** Jessica AM Bastiaansen, Davide Piccini, Lorenzo Di Sopra, Christopher, W Roy, Robert R Edelman, Ioannis Koktzoglou, Jerome Yerly, Matthias Stuber

arXiv: 1903.03498 · 2019-08-20

## TL;DR

This study develops and optimizes a fast interrupted steady-state (FISS) MRI sequence for native fat suppression in 5D whole-heart imaging at 1.5T and 3T, enabling high-quality, free-running cardiac scans within 8 minutes.

## Contribution

The paper introduces an optimized FISS sequence for native fat suppression in 5D whole-heart MRI at 1.5T and 3T, with detailed parameter ranges and validation in volunteers.

## Key findings

- FISS provides effective fat suppression compared to bSSFP.
- Optimal TR and NR parameters depend on field strength and desired suppression bandwidth.
- High-resolution 5D whole-heart MRI can be performed in under 8 minutes with FISS.

## Abstract

Purpose: To implement, optimize and test fast interrupted steady-state (FISS) for natively fat-suppressed free-running 5D whole-heart MRI at 1.5T and 3T. Methods: FISS was implemented for fully self-gated free-running cardiac- and respiratory-motion-resolved radial imaging of the heart at 1.5T and 3T. Numerical simulations and phantom scans were performed to compare fat suppression characteristics and to determine parameter ranges (readouts per FISS module (NR) and repetition time (TR)) for effective fat suppression. Subsequently, free-running FISS data were collected in ten healthy volunteers. All acquisitions were compared with a continuous bSSFP version of the same sequence, and both fat suppression and scan times were analyzed. Results: Simulations demonstrate a variable width and location of suppression bands in FISS that was dependent on TR and NR. For a fat suppression bandwidth of 100Hz and NR below 8, simulations demonstrated that a TR between 2.2ms and 3.0ms is required at 1.5T while a range of 3.0ms to 3.5ms applies at 3T. Fat signal increases with NR. These findings were corroborated in phantom experiments. In volunteers, fat SNR was significantly decreased using FISS compared with bSSPF at both field strengths. After protocol optimization, high-resolution (1.1mm x 1.1mm x 1.1mm) 5D whole-heart free-running FISS can be performed with effective fat suppression in under 8 min at 1.5T and 3T at a modest scan time increase compared to bSSFP.Conclusion: An optimal FISS parameter range was determined enabling natively fat-suppressed 5D whole-heart free-running MRI with a single continuous scan at 1.5T and 3T, demonstrating potential for cardiac imaging and noncontrast angiography.

---