The neutron electric dipole moment experiment at the Spallation Neutron Source
K.K.H. Leung, M. Ahmed, R. Alarcon, A. Aleksandrova, S. Bae{\ss}ler,, L. Barr\'on-Palos, L. Bartoszek, D.H. Beck, M. Behzadipour, J. Bessuille,, M.A. Blatnik, M. Broering, L.J. Broussard, M. Busch, R. Carr, P.-H. Chu, V., Cianciolo, S.M. Clayton, M.D. Cooper, C. Crawford

TL;DR
The paper describes the design and expected sensitivity of the nEDM@SNS experiment, which aims to measure the neutron electric dipole moment with unprecedented precision using innovative cryogenic techniques and systematic error controls.
Contribution
It introduces a novel experimental setup combining polarized ultracold neutrons, $^3$He, and superfluid helium to achieve higher sensitivity and systematic error mitigation in neutron EDM measurements.
Findings
Design of a cryogenic apparatus with two small measurement cells
Enhanced electric field strength and precession time compared to previous methods
Implementation of two measurement modes for result validation
Abstract
Novel experimental techniques are required to make the next big leap in neutron electric dipole moment experimental sensitivity, both in terms of statistics and systematic error control. The nEDM experiment at the Spallation Neutron Source (nEDM@SNS) will implement the scheme of Golub & Lamoreaux [Phys. Rep., 237, 1 (1994)]. The unique properties of combining polarized ultracold neutrons, polarized He, and superfluid He will be exploited to provide a sensitivity to . Our cryogenic apparatus will deploy two small () measurement cells with a high density of ultracold neutrons produced and spin analyzed in situ. The electric field strength, precession time, magnetic shielding, and detected UCN number will all be enhanced compared to previous room temperature Ramsey measurements. Our He co-magnetometer offers unique control of…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
11institutetext: North Carolina State University, Raleigh, NC 27695, USA 22institutetext: Triangle Universities Nuclear Laboratory, Durham, NC 27708, USA 33institutetext: Duke University, Durham, NC 27708, USA 44institutetext: North Carolina Central University, Durham, NC 27707, USA 55institutetext: Arizona State University, Tempe, AZ 85287, USA 66institutetext: University of Kentucky, Lexington, KY 40506, USA 77institutetext: University of Virginia, Charlottesville, VA 22904, USA 88institutetext: Oak Ridge National Laboratory, Oak Ridge, TN 37831, USA 99institutetext: Universidad Nacional Autonóma de Mexico, Apartado Postal 20-364, México 1010institutetext: Bartoszek Engineering, Aurora, IL 60506, USA 1111institutetext: University of Illinois Urbana-Champaign, Champaign, IL 61801, USA 1212institutetext: Massachusetts Institute of Technology, Cambridge, MA 02139, USA 1313institutetext: California Institute of Technology, Pasadena, CA 91125, USA 1414institutetext: Los Alamos National Laboratory, Los Alamos, NM 87545, USA 1515institutetext: Mississippi State University, Mississippi State, MS 39762, USA 1616institutetext: University of Tennessee, Knoxville, TN 37996, USA 1717institutetext: Simon Fraser University, Burnaby, BC, V5A 1S6, Canada 1818institutetext: Tennessee Technological Institute, Cookeville, TN 38501, USA 1919institutetext: Yale University, New Haven, CT 06520, USA 2020institutetext: Indiana University, Bloomington, IN 47408, USA 2121institutetext: Brown University, Providence, RI 02912, USA 2222institutetext: Harvard University, Cambridge, MA 02138, USA 2323institutetext: Valparaiso University, Valparaiso, IN 46383, USA
The neutron electric dipole moment experiment at the Spallation Neutron Source
\firstnameK.K.H. \lastnameLeung\fnsep 1122 [email protected]
\firstname M. \lastname Ahmed 2 2334 4
\firstname R. \lastname Alarcon 5 5
\firstname A. \lastname Aleksandrova 6 6
\firstname S. \lastname Baeßler 7 78 8
\firstname L. \lastname Barrón-Palos 9 9
\firstname L. \lastname Bartoszek 10 10
\firstname D.H. \lastname Beck 11 11
\firstname M. \lastname Behzadipour 6 6
\firstname J. \lastname Bessuille 12 12
\firstname M.A. \lastname Blatnik 13 13
\firstname M. \lastname Broering 6 6
\firstname L.J. \lastname Broussard 8 8
\firstname M. \lastname Busch 2 23 3
\firstname R. \lastname Carr 13 13
\firstname P.-H. \lastname Chu 3 3
\firstname V. \lastname Cianciolo 8 8
\firstname S.M. \lastname Clayton 14 14
\firstname M.D. \lastname Cooper 14 14
\firstname C. \lastname Crawford 6 6
\firstname S.A. \lastname Currie 14 14
\firstname C. \lastname Daurer 11 11
\firstname R. \lastname Dipert 5 5
\firstname K. \lastname Dow 12 12
\firstname D. \lastname Dutta 15 15
\firstname Y. \lastname Efremenko 8 816 16
\firstname C.B. \lastname Erickson 11 11
\firstname B.W. \lastname Filippone 13 13
\firstname N. \lastname Fomin 16 16
\firstname H. \lastname Gao 3 3
\firstname R. \lastname Golub 1 12 2
\firstname C.R. \lastname Gould 1 12 2
\firstname G.L. \lastname Greene 8 816 16
\firstname D.G. \lastname Haase 1 12 2
\firstname D. \lastname Hasell 12 12
\firstname A.I. \lastname Hawari 1 1
\firstname M.E. \lastname Hayden 17 17
\firstname A.T. \lastname Holley 18 18
\firstname R.J. \lastname Holt 13 13
\firstname P.R. \lastname Huffman 1 1228 8
\firstname E. \lastname Ihloff 12 12
\firstname T.M. \lastname Ito 14 14
\firstname J. \lastname Kelsey 12 12
\firstname Y.J. \lastname Kim 14 14
\firstname J. \lastname Koivuniemi 11 11
\firstname E. \lastname Korobkina 1 12 2
\firstname W. \lastname Korsch 6 6
\firstname S.K. \lastname Lamoreaux 19 19
\firstname E. \lastname Leggett 15 15
\firstname A. \lastname Lipman 1 12 2
\firstname C.-Y. \lastname Liu 20 20
\firstname J. \lastname Long 20 20
\firstname S.W.T. \lastname MacDonald 14 14
\firstname M. \lastname Makela 14 14
\firstname A. \lastname Matlashov Present address: IBS Center for Axion and Precision Physics Research 14 14
\firstname J. \lastname Maxwell 12 12
\firstname M. \lastname McCrea 6 6
\firstname M. \lastname Mendenhall 13 13
\firstname H.O. \lastname Meyer 20 20
\firstname R. \lastname Milner 12 12
\firstname P. \lastname Mueller 8 8
\firstname N. \lastname Nouri 6 619 19
\firstname C.M. \lastname O’Shaughnessy 14 14
\firstname C. \lastname Osthelder 13 13
\firstname J.-C. \lastname Peng 11 11
\firstname S. \lastname Penttila 8 8
\firstname N.S. \lastname Phan 14 14
\firstname B. \lastname Plaster 6 6
\firstname J. \lastname Ramsey 8 814 14
\firstname T. \lastname Rao 11 11
\firstname R. P. \lastname Redwine 12 12
\firstname A. \lastname Reid 1 12 2
\firstname A. \lastname Saftah 6 6
\firstname G.M. \lastname Seidel 21 21
\firstname I.F. \lastname Silvera 22 22
\firstname S. \lastname Slutsky 13 13
\firstname E. \lastname Smith 14 14
\firstname W.M. \lastname Snow 20 20
\firstname W. \lastname Sondheim 14 14
\firstname S. \lastname Sosothikul 1 12 2
\firstname T.D.S. \lastname Stanislaus 23 23
\firstname X. \lastname Sun 13 13
\firstname C.M. \lastname Swank 13 13
\firstname Z. \lastname Tang 14 14
\firstname R. \lastname Tavakoli Dinani 17 17
\firstname E. \lastname Tsentalovich 12 12
\firstname C. \lastname Vidal 12 12
\firstname W. \lastname Wei 13 1314 14
\firstname C.R. \lastname White 1 12 2
\firstname S.E. \lastname Williamson 11 11
\firstname L. \lastname Yang 11 11
\firstname W. \lastname Yao 8 8
\firstname A.R. \lastname Young 1 12 2
Abstract
Novel experimental techniques are required to make the next big leap in neutron electric dipole moment experimental sensitivity, both in terms of statistics and systematic error control. The nEDM experiment at the Spallation Neutron Source (nEDM@SNS) will implement the scheme of Golub & Lamoreaux [Phys. Rep., 237, 1 (1994)]. The unique properties of combining polarized ultracold neutrons, polarized 3He, and superfluid 4He will be exploited to provide a sensitivity to . Our cryogenic apparatus will deploy two small () measurement cells with a high density of ultracold neutrons produced and spin analyzed in situ. The electric field strength, precession time, magnetic shielding, and detected UCN number will all be enhanced compared to previous room temperature Ramsey measurements. Our 3He co-magnetometer offers unique control of systematic effects, in particular the Bloch-Siegert induced false EDM. Furthermore, there will be two distinct measurement modes: free precession and dressed spin. This will provide an important self-check of our results. Following five years of “critical component demonstration,” our collaboration transitioned to a “large scale integration” phase in 2018. An overview of our measurement techniques, experimental design, and brief updates are described in these proceedings.
1 Introduction
The existence of a permanent neutron electric dipole moment would be a direct violation of time-reversal () symmetry. This is equivalent to violation of combined charge and parity () symmetry via the theorem Schwinger1951 ; Luders1957 , which has so far withstood experimental scrutiny. violation has been measured in the meson and meson systems Christenson1964 ; Aubert2002 ; Abe2002 and is successfully incorporated in the standard model as a complex phase in the Cabibbo-Kobayashi-Maskawa quark mixing matrix Cabibbo1963 ; Kobayashi1973 . While violation is a key ingredient to generate a baryon-antibaryon asymmetry Sakharov1967 , the amount of known standard model violation is much too small to adequately explain the observed size of this asymmetry in the Universe.
In electroweak baryogenesis mechanisms proposed by minimal supersymmetric extensions to the standard model, the lower limit on the size of the neutron electric dipole moment (EDM) required is to be consistent with the baryon-to-photon ratio measured from the Cosmic Microwave Background Li2009 ; Cirigliano2010 . This is significantly larger than the value inferred from the standard model. Therefore, searching for a neutron EDM down to the level provides a promising route for the discovery of new physics. If an observation was made at this level, it would represent a clear signal above the standard model calculated “background”. Even a negative result would have significant impact as it would rule out a large class of beyond standard model theories.
In 1950 the first neutron EDM experiment was performed using Ramsey’s separated oscillatory fields technique Ramsey1950 on a neutron beam in an applied electric field Smith1957 . The refinement of this method over the next three decades Ramsey1982 improved its sensitivity by four orders-of-magnitude Dress1977 . Since the 1980s, a transition to using stored ultracold neutrons (UCNs) and the implementation of a cohabiting magnetometer species (“co-magnetometer”) to correct for temporal and spatial magnetic field inhomogeneities (while still using Ramsey’s technique) were used to attain a sensitivity improvement of a further two orders-of-magnitude Lamoreaux2009 . The current world limit is (90 % C.L.) Baker2006 ; Pendlebury2015 . In this experiment, UCNs from the PF2 turbine source at the Institut Laue-Langevin (ILL) were used to fill a 21 L cylindrical measurement cell (with a diameter) to reach an initial polarized density of Baker2014 . An electric field of was used, along with a Ramsey free precession time . The choice of is determined by a combination of the UCN storage time and spin coherence times of the UCNs and the co-magnetometer. The detected UCN counts after corresponded to a density in the cell of with a polarization observable (i.e. the contrast of the Ramsey fringes) of .
Relocation of this apparatus from the ILL to a new solid deuterium source at the Paul Scherrer Institute Anghel2009 ; Becker2015 gave a 10% increase in the UCN numbers, but more importantly, improvements in other systems allowed , , and Roccia2018 . The shot noise statistical sensitivity of a neutron EDM experimental cycle is given by:
[TABLE]
where is the number of detected UCNs. Therefore, the sensitivity of the experiment per day improved by more than a factor of three.
There is ample room for increasing the number of UCNs in future neutron EDM experiments. However, there exists a Bloch-Siegert shift induced false EDM Commins1991 ; Pendlebury2004 (often called the “geometric-phase induced frequency shift”) caused by the interaction of the motion-induced field with magnetic field gradients, where is the velocity. For a spin species undergoing diffusive motion (e.g. the co-magnetometer), the size of this false EDM is proportional to , where is the long dimension of the cell. Furthermore, the transverse spin coherence time scales as McGregor1990 . Therefore, there are limits to the extent that can be increased by increasing the size of the cell alone.
To reach a sensitivity new techniques are needed. A novel experimental scheme that can reach this coveted precision in both statistics and systematics was proposed in Golub1994a . Here, unique properties of polarized 3He and superfluid 4He are exploited to allow a neutron EDM experiment to be performed on a high density of polarized UCNs located inside two small measurement cells. This experiment, which we refer to as nEDM@SNS, will be located at the Spallation Neutron Source at Oak Ridge National Laboratory. A more in-depth description of our experiment will be published shortly Filippone2019 .
2 nEDM@SNS experimental scheme
At the heart of the apparatus will be two measurement cells both in length and with cross-sections filled with isotopically pure superfluid 4He cooled to . Along the short horizontal dimension of both cells, a highly homogenous magnetic field will be applied, a value chosen to balance between expected systematic uncertainties and statistical precision. Polarized 3He will be preloaded into the cells at an isotopic concentration of and a polarization . A high voltage electrode will be located between the two cells, while two ground electrodes will be situated on either side (see Fig. 1). Therefore, will be parallel to in one cell and anti-parallel to it in the other.
The FNPB cold neutron beam at the SNS Fomin2015 , collimated and focused to be smaller than the cells’ inner dimensions, will be turned on to produce and “fill” UCNs inside the cell Golub1975 ; Golub1977 . This we call in situ UCN production. The initially white beam will be polarized using a super-mirror and will be chopped to to suppress backgrounds caused by cold neutrons with wavelengths other than at the single-phonon UCN production peak. On the inside of the cell there will be a deuterated polystyrene (dPS) coating, which has a neutron optical potential . The volumetric production rate of polarized UCNs that can be stored by this potential will be .
The UCN density in the cell will build-up with a time constant given by: . These time constants are: the neutron -decay lifetime ; the UCN upscattering loss lifetime set by interactions with the superfluid helium excitations, expected to be at Golub1979 ; Golub1983 ; the UCN wall loss lifetime at the cell walls, where the design goal is ; and the UCN-3He capture lifetime, which during filling will be . Combining these time constants and assuming good match between the neutron beam dimensions with the cell, the saturated UCN density in the cell will thus be . The UCN polarization is expected to be high, , since the cold neutrons retain their polarization after scattering off superfluid 4He phonons, and the UCNs in the opposite spin-state will be absorbed by the 3He during accumulation with a time constant of approximately 200 s.
With in situ UCN production, and also having the UCN spins analyzed in the cell (which we call “in situ spin analysis,” described next in Sec. 2.1), the combined UCNs produced in the two cells can be utilized. This avoids UCN loss and depolarization that can occur during transport from an external UCN source and then to an external spin analysis system. These problems have a long history of degrading the expected performance of UCN experiments. The physics of the super-thermal UCN production process in superfluid helium has been well-studied Golub1983 ; Yoshiki1992 ; Baker2003 ; Zimmer2007a ; Piegsa2014 ; Leung2016 .
Our in situ UCN production and in situ spin analysis techniques provide a large statistical sensitivity increase compared to neutron EDM experiments that use external UCN sources because of the increase in detected UCN number . The other statistical advantages of our experimental design include the suppression of UCN upscattering loss from the cryogenic cell walls and the large electric field Gerhold1998 ; Ito2016 that can be supported by the superfluid helium due to its high dielectric breakdown strength.
For the suppression and control of systematic effects, our cryogenic environment allows the natural implementation of superconducting magnetic shielding, which is particularly useful for suppressing low frequency ambient field drifts. Furthermore, by making small changes to the temperature of the superfluid, the mean free path of the 3He co-magnetometer can be changed drastically. This has important consequences for controlling and suppressing important systematic errors, as will be described in Sec. 2.3. First, the role of the 3He as UCN spin analyzer will be described.
2.1 UCN spin analysis with polarized 3He
UCNs experience the strongly spin-dependent capture reaction . For anti-parallel neutron and 3He nuclear spins, the capture cross-section at thermal energies is . This translates to a cross-section of at a relative velocity of in our experiment. When the spins are parallel, the absorption cross-section is small; the experimental limit on the triplet to singlet absorption cross-section ratio is Passell1966 .
After a capture event, the charged products will produce scintillation light in the superfluid helium at wavelengths of . This EUV light inside the cell will be converted to blue light using deuterated 1,1,4,4-tetraphenyl-1,3-butadiene (dTPB) that is embedded in the dPS polymer matrix on the inner cell walls. Since the UCNs do not leave the measurement cell, the analysis of their spin orientation is performed in situ. The detection of this scintillation light will also give live information on the average relative angle between the UCN and 3He spins throughout a measurement. This is different than the Ramsey technique where the phase of the UCN spins (converted to a longitudinal polarization) is only measured after the end of the precession period. This opens up possibilities for different measurement schemes.
Quantitatively, the average scintillation light event rate will be given by:
[TABLE]
where is the number of UCNs remaining in the cell at time , and are the 3He and UCN polarizations, is the difference between the average phase angle of the two spin species, and is the time-averaged n-3He absorption time constant.
UCN -decay, , produces a time-dependent “background” in . The scintillation light produced from this reaction comes primarily from the electron, which has an energy between [math] and . Due to a combination of the stopping power differences between electrons, protons and tritons, and the effects of the electric field, the n-3He capture peak in the number of EUV photon spectrum is expected to sit on top of a broad -decay bump Ito2012 ; Ito2013 . By making appropriate cuts in the detected spectrum, a proportion of the -decay events can be rejected Archibald2006 . For our scintillation light detection system design (described in Sec. 3), the statistically optimized acceptance probabilities in Eq. 2 are and for capture events and neutron -decay, respectively. The rate of ambient background events that will fall inside the spectrum cuts are included as in Eq. 2.
2.2 Free precession measurement mode
Our experiment can be performed employing two distinct measurement modes. In this section, we describe the free precession mode. (The dressed spin mode is described in Sec. 2.4.) In both modes the measurements start after the cell is filled with UCNs from the cold neutron beam followed by a pulse, an applied AC magnetic field that rotates both the 3He and UCN spins into the plane transverse to .
In the free precession mode, the two spin species will simply be left to freely precess after the pulse. Due to the interaction with the combined magnetic and electric fields, evolves as:
[TABLE]
where and are the neutron and 3He gyromagnetic ratios, is the phase at the start of the measurement, and is the angular precession frequency difference between the neutron and 3He. The ‘’ or ‘’ signs and superscripts are for parallel or anti-parallel to , respectively. For , since and , . A negative frequency corresponds to clockwise motion looking down the axis. Of course, the field would also interact with any 3He EDM, but the size of this effect will be greatly suppressed due to screening by atomic electrons Schiff1963 ; Flambaum2012 .
If we include transverse spin relaxation of the UCNs and 3He, then and . Our design goals for both and are . The scintillation light event rate in the free precession mode will thus be given by:
[TABLE]
Defining , the number of UCNs in the cell will be given by:
[TABLE]
where is the initial number of UCNs. The integral produces small short-lived oscillations in that for the level of description in these proceedings can be neglected so that:
[TABLE]
In the absence of systematic effects, the difference between two measurements when is reversed relative to would be given by:
[TABLE]
To search for a non-zero neutron EDM signal, the difference between extracted from simultaneous measurements in the two cells (i.e. made at different spatial positions) and from measurements as the polarity on the high voltage electrode alternated (i.e. made at different times) will be analyzed.
Monte-Carlo simulations of the scintillation event rate versus time during the free precession period have been made and fitted (see Fig. 2). The fit is performed over the whole free precession measurement time . However, since we are able to observe the live UCN precession with our technique, the frequency analysis can be performed in short time windows. This will be useful for correcting temporal magnetic field drifts (see Sec. 2.3).
At the end of every measurement cycle, the (partially) depolarized 3He will be removed from the cell and new polarized 3He injected into the cell (described in Sec. 3). The field may also be reversed systematically between measurement cycles. The combined time for these operations is the dead time. For the sensitivity discussions next, a dead time of is assumed, along with an ambient background rate of . The influence of these parameters on the sensitivity have been studied. Additionally, is assumed to be fixed in the fitting routine for the discussions below. This requires to be reproducible after the pulse.
Our simulations along with the above assumptions lead to the conclusion that the statistically optimum operating values for extracting from repeated measurement cycles include: (a) a value of that produces (i.e. , assuming ), (b) , and (c) a cold neutron beam fill time . The uncertainty in extracting from these operating parameters is per free precession cycle per cell. For an electric field , the precision in extracting a neutron EDM is for 300 live days of running. We expect to be able to achieve this in around three calendar years.
Before describing the dressed spin measurement mode, we first discuss the role of the 3He as a co-magnetometer and the key systematic effects involved in the free precession mode.
2.3 3He co-magnetometer and systematic effects
When writing Eq. (7), it was implied that the static field in the two measurements are identical. However, since they are made for cells at different spatial positions or in the same cell during different times, corrections for magnetic field gradients and drifts will inevitably be required. Therefore, high-precision magnetometry is needed.
A magnetometer system located outside the measurement cell and away from the large electric field can be used to infer the magnetic field inside the cells. However, such external magnetometry systems are insensitive to fields produced by leakage currents Lamoreaux2009 or magnetization contamination Pendlebury2015 near the cell, both of which can produce sizable systematic effects. For high precision neutron EDM experiments, co-magnetometery is needed.
The polarized 3He located in the same volume as the UCNs can be used as a co-magnetometer. For a concentration of , the 3He number density is (compared with for the UCN density). The precessing 3He magnetization will generate an oscillating field with an amplitude of several fT close to the cell. This will be measured by an array of Superconducting Quantum Interference Device (SQUID) magnetometers near the cell Kim2013 .
While co-magnetometry drastically reduces systematic effects due to field variations, it adds additional effects at a much smaller scale. These come in the form of different ensemble averaged magnetic fields between the co-magnetometer atoms and the UCNs, despite the fact they occupy the same volume, as well as different precession frequency shifts experienced by the two.
The observed in a cell should be written as:
[TABLE]
where and are the ensemble averaged magnetic fields, and and are the frequency shifts of the two species. These effects only arise because of the differences in motion and precession between the co-magnetometer atoms and UCNs.
Firstly, the 3He atoms are in thermal equilibrium with the superfluid helium bath. Their velocity, taking into account the 3He effective mass, will be . The UCNs, however, will have velocities . It is worth mentioning that this velocity difference will be smaller than for other commonly used co-magnetometers at room temperature. Secondly, UCNs undergo ballistic motion reflecting only off the cell walls whereas the 3He motion will be diffusive. At our design concentration the 3He mean-free-path is approximately given by Baym2013 , dominated by scattering off phonons in the superfluid. Thirdly, as already mentioned, the spin precession frequency of the 3He will only be faster than the UCNs. Again, this will be a smaller difference than compared with typical co-magnetometers. Furthermore, it should be noted that the sign of and are both negative. This will suppress effects such as that caused by Earth’s rotation Lamoreaux2009 .
The most serious of these frequency shifts that can produce a false EDM signal comes from an interaction between magnetic field gradients and the motional magnetic field Lamoreaux2005 ; Barabanov2006 ; Golub2011 ; Swank2012a ; Steyerl2014 ; Golub2015 ; Pignol2015 ; Swank2016 . The Bloch-Siegert induced frequency shift is generally larger for the co-magnetometer. The magnitude of for and our electric field will be . This field is transverse to with a direction that fluctuates after every collision with the walls or phonons in the superfluid. This results in Bloch-Siegert frequency shifts Bloch1940 . The combination of this field with magnetic field gradients transverse to produces a frequency shift that is proportional to and hence will appear as a false EDM signal. The size of this frequency shift is also generally dependent on the gyromagnetic ratio, the static field strength , the field gradients, the collisional frequency, and (as already mentioned in the introduction) on the dimensions of the cell. However, closed form expressions exist for the ballistic and highly diffusive cases Pendlebury2004 ; Pignol2015 . We will call this the "Bloch-Siegert induced false EDM”.
A technique to suppress this false EDM effect is to apply a magnetic field gradient and scan its value to interpolate for when the total gradient is zero Pendlebury2015 . For our 3He co-magnetometer, we have another important handle to control this effect. By scanning the temperature of the superfluid helium bath the 3He mean-free-path, and thus the collisional frequency, can be changed drastically without influencing the UCNs. For example a scan will cause the collisional frequency to change by a factor of about 50 because of the strong temperature dependence of the phonon density. Combined with previous theoretical treatments of this problem Lamoreaux2005 ; Barabanov2006 ; Golub2011 ; Swank2012a ; Steyerl2014 ; Golub2015 ; Pignol2015 ; Swank2016 , we have a powerful tool for studying and controlling this important false EDM effect. For instance, we can suppress the false EDM by appropriately tuning the temperature, or reduce the temperature temporarily to magnify this effect and use it to tune out field gradients.
An effect introduced by the polarized 3He is the pseudo-magnetic field experienced by the UCNs Abragam1973 . This comes from the spin-dependence of the real part of the n-3He scattering length Huber2009 ; Zimmer2002 . The pseudo-magnetic field experienced by the UCNs will be parallel to the 3He spins. For and , the magnitude of this pseudo-magnetic field is . The effects of this field can be suppressed by precise operation of the experiment (see Sec. 3.6 about the Systematics and Operational Studies apparatus).
2.4 Dressed spin measurement mode
The second measurement mode is the dressed spin mode, where a strong off-resonant AC magnetic field transverse to with magnitude and angular frequency will be applied Haroche1970 . In the limit , the dressing field causes the gyromagnetic ratio of an unperturbed species to be modified to:
[TABLE]
where is the zeroth-order Bessel function of the first kind. A higher-order treatment of spin dressing and descriptions of the spin motions can be found in Swank2018a ; Tavakoli-Dinani2018 .
By an appropriate choice of dressing field parameters, the effective gyromagnetic ratios of the 3He and UCNs can be made the same, i.e. . This is first satisfied when . For , a dressing frequency can be used.
Considering for now the case with no field, by applying the dressing field satisfying the relative azimuthal angle between the two species can be made to remain fixed. Furthermore, by applying the dressing field slightly above or below (e.g. by shifting ), then can be increased or decreased.
The pseudo-magnetic field experienced by UCNs when will cause their spins to precess out of the transverse plane at a frequency of about the 3He spin axis. This motion can be mitigated by appropriately modulating the dressing field away from at a modulation frequency (e.g. ) to make the motion about symmetric. To a first order approximation, this will allow the out-of-plane motion during one half of the modulation cycle to be reversed during the next half cycle.
The effect of the field on the neutron EDM will be an additional angular frequency so that when at critical dressing the evolution of will be given by:
[TABLE]
in the limit. Here, is some user chosen phase between the neutron and 3He spins.
There are different spin dressing modulation techniques and feedback modes (e.g. for stabilizing the spin dressing field) that can be implemented with different statistics and suppression of systematic drifts. These are described in Refs. Golub1994a ; Swank2018a ; Filippone2019 . With these techniques, the neutron EDM signal (from the additional term in Eq. 10) can be extracted from a single dressed spin measurement in a single cell. Therefore, the SQUID readout of the 3He will not be needed.
The optimized 1 statistical precision from the dressed spin mode with pulsed modulation and using both cells is expected to be for 300 live days of data taking Filippone2019 . The same parameters as in Sec. 2.2 are assumed except for , which is more statistically advantageous for this mode. This sensitivity is better than in the free precession mode. Furthermore, this technique will contain different systematics. For instance, there will be no Bloch-Siegert induced false EDM from gradients in the dressing field, and effects from static field inhomogeneities can be suppressed Swank2018a . The false EDM due to static field gradients will still be present but can be suppressed with temperature tuning Swank2018a , similar to the free precession mode. With the dressed spin mode, precise simultaneous control of the two spin species will be important. The techniques required to achieve this and their associated effects will be studied with the Systematics and Operation Studies apparatus described in Sec. 3.6. Having these two measurement modes with different systematic effects will allow us to provide an important self-verification of our results.
3 Apparatus design
An overview of our apparatus is shown in Fig. (3). It consists of three main modular sub-systems: the Central Detector System, the 3He Services System, and the Magnetic Field Module. These systems are being developed and tested in parallel at different institutes before their integration at the SNS. A brief overview of each system, as well as the Systematics and Operational Studies apparatus, is given below.
3.1 Central detector system
The Central Detector System (CDS), shown in greater detail in Fig. (1), will be a volume filled with superfluid helium cooled with a custom non-magnetic dilution refrigerator to temperatures in the range . This will enable the study of systematic effects as described in Sec. 2.3. The CDS will contain the two measurement cells and associated cell valves, a high voltage electrode, two ground electrodes, a high voltage amplification system, SQUID arrays, and scintillation light detection systems.
Materials inside the CDS volume have stringent requirements in terms of being non-magnetic (to avoid producing field gradients) and non-electrically conductive to avoid Johnson noise in the SQUIDs and eddy current heating caused by the spin dressing field. The volume itself will be constructed from an epoxy-impregnated fiberglass composite material similar to G10. Large scale ( diameter) detachable cryogenic seals made using a Kapton gasket placed between G10 plates have been developed for the seal of the CDS volume. Cryogenic stress-relieved epoxy mounts for large silicon wafers, required for cold neutron beam windows, have also been developed.
The measurement cells are required to have low UCN reflection wall loss probability and low 3He and UCN wall depolarization rates. They must be 3He leak tight (to exclude 3He which exists at natural isotopic abundance levels in the superfluid 4He of the CDS), and be constructed from high purity and deuterated materials where needed.
Full-sized cells made from protonated PMMA plates with the dTPB-doped dPS (dPS+dTPB) coating on the inside have been produced. The neutron optical potential of the coatings have been verified to be with neutron reflectometry. The UCN storage properties of these cells have been characterized using the LANL solid deuterium based UCN source Saunders2013 ; Ito2018 and an apparatus that can cool the cells down to in vacuum. A UCN wall loss time has been demonstrated for 50% of UCNs loaded into the cell from this source with stainless steel guides. The dPS+dTPB coating has also been demonstrated to have low depolarization probability per 3He wall collision () down to Ye2009 ; Yoder2010 in small () cells.
A valve is needed for loading and unloading a cell with 3He (see Sec. 3.2). This valve when closed is required to be UCN and 3He storage friendly and UCN and 3He polarization friendly, similar to the cells. A prototype deuterated plastic cell valve system has also been tested in the cryogenic UCN storage test apparatus. This valve has demonstrated UCN loss at levels well below the wall loss rate. The 3He holding time at this valve has also been demonstrated to be sufficiently long from the results of extrapolating helium gas holding measurements performed down to .
In another apparatus designed for studying electrical breakdown in superfluid helium Ito2016 , a diameter copper implanted PMMA electrode has been demonstrated to support a stable electric field . Being a stochastic process, the dielectric breakdown probability scales with the electrode surface area. The leakage currents measured for PMMA in various geometries inserted in between the electrodes were demonstrated to be low enough to reduce leakage current systematic effects to for pessimistic assumptions.
In order to reach the design field inside the cells, a potential of is required at the high voltage electrode. This is too high for a direct feed system. A suitable high voltage amplification system has been identified in the form of a cryogenic Cavallo multiplier system. Here, charge is transferred to the high voltage electrode in discrete steps by a moving electrode. A room temperature prototype has been used to demonstrate this technique Clayton2018 .
After a capture event approximately EUV photons with wavelengths are expected to be produced in the superfluid helium in the presence of the field Ito2012 . These photons need to be detected with sufficiently high efficiency to enable discrimination against backgrounds (from neutron decay and other sources). The scintillation light detection efficiency has been measured in small () prototype cells coated with dPS+dTPB. These cells were filled with liquid helium and scintillation light was produced with a 210Po -source. The light collection system employed wavelength shifting fibers to convert blue light from the TPB to green light. This green light was then guided to silicon photomultipliers (SiPM). Using the results from these tests for the design of a similar system in the nEDM@SNS apparatus, photoelectrons are expected to be detected by the SiPMs per event. These results were used to determine the values of and described in Sec. 2.1.
SQUIDs will be used to detect the nuclear precession signal of the 3He co-magnetometer. The intrinsic and flux noise of a prototype SQUID system with long pick-up coil leads has been demonstrated to meet the noise requirements at Kim2013 .
3.2 3He services system
The cryostat of the 3He Services (He3S) System, located several meters above the CDS, will provide the temperature gradient (relative to the measurement cell temperature) for loading polarized 3He before each measurement cycle and unloading (partially) depolarized 3He afterwards. This 3He “heat flush” injection and removal technique is described in Baym2015 for 3He concentrations and temperatures relevant to our proposed experiments. Experimental heat flush tests have been performed using superfluid helium with 3He at concentrations of .
The polarized 3He source will be a cryocooler based atomic beam, which was previously used in the experiments described in Esler2007 ; Eckel2012 . With a nozzle cooled to , a cold effusive source of 3He atoms is produced. This unpolarized 3He passes through a permanent magnetic quadrupole that filters out one spin-state, producing a polarized 3He beam with a flux of . This beam will be injected onto the surface of a small superfluid helium filled volume (called the “Injection Volume”) in the He3S cryostat. A spin transport magnetic field system will be used to tailor the magnetic field so as to maintain the 3He polarization during injection. Scattering of the 3He beam by 4He vapor arising from evaporation of the superfluid film from the Injection Volume will be suppressed by a film burner that has been fabricated and successfully tested.
After accumulation, the polarized 3He in the Injection Volume will be transported to the cell. Polarized 3He friendly coatings for the transfer tube to the cell have been developed Yoder2010 . The 3He removed from the cell after each measurement will be concentrated by heat flushing to a small volume. This volume, called the “Sequestration Volume,” will then be isolated, and the liquid inside evaporated and removed with room temperature pumps.
A separate dilution refrigerator will be used for cooling the He3S cryostat. Since this will be located further away from the cells than the refrigerator for the CDS, its non-magnetic requirements are slightly relaxed. This dilution refrigerator is currently under construction and has been cooled to , demonstrating superfluid leak tightness of the components.
3.3 Magnetic field module
The Magnetic Field Module (MFM), which surrounds the CDS, will provide the static field as well as the AC fields (transverse to ) for the rotation and spin dressing. Field gradients cause spin relaxation and give rise to the Bloch-Siegert induced false EDM effect, as discussed in Secs. 2.2 and 2.3. The magnetic field gradient requirements for the field are less than a few , while those for the spin dressing field are less stringent by a factor of about 15.
To produce the uniform horizontal magnetic fields, modified saddle-shaped cos coil designs will be used. In order from smallest radius outwards, the components of the MFM (all of which have a vertical cylindrical geometry) are: the spin dressing coil, an eddy current shield (a thin conductor to reduce heating of external material caused by the spin dressing field), the coil, a ferromagnetic shield (which provides a flux return for the field, improving its uniformity inside the coil), a superconducting Pb magnetic shield (for suppression of time varying fields), and a magnetic cloak (comprising strips of ferromagnetic material to produce a tunable effective permeability, described more in Sec. 3.4). The Pb magnetic shield is closed at the top and bottom with Pb end caps, and with the curved saddle wires of the cos coils located outside. The end caps also serve as magnetic mirrors to increase the effective length of the cos coils, improving field uniformity.
The cryostat for the MFM will have an outer vacuum jacket, a liquid nitrogen cooled shield, and an inner magnet volume (IMV) for cooling the components described above. The outer shell of the IMV will be cooled with liquid helium lines, and inside the IMV will be helium gas to provide heat exchange. This will facilitate operation of the superconducting Pb shield, as well as superconducting Pb alloy coil wires at temperatures below . The inner shell of the IMV is required to be made from a non-electrically conductive material because of eddy current heating and Johnson noise requirements.
The construction and testing of a half scale prototype MFM system has enabled us to demonstrate our ability to produce the required field gradients, maintain adequate shielding, and sustain necessary heat loads Galvan2011 ; Slutsky2017 . The outer aluminum vacuum vessel and liquid nitrogen shields for the final MFM have been fabricated and cryogenically tested.
An array of fluxgate magnetometers will be located between the inner wall of the MFM and the CDS Nouri2014 ; Nouri2014a ; Nouri2015 . This magnetometer array will be used to reconstruct the field at the two measurement cells by solving Laplace’s equation for the magnetic scalar potential () assuming no current or magnetization sources inside the boundary of the magnetometer array (i.e. ). This system will be used for initial tuning of magnetic field gradients.
3.4 Magnetic shield enclosure
A multi-layered ferromagnetic Magnetic Shield Enclosure (MSE) with internal dimensions approximately will be located outside the MFM. A magnetically sealed door will provide internal access. Specifications for the MSE include field gradients over a volume at the cells, and dynamic shielding factors at and between and . A tri-axial system of rectangular coils will surround the MSE to compensate for the Earth’s field and other background fields.
A set of coils that produce a uniform field parallel to will be located inside the MSE. This uniform field will help maintain 3He polarization during transport from the He3S System. It will also help reduce field distortions caused by gaps in the superconducting Pb shield end caps of the MFM. Along the side wall of the Pb shield, the magnetic field of the internal MSE coils will encounter the flux repelling superconductor. The ferromagnetic cloak inside the MFM (mentioned earlier) will be used to cancel the effects of the superconductor in order to match the field inside and outside the Pb shield.
3.5 Cold neutron beam
The cold neutron beam will pass through the MSE and the MFM, and enter the CDS to reach the superfluid helium inside the measurement cells in order to produce UCNs. The beam line has been redesigned to incorporate a super-mirror for polarization and a chopper system for monochromation to , rather than the intercalated graphite Bragg-reflection monochrometers initially proposed Fomin2015 . Despite the fact that the revised neutronics system will require longer guides, better optimization of angles and beam cross-sections will improve the integrated cold neutron flux by . This will provide a volumetric UCN production rate in the cells of for our trapping potential ( for dPS minus for the superfluid helium).
3.6 Systematics and operational studies apparatus
The Systematics and Operational Studies (SOS) apparatus will be used to perform measurements required for finalizing the design of various components and protocols of the nEDM@SNS experiment. It will be a smaller apparatus with a cryogenic turnaround time of approximately two weeks, as compared to two months for the main apparatus.
The design of the SOS apparatus does not include a high voltage system and also contains only one measurement cell. This reduces the volume of superfluid helium that is required from to . Polarized 3He will be provided by a room-temperature metastability exchange optical pumping (MEOP) based source. This source enables 3He to be loaded to concentrations that are several orders of magnitude higher compared with the atomic beam source, at the expense of a lower . These higher concentrations will be useful for certain measurements that don’t require long UCN storage times. The UCNs will be injected from an external source rather than being produced in situ. The magnetic field gradient requirements in the SOS apparatus are also relaxed by a factor of three relative to the main apparatus. These design specifications make the SOS apparatus easier to implement.
The key measurements that will be performed with the SOS apparatus will include:
Characterizing the UCN storage properties, 3He and UCN wall spin relaxation times, and fluorescence properties of the measurement cells prior to installation in the nEDM@SNS apparatus. 2. 2.
Characterizing UCN and 3He motional correlation functions which are essentially unaffected by the -field, and which can then be used to validate and control the Bloch-Siegert induced false EDM. 3. 3.
Identifying techniques required for the high-precision simultaneous control of the two spin species, such as those required for the pulse and spin dressing.
Note in regard to the latter that spin dressing of 3He in an atomic beam Esler2007 ; Eckel2012 and in a cell Chu2011a has been previously demonstrated. Similarly, critical spin dressing on 1H and 19F in a liquid sample has also been performed Tavakoli-Dinani2018 . However, the duration and precision of control in these experiments are several orders of magnitude below what is required for the nEDM@SNS experiment.
The SOS apparatus will enable us to divert key studies from the critical path of the main nEDM@SNS experiment and will decrease the time needed to reach a physics result. The design of the SOS apparatus and the associated measurement program will be described in an upcoming paper Leung2018 .
The aluminum cryostat needed for the SOS apparatus has been constructed and commissioned to . The MEOP source achieved the design goal of inside the optical pumping cell. The dilution refrigerator that will be used has been cooled and observed to deliver the cooling power required. Various demanding cryogenic components have been fabricated and tested, such as superfluid leak tight cryogenic components and seals made only from plastics, and components for the 3He removal system.
4 Conclusion
The nEDM@SNS experiment will search for a neutron EDM down to the level which is compelling for tests of electroweak baryogenesis and beyond standard model physics. This unprecedented sensitivity will be reached using a high UCN density produced directly in superfluid helium inside the measurement cell, as well as by performing in situ and live UCN spin analysis using polarized 3He. The relatively small measurement cell size combined with a unique cryogenic 3He co-magnetometer allows many key systematic effects to be controlled and sufficiently suppressed, in particular the Bloch-Siegert induced false EDM effect. The apparatus allows two distinct measurement techniques with different systematics: the free precession mode and the dressed spin mode. These two modes of operation exhibit different sensitivities to effects giving rise to statistical and systematic uncertainties. This will provide the opportunity for a powerful self-check of our results. The combined systematic error of our experiment is expected to be below .
The nEDM@SNS collaboration has devoted significant effort to the development of the apparatus and techniques for the realization of this pioneering experiment proposed by Golub1994a . Over the last five years we have experimentally demonstrated technical readiness of many high fidelity prototypes and standalone final components. The valuable lessons learned have been applied to an engineering design that we believe will reduce schedule risks. In mid-2018, we transitioned to a “large scale integration” phase during which we will make large-scale procurements, construct final components, and integrate them at the SNS. The acquisition of first physics data is expected to begin toward the end of 2023. The apparatus is expected to reach statistical precision after only 1 week of measurement time.
5 Acknowledgements
This work was supported in part by the U.S. Department of Energy under grants DE-AC02-06CH11357, DE-AC05-00OR22725, DE-AC52-06NA25396, DE-FG02-94ER40818, DE-FG02-97ER41042, DE-FG02-99ER41101, DE-SC0014622, DE-SC0008107, DE-SC0005367, DE-FG02-88ER40416, and 2017LANLEEDM, and the U.S. National Science Foundation under grants 1306547, 1205977, 1506459, 1812340, 1440011, 1506416, 1822502, 1812377, and 1506451.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1(1) J. Schwinger, Phys. Rev. 82 , 914 (1951)
- 2(2) G. Lüders, Ann. Phys. (N. Y.) 2 , 1 (1957)
- 3(3) J.H. Christenson, J.W. Cronin, V.L. Fitch, R. Turlay, Phys. Rev. Lett. 13 , 138 (1964)
- 4(4) B. Aubert, D. Boutigny, J.M. Gaillard, A. Hicheur, Y. Karyotakis, J.P. Lees, P. Robbe, V. Tisserand, A. Zghiche, A. Palano et al. (The BABAR Collaboration), Phys. Rev. Lett. 89 , 201802 (2002)
- 5(5) K. Abe, K. Abe, T. Abe, I. Adachi, H. Aihara, M. Akatsu, Y. Asano, T. Aso, V. Aulchenko, T. Aushev et al. (Belle Collaboration), Phys. Rev. D 66 , 071102 (2002)
- 6(6) N. Cabibbo, Phys. Rev. Lett. 10 , 531 (1963)
- 7(7) M. Kobayashi, T. Maskawa, Progr. Theor. Exp. Phys. 49 , 652 (1973)
- 8(8) A.D. Sakharov, Pisma Zh. Eksp. Teor. Fiz. 5 , 32 (1967), [JETP Lett. 5 (1967) 24] [Sov. Phys. Usp. 34 (1991) 392]