# Noncontrast free-breathing respiratory self-navigated coronary artery   cardiovascular magnetic resonance angiography at 3 T using lipid insensitive   binomial off-resonant excitation (LIBRE)

**Authors:** Jessica A.M. Bastiaansen, Ruud B. van Heeswijk, Matthias Stuber,, Davide Piccini

arXiv: 1903.01325 · 2019-07-22

## TL;DR

This study demonstrates that LIBRE RF pulses provide superior fat suppression and improve coronary artery imaging quality in free-breathing 3T MRI compared to traditional methods, enhancing the robustness of coronary MRA.

## Contribution

The paper introduces the use of lipid insensitive binomial off-resonant excitation (LIBRE) pulses for self-navigated coronary MRA at 3T, showing improved fat suppression and image quality.

## Key findings

- LIBRE achieved near complete fat suppression in vitro and in vivo.
- LIBRE significantly increased RCA vessel sharpness and length.
- LIBRE outperformed conventional fat saturation and water excitation methods.

## Abstract

Robust and homogeneous lipid suppression is mandatory for coronary magnetic resonance angiography (MRA) since coronary arteries are commonly embedded in fat. However, effective large volume lipid suppression becomes challenging when performing radial whole-heart coronary MRA and the problem may even be exacerbated at increasing magnetic field strengths. Incomplete fat suppression also generates artifacts, and may affect advanced motion correction methods. The aim was to evaluate a recently reported lipid insensitive MRI method for self-navigated coronary MRA at 3T. Lipid insensitive binomial off resonant excitation (LIBRE) radiofrequency (RF) excitation pulses were included into a self-navigated 3D radial GRE coronary MRA sequence at 3T. LIBRE was compared against conventional fat saturation (FS) and binomial 1-180{\deg}-1 water excitation (WE). First, fat suppression of all techniques was numerically characterized using Matlab and experimentally validated in phantoms and in legs of human volunteers. Subsequently, free-breathing self-navigated coronary MRA was performed using the LIBRE pulse as well as FS and WE in ten volunteers. Results obtained in the simulations were confirmed by the experimental validations as LIBRE enabled near complete fat suppression for 3D radial imaging in vitro and in vivo. For self-navigated whole-heart coronary MRA at 3T, fat SNR was significantly attenuated using LIBRE compared with conventional FS. LIBRE increased the RCA vessel sharpness significantly (37 +/- 9% (LIBRE) vs. 29 +/- 8% (FS) and 30 +/- 8% (WE), both p<0.05) and led to a significant increase in the measured RCA vessel length to (83 +/- 31 mm (LIBRE) vs. 56 +/- 12 mm (FS) and 59 +/- 27 (WE) p<0.05). LIBRE enables robust large volume fat suppression and significantly improves coronary artery image quality at 3T compared to the use of conventional fat suppression and water excitation.

---
Source: https://tomesphere.com/paper/1903.01325