Au-Ge alloys for wide-range low-temperature on-chip thermometry
J. R. A. Dann, P. C. Verpoort, J. Ferreira de Oliveira, S. E. Rowley,, C. J. B. Ford, G. J. Conduit, and V. Narayan

TL;DR
This paper introduces a Au-Ge alloy that functions effectively as a resistance-based thermometer from room temperature down to 0.2 K, exhibiting a logarithmic resistivity dependence suitable for low-temperature on-chip thermometry.
Contribution
The study demonstrates the potential of Au-Ge alloys as practical, wide-range low-temperature thermometers with favorable sensitivity and resistivity characteristics, and explains the underlying Kondo-like behavior.
Findings
Resistivity shows logarithmic temperature dependence from room temperature to 0.2 K.
The alloy's sensitivity compares favorably to commercial sensors.
Kondo-like behavior explains the resistivity's temperature dependence.
Abstract
We present results of a Au-Ge alloy that is useful as a resistance-based thermometer from room temperature down to at least \SI{0.2}{\kelvin}. Over a wide range, the electrical resistivity of the alloy shows a logarithmic temperature dependence, which simultaneously retains the sensitivity required for practical thermometry while also maintaining a relatively modest and easily-measurable value of resistivity. We characterize the sensitivity of the alloy as a possible thermometer and show that it compares favorably to commercially-available temperature sensors. We experimentally identify that the characteristic logarithmic temperature dependence of the alloy stems from Kondo-like behavior induced by the specific heat treatment it undergoes.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Au-Ge alloys for wide-range low-temperature on-chip thermometry
J. R. A. Dann
Cavendish Laboratory, University of Cambridge, J.J. Thomson Avenue, Cambridge CB3 0HE, UK
P. C. Verpoort
Cavendish Laboratory, University of Cambridge, J.J. Thomson Avenue, Cambridge CB3 0HE, UK
J. Ferreira de Oliveira
Cavendish Laboratory, University of Cambridge, J.J. Thomson Avenue, Cambridge CB3 0HE, UK
Centro Brasileiro de Pesquisas Físicas, Rua Dr Xavier Sigaud 150, Rio de Janeiro, 22290-180, Brazil
S. E. Rowley
Cavendish Laboratory, University of Cambridge, J.J. Thomson Avenue, Cambridge CB3 0HE, UK
Centro Brasileiro de Pesquisas Físicas, Rua Dr Xavier Sigaud 150, Rio de Janeiro, 22290-180, Brazil
A. Datta
Department of Materials Science, University of Cambridge, 27 Charles Babbage Road, Cambridge CB3 0FS, UK
S. Kar-Narayan
Department of Materials Science, University of Cambridge, 27 Charles Babbage Road, Cambridge CB3 0FS, UK
C. J. B. Ford
Cavendish Laboratory, University of Cambridge, J.J. Thomson Avenue, Cambridge CB3 0HE, UK
G. J. Conduit
Cavendish Laboratory, University of Cambridge, J.J. Thomson Avenue, Cambridge CB3 0HE, UK
V. Narayan
Cavendish Laboratory, University of Cambridge, J.J. Thomson Avenue, Cambridge CB3 0HE, UK
Abstract
We present results of a Au-Ge alloy that is useful as a resistance-based thermometer from room temperature down to at least . Over a wide range, the electrical resistivity of the alloy shows a logarithmic temperature dependence, which simultaneously retains the sensitivity required for practical thermometry while also maintaining a relatively modest and easily-measurable value of resistivity. We characterize the sensitivity of the alloy as a possible thermometer and show that it compares favorably to commercially-available temperature sensors. We experimentally identify that the characteristic logarithmic temperature dependence of the alloy stems from Kondo-like behavior induced by the specific heat treatment it undergoes.
I Introduction
Measuring the temperature () dependence of material properties is crucial in building up our understanding of condensed matter systems. It is also important for the control and measurement of electronic devices, particularly quantum devices, that operate at low . Cooling materials to very low has often revealed rich and unexpected physics with relevance for current and future technologies. Understanding and quantifying low-temperature phenomena requires effective thermometry across the entire range of interest.
A limitation of present-day thermometry schemes is the absence of a single sensor spanning a broad range without loss of sensitivity. It is therefore customary to use multiple thermometers, each of which is designed for use in different parts of the range of interest. For example, in resistance ()-based thermometry, in which is inferred from the electrical resistance of a ‘thermistor’, metallic or semiconducting sensors are used depending on the range in question: high is most conveniently measured using metallic thermistors which, however, lose sensitivity below 5\text{,}\mathrm{K}$$, where is dominated by -independent impurity scattering. At low , semiconducting thermistors (e.g., Ge, RuO2, carbon glasses) are employed, in which the increases as is lowered with activated behavior, . Yet despite the large sensitivity that such thermistors offer, the rapidly increasing can raise substantial issues with measurements that avoid self-heating of the thermometer.
A variety of other techniques are also available for thermometry at low , such as nuclear orientation thermometry Marshak (1983), magnetic thermometry Schuster et al. (1994), shot-noise thermometry Spietz et al. (2003), Johnson-noise thermometry White et al. (1996), Coulomb-Blockade thermometry Pekola et al. (1994); Kauppinen et al. (1998), capacitance thermometry Lawless (1971), lithographically-defined on-chip thermocouples Chickering et al. (2009); Narayan et al. (2012); Billiald et al. (2015); Narayan et al. (2014), non-invasive charge-sensing thermometers Mavalankar et al. (2013), and normal-superconductor-based tunnel junctions Feshchenko et al. (2015); Saira et al. (2016).
Here we present a novel thermistor which offers several of the individual advantages presented by these techniques, namely: 1) it is sensitive over a broad -range; 2) its resistance remains easily measurable and small enough to allow efficient thermalization and avoid self-heating; 3) it can be lithographically patterned to reliably measure spatial gradients and/or be fabricated on chips as part of microelectronic devices.
The thermistor we report is made of a Au-Ge alloy ( ) and might be suitable as a -based thermometer that can operate over a wide range from up to , and possibly even up to room . In particular, we demonstrate that for a range of compositions shows a logarithmic dependence on across a range of three orders of magnitude. The material satisfies the conditions: 1) the sensitivity remains usably high between room and with increasing with lowering , 2) the absolute value of remains sufficiently small that self-heating should be negligible, and thermalization should not be an issue, 3) this material is suitable for lithographic -scale fabrication, and thus for use on chips as part of nano-scale devices, as well as on larger scales for use in other environments, such as cryogenic refrigerators, and, in addition, 4) the linearity of as a function of allows for a relatively simple calibration.
II Methods
The alloy was fabricated into Hall bar devices using a photolithography and lift-off technique. The required thicknesses of Ge and Au were thermally evaporated in separate layers and the Hall bars were then heated in a reducing environment of nitrogen and hydrogen at 450\text{,}\mathrm{\SIUnitSymbolCelsius}$$ for two minutes before being rapidly cooled to room within a few minutes, as shown in Fig. S1. Finally, Ohmic Ti/Au contacts were deposited in a second photolithography step. The composition of the material was measured in two separate ways: 1) the thickness of each layer was measured using a crystal monitor during the evaporation; and 2) energy dispersive X-ray spectroscopy measurements (see Fig. S2) were carried out after the heat treatment. The compositions from the two methods were consistent within 0.2%. The Hall bars were approximately long, wide, and thick. Several different substrates have been successfully employed, namely silicon, sapphire, and GaAs. Unless otherwise stated, all measurements presented here were carried out on samples deposited on silicon substrates. The results are not observed to be affected by the choice of three substrates, and showed the same main characteristics over multiple separate processing batches. Resistance measurements were carried out using standard four-terminal low-frequency lock-in techniques. Unless otherwise stated, a frequency of was used and the excitation current was . The measurements reported here were performed in three different cryostats with base temperatures of , , and . While the qualitative nature of the results was unaffected between cryostats, in order to make quantitative comparisons of the results (Fig. 1), we have introduced multiplicative factors of 1-2% in Fig. 1 to ensure that data from different cryostats falls on the same curve.
III Results and Discussion
Figure 1 shows the characteristics of a film. Over most of the range, the electrical resistivity (here and are the width and length of the Hall bar respectively, and is the film thickness) displays a logarithmic dependence increasing by 30\text{,}\mathrm{\SIUnitSymbolMicro\SIUnitSymbolOhm}\text{,}\mathrm{m}$$ per decade in . The data for the main panel of this figure has been taken in two separate cryostats to span the full range. Notably, changes monotonically from room down to , below which the dependence begins to flatten out. Above , is constant (and so is linear) until , where there is a distinct change in slope. Between these limits the data is fitted to
[TABLE]
where and are fitting parameters depending on the exact composition of the alloy.
Not only is the logarithmic form of applicable over a wide range, but also shows little deviation from this functional form. This accuracy in the form of should make this material useful for -based thermometry, as any deviations adversely impact the accuracy of the measured . Furthermore, the simplicity and accuracy of Eq. (1) makes calibration straightforward. Calibrating a thermometer would only require measurements at two well-defined values of , for example when immersed in liquid nitrogen and in liquid helium. More calibration points could improve the calibration and/or extend the range over which it is calibrated.
The relatively straightforward calibration offsets an unexpected cooldown dependence of R. As shown in Fig. S3, we observe a non-systematic variation of R between cooldowns. While this effect is small, it is certainly within experimental resolution. The origin of this effect is unexplained as of now, and may have to do with the differential thermal expansion of the film with respect to the substrate. However, we can rule out the role of atomic diffusion as the resistance of the samples is stable when stored at room temperature for several weeks or months. Likewise, the resistance does not vary when maintained at sub-Kelvin temperatures, implicating the temperature change as responsible for the observed cooldown dependence. We have noticed that encapsulating the film in a few hundred of Ge substantially improves the stability. However, even in the absence of capping, a simple two-point calibration should mitigate the error introduced by this variation.
Figure 2 shows that down to 0.2\text{,}\mathrm{K} the sensitivity $S\equiv\derivative*{R}{T}$ of the $\mathrm{Au}_{0.07}\mathrm{Ge}_{0.93}$ alloy as a thermometer increases as $T\rightarrow$0\text{\,}\mathrm{K}. Although this dependence is found even in semiconductor materials, in those the advantage gained due to the increasing is offset by the correspondingly large values of . This is the primary difference between semiconductors and the alloy, in which the absolute value of remains moderate. Therefore, we examine not only but also the ‘dimensionless sensitivity’ which, importantly, is found to vary by less than a factor of 2 over three decades in . This is to be contrasted with commercially-available thermometers, in which the dimensionless sensitivity typically varies by more than an order of magnitude over comparable scales, restricting the range over which they can be used lak . The weak dependence of the dimensionless sensitivity on suggests not only that a single sensor element can be used over a wide range, but also that the same electronic instrument required to measure can be used over the whole range.
It is instructive to consider how the fractional error in a reading varies with . Normally, the dimensionless sensitivity gives an indication of this, assuming that the fractional error in the resistance measurement remains constant throughout the range of operation. However, since of the Au-Ge film changes only relatively slowly with , changing by less than one order of magnitude over the entire range studied, it is reasonable to consider the case of resistance measurements having a fixed uncertainty over the entire temperature range. Using Eq. (1), this leads to an uncertainty in the measured of . Over the entire range of applicability of Eq. (1) the fractional uncertainty in the measurement is therefore constant. Therefore, the Au-Ge alloys described here have the potential to be used to make thermometers that are equally useful at low and relatively high .
We now explore how changing the composition affects the behavior of the material. A similar logarithmic dependence has been observed in samples with a variety of compositions of (). Figure 3 shows that is the largest for , for which the uncertainties in the fitting parameters are also smallest. While this suggests that both and the accuracy may be increased by further reducing , we also note that this would increase the absolute value of .
What is the response of the alloy to a magnetic field ? As shown in Fig. 4 for 1\text{,}\mathrm{K} the films show an approximately linear magnetoresistance at high $B$ Dreizin and Dykhne ([1973](#bib.bib16)); Stroud and Pan ([1976](#bib.bib17)); Parish and Littlewood ([2005](#bib.bib18)) with weak anti-localization at small $B$. This allows, in principle, for a simple calibration-based read-out of $T$ in the presence of a finite $B$. We also note that at $1.8\text{\,}\mathrm{K}$ the magnetoresistance is $\approx$0.5\text{\,}\mathrm{\char 37\relax}\text{/}\mathrm{T}, implying that the induced error is 0.13\text{,}\mathrm{K}\text{/}\mathrm{T}$$. Below , the -response of becomes more complex with showing non-monotonic behavior and slow transients similar to that seen in Narayan et al. (2019); Verpoort and Narayan (2019).
The post-deposition heat treatment of the alloy has a crucial impact on its performance as a thermometer. Figure 5 shows for a sample of , with the material deposited at the same time as for the sample shown in Fig. 1, but importantly with the only difference being that it was annealed at , which is below the eutectic point of Au-Ge alloys (\approx$$360\text{\,}\mathrm{\SIUnitSymbolCelsius}) Okamoto and Massalski (1984). The sample in Fig. 5 does not show a logarithmic , but instead is closer to a power-law dependence, consistent with a previous work that suggested the use of alloys as thermometers, where the films were heated at Béthoux et al. (1995). Clearly, the overall characteristics are far less satisfactory in terms of how sensitivity and accuracy vary with temperature compared to those shown in Fig. 1.
The observed logarithmic is clearly correlated to the heat treatment of the films. Cochrane et al. Cochrane et al. (1975) previously reported logarithmic dependences in moderately-disordered materials and explained their findings based on disorder-induced, non-magnetic localized microscopic degrees of freedom, which act as sources for Kondo-type scattering events. These localized degrees of freedom correspond to different configurations of atomic positions that are separated by potential wells and have a finite tunnelling amplitude. Unlike in the case of localized magnetic spins, the resulting eigenstates are not perfectly degenerate, but instead have their energy levels split by . This gives a resistance dependence of
[TABLE]
where is the electron bandwidth. Consequently, the logarithmic behavior ceases when , and becomes independent of .
The visibly-disordered character of the films suggests that the observed logarithmic arises due to similar physics, as indicated by transmission electron micrograph (TEM) images (inset Fig. 3) that reveal that the films are phase-separated. Consideration of the phase diagram Okamoto and Massalski (1984) suggests that the two phases are Au-rich (containing Ge) islands and an almost pure Ge matrix. This is consistent with energy dispersive X-ray spectroscopy measurements. Generally, it is observed that as the fraction of Au is decreased, the logarithmic fit for becomes better (Fig. 3), and this might suggest that the logarithmic conductivity arises dominantly through the Ge matrix. In samples with large contiguous regions of Au-rich material, conduction through the Au-rich regions becomes more important, and causes a correction to the dependence. We speculate that the Ge matrix is slightly disordered, presumably due to the rapid cooling after the heat treatment, and/or due to the presence of Au impurities that disturb the crystallinity. The flattening of at low (Fig. 1) would suggest that 0.2\text{,}\mathrm{K} in the films. The observed ‘shoulder’ in at 200\text{,}\mathrm{K}$$ is also consistent with the observations in Ref. Cochrane et al. (1975) and we speculate that these can arise due to higher energy levels of the Kondo-type scattering centre. Notably, over the explored range we see no indication of a minimum in , which is expected at a certain , depending on the concentration of Kondo-type scatterers Kondo (1964). Separately, we also note that the observed linear magnetoresistance (Fig. 4) is consistent with the picture of microscopic disorder playing a pivotal role in the transport, as linear magnetoresistance is characteristic of disordered materials Dreizin and Dykhne (1973); Stroud and Pan (1976); Parish and Littlewood (2005).
We note that there are other possible sources of a logarithmic temperature dependence, such as granular systems Beloborodov et al. (2007) and thin disordered films Lee and Ramakrishnan (1985), but these apply to electrical conductivity rather than resistivity as seen here. As shown in Fig. S4, clearly fits a -dependence more consistently.
IV Conclusion
In conclusion we have presented a alloy with clear practical advantages for simple -based thermometry from room down to 0.2\text{,}\mathrm{K}$$. Principally, the use of a single sensor over this range eliminates the necessity of matching calibrations of different thermometers at the limits of their applicability, while also reducing costs towards multiple sensors and the associated wiring and measurement apparatus. Furthermore, the reasonably low values of over the entire range alleviates the need for apparatus that can measure with high precision and without appreciable self-heating. The possibility of lithographically patterning the alloy suggests that -scale thermometers can be deposited directly onto substrates of choice, allowing for thermal and thermoelectric measurements on small samples and at low . The accuracy of the logarithmic -dependence over a wide range allows for a simple 2 point calibration. We find that the characteristic logarithmic behavior of the alloy is linked to its composition and the precise conditions of preparation, in particular annealing above the eutectic temperature. We suggest that the mechanism driving the logarithmic resistance dependence on temperature is the same as that reported previously in moderately-disordered metallic alloys, namely where the disorder induces Kondo-like scattering centres. Strikingly, however, the range over which we observe the logarithmic behavior is significantly larger than in any previous report we are aware of, with the anticipated ‘upturn’ in resistance not occurring up to room temperature.
V Acknowledgements
J.R.A.D., P.C.V., G.J.C., and V.N. acknowledge funding from the Engineering and Physical Sciences Research Council (EPSRC), UK. G.J.C. and S.E.R. acknowledge funding from the Royal Society, UK. J.F.O. would like to thank the Brazilian Agencie CNPq. A.D. and S.K-N. acknowledge financial support through an ERC Starting Grant (Grant No. ERC-2014-STG-639526, NANOGEN). We thank J. J. Rickard for support with TEM.
The reference list from the paper itself. Each links out to its DOI / PubMed record.
- 1Marshak (1983) H. Marshak, Journal of Research of the National Bureau of Standards 88 , 175 (1983).
- 2Schuster et al. (1994) G. Schuster, D. Hechtfischer, and B. Fellmuth, Reports on Progress in Physics 57 , 187 (1994).
- 3Spietz et al. (2003) L. Spietz, K. Lehnert, I. Siddiqi, and R. Schoelkopf, Science 300 , 1929 (2003).
- 4White et al. (1996) D. White, R. Galleano, A. Actis, H. Brixy, M. De Groot, J. Dubbeldam, A. Reesink, F. Edler, H. Sakurai, R. Shepard, et al. , Metrologia 33 , 325 (1996).
- 5Pekola et al. (1994) J. Pekola, K. Hirvi, J. Kauppinen, and M. Paalanen, Physical review letters 73 , 2903 (1994).
- 6Kauppinen et al. (1998) J. Kauppinen, K. Loberg, A. Manninen, J. Pekola, and R. Voutilainen, Review of scientific instruments 69 , 4166 (1998).
- 7Lawless (1971) W. Lawless, Review of Scientific Instruments 42 , 561 (1971).
- 8Chickering et al. (2009) W. Chickering, J. Eisenstein, and J. Reno, Physical review letters 103 , 046807 (2009).