# Evidence of fast neutron sensitivity for 3He detectors and comparison   with Boron-10 based neutron detectors

**Authors:** Giacomo Mauri, Francesco Messi, Kalliopi Kanaki, Richard Hall-Wilton,, Francesco Piscitelli

arXiv: 1902.09870 · 2020-03-13

## TL;DR

This study measures and compares the fast neutron sensitivity of 3He detectors to boron-10 detectors, revealing significantly higher sensitivity in 3He detectors which impacts their background rejection capabilities.

## Contribution

First estimation of 3He detector sensitivity to fast neutrons up to 10 MeV, with comparative analysis against boron-10 detectors and validation through simulations.

## Key findings

- 3He detectors have a sensitivity of approximately 4x10^-3 to fast neutrons.
- 3He detectors exhibit two orders of magnitude higher sensitivity than boron-10 detectors.
- Experimental results are supported and validated by simulation data.

## Abstract

The 3He-based neutron detectors are no longer the default solution for neutron scattering applications. Both the inability of fulfilling the requirements in performance, needed for the new instruments, and the shortage of 3He, drove a series of research programs aiming to find new technologies for neutron detection. The characteristics of the new detector technologies have been extensively tested to prove their effectiveness with respect to the state-of-the-art technology. Among these, the background rejection capability is crucial to determine. The signal-to-background ratio is strongly related to the performance figure-of-merit for most instruments. These are designed to exploit the high flux expected from the new high intensity neutron sources. Therefore, an inadequate background rejection could significantly affect the measurements, leading to detector saturation and misleading events. This is of particular importance for the kind of techniques in which the signals are rather weak. For the first time, the sensitivity of 3He detectors to fast neutrons, up to En = 10 MeV, has been estimated. Two independent measurements are presented: a direct calculation based on a subtraction method used to disentangle the thermal and the fast neutron contribution, while a further evidence is calculated indirectly through a comparison with the recently published data from a 10B-based detector. Both investigations give a characterization on the order of magnitude for the sensitivity. A set of simulations is presented as well in order to support and to validate the results of the measurements. A sensitivity of 4x10-3 is observed from the data. This is two orders of magnitude higher than that previously observed in 10B-based detectors.

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/1902.09870/full.md

## Figures

31 figures with captions in the complete paper: https://tomesphere.com/paper/1902.09870/full.md

## References

50 references — full list in the complete paper: https://tomesphere.com/paper/1902.09870/full.md

---
Source: https://tomesphere.com/paper/1902.09870