Study of uniformity of characteristics over the surface for triple GEM detector
S. Chatterjee, S. Chakraborty, S. Roy, S. Biswas, S. Das, S. K. Ghosh,, S. K. Prasad, S. Raha

TL;DR
This paper investigates the uniformity of key performance characteristics such as gain, energy resolution, and count rate across a triple GEM detector's surface using Fe55 X-ray source and standard electronics.
Contribution
It provides a detailed methodology and experimental results on the uniformity of a triple GEM detector's characteristics, which is essential for detector calibration and performance assessment.
Findings
Uniformity of gain across the detector surface
Energy resolution consistency over active area
Count rate variation and stability
Abstract
Study of the uniformity of gain, energy resolution and count rate over active area of a triple GEM detector has been performed using a strong Fe55 X-ray source with premixed gas of Argon and CO2 in 70/30 ratio and conventional NIM electronics. The detail method of measurement and experimental results are presented in this article.