# Reconstruction Algorithm Design for Mitigating the Orientation Dependent   Conspicuity of Fiber-Like signals in Digital Breast Tomosynthesis

**Authors:** Sean D. Rose, Ingrid Reiser, Emil Y. Sidky, Xiaochuan Pan

arXiv: 1803.04554 · 2018-03-14

## TL;DR

This paper investigates how fiber-like structures in digital breast tomosynthesis exhibit orientation-dependent visibility and proposes reconstruction algorithm modifications to reduce this effect while balancing depth resolution.

## Contribution

The study introduces a reconstruction algorithm design that mitigates orientation-dependent conspicuity of fiber-like signals in DBT, revealing a tradeoff with depth resolution.

## Key findings

- Orientation-dependent signal conspicuity exists in DBT.
- Algorithm design can reduce orientation dependence.
- Tradeoff between minimizing dependence and preserving depth resolution.

## Abstract

There are a number of clinically relevant tasks in digital breast tomosynthesis (DBT) involving the detection and visual assessment of fiber-like structures such as Cooper's ligaments, blood vessels, and spiculated lesions. Such structures can exhibit orientation dependent variations in conspicuity. This study demonstrates the presence of in-plane orientation-dependent signal conspicuity for fiber-like signals in DBT and shows how reconstruction algorithm design can mitigate this phenomenon. We uncover a tradeoff between minimizing orientation-dependence and preserving depth resolution that is dictated by the regularization strength employed in reconstruction.

## Full text

_Full body text omitted from this summary view._ Fetch the complete paper as Markdown: https://tomesphere.com/paper/1803.04554/full.md

## Figures

3 figures with captions in the complete paper: https://tomesphere.com/paper/1803.04554/full.md

## References

5 references — full list in the complete paper: https://tomesphere.com/paper/1803.04554/full.md

---
Source: https://tomesphere.com/paper/1803.04554