Prototype high angular resolution LEKIDs for NIKA2
Shibo Shu, Martino Calvo, Samuel Leclercq, Johannes Goupy, Alessandro, Monfardini, Eduard Driessen

TL;DR
This paper presents a prototype of high angular resolution LEKID arrays for NIKA2, demonstrating an 8% in-lab resolution improvement with smaller pixels, promising enhanced on-sky performance.
Contribution
Development and experimental comparison of a new LEKID pixel design with smaller inductor size for improved angular resolution in NIKA2.
Findings
8% in-lab resolution improvement
Potential 10.2" on-sky FWHM resolution
Prototype validation of pixel size reduction
Abstract
The current resolution of the NIKA2 260 GHz arrays is limited by the inductor size on the individual pixels. In view of future updates of the instrument, we have developed a prototype array with smaller pixels that is experimentally compared to the current pixel design. We find an in-lab improvement increase of the angular resolution of 8%, promising an on-sky FWHM resolution of 10.2" using this new pixel design.