Upgrade of compact neutron spectrometer for high flux environments
M. Osipenko, A. Bellucci, V. Ceriale, D. Corsini, G. Gariano, F., Gatti, M. Girolami, S. Minutoli, F. Panza, M. Pillon, M. Ripani, D.M. Trucchi

TL;DR
This paper presents a new compact Li6-based neutron spectrometer designed for high flux environments, featuring improved charge collection, radiation hardness, and resolution, tested successfully in reactor and neutron generator settings.
Contribution
The paper introduces a novel neutron spectrometer with enhanced performance and robustness, utilizing commercial diamond crystals and RF preamplification for high flux applications.
Findings
Achieved 72 keV energy resolution (RMS)
Attained 68 ps coincidence timing resolution (RMS)
Validated performance through tests in reactor and neutron generator environments
Abstract
In this paper a new version of Li6-based neutron spectrometer for high flux environments is described. The new spectrometer was built with commercial single crystal Chemical Vapour Deposition diamonds of electronic grade. These crystals feature better charge collection as well as higher radiation hardness. Ohmic metal contacts were deposited on the diamonds suppressing build-up of space charge observed in the previous prototypes. New passive preamplification of signal at detector side was implemented to improve the resolution. This preamplification is based on RF transformer not sensitive to high neutron flux. Compact mechanical design allowed to reduce detector size to a tube of 1 cm diameter and 13 cm long. The spectrometer was tested in thermal column of TRIGA reactor and at DD neutron generator. The test results indicate an energy resolution of 72 keV (RMS) and coincidence timing…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.