Evaluation of Direct Haptic 4D Volume Rendering of Partially Segmented Data for Liver Puncture Simulation
Andre Mastmeyer, Dirk Fortmeier, Heinz Handels

TL;DR
This study evaluates a novel haptic rendering approach for liver puncture simulation using partially segmented data, demonstrating low force errors and high plausibility across numerous test paths.
Contribution
It introduces and quantitatively assesses a new direct needle force volume rendering method for liver puncture simulation using partially segmented data.
Findings
Mean force error of 0.12 N RMS compared to gold-standard models
Only 12% of forces affected by errors along test paths
Haptic rendering plausibility confirmed on over 31,000 test paths
Abstract
This work presents an evaluation study using a force feedback evaluation framework for a novel direct needle force volume rendering concept in the context of liver puncture simulation. PTC/PTCD puncture interventions targeting the bile ducts have been selected to illustrate this concept. The haptic algorithms of the simulator system are based on (1) partially segmented patient image data and (2) a non-linear spring model effective at organ borders. The primary aim is to quantitatively evaluate force errors caused by our patient modeling approach, in comparison to haptic force output obtained from using gold-standard, completely manually-segmented data. The evaluation of the force algorithms compared to a force output from fully manually segmented gold-standard patient models, yields a low mean of 0.12 N root mean squared force error and up to 1.6 N for systematic maximum absolute…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.