# Electron Lifetime Determination in Semiconductor Gamma Detector Arrayes

**Authors:** Uri Lachish

arXiv: 1701.03459 · 2017-01-16

## TL;DR

This paper presents a modified analysis method for accurately determining electron lifetime in semiconductor gamma detector arrays, addressing limitations of the traditional Hecht equation.

## Contribution

It introduces a revised expression for mobility-lifetime calculation that improves accuracy over standard methods for detector array measurements.

## Key findings

- The modified expression yields more accurate electron lifetime values.
- Standard Hecht equation is inadequate for array measurements.
- Improved analysis enhances detector performance evaluation.

## Abstract

Hecht equation is not adequate to analyzing standard measurements, of the mobility-lifetime product, carried out with single pixels of detector arrays. A modified expression is calculated, in order to have correct mobility-lifetime values out of experimental data.

---
Source: https://tomesphere.com/paper/1701.03459