High perfomance selectable value transportable high dc Voltage standard
Flavio Galliana, Roberto Cerri, Luca Roncaglione Tet

TL;DR
This paper presents a high-performance, transportable high dc voltage standard covering 10 V to 100 V, designed for laboratory and inter-laboratory use, with enhanced accuracy and stability over existing calibrators.
Contribution
Development of a portable high dc voltage standard with improved accuracy, stability, and transportability for use in calibration and inter-laboratory comparisons.
Findings
Better noise performance than existing calibrators
Enhanced short-mid-term stability
Superior transportability and insensitivity
Abstract
At National Institute of Metrological Research (INRIM), a selectable-value Transportable High dcVoltage Standard (THVS) operating in the range from 10 V to 100 V in steps of 10 V, was developed. This Standard was built to cover the lack of high level dc Voltage Standards at voltages higher than 10 V to employ as laboratory (local) or travelling Standards for Inter-Laboratory Comparisons (ILCs). A ground-mobile electronic technique was used to enhance the accuracy of the THVS at the higher values. The THVS shows better noise, better short-mid-term stability than top level dc Voltage and multifunction calibrators (MFCs) and better suitability and insensibility to be transported than these instruments. The project is extensible to 1000 V.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsAdvanced Electrical Measurement Techniques · Analog and Mixed-Signal Circuit Design · Physics of Superconductivity and Magnetism