Long Time Stability of the Energy Scale Calibration of a Quantum 2000
Uwe Scheithauer

TL;DR
This paper demonstrates that the energy scale calibration of the Quantum 2000 XPS instrument remains stable over an 8-year period when calibrated according to ISO 15472 standards, ensuring reliable measurements.
Contribution
It adapts ISO 15472 calibration procedures to the Quantum 2000 and evaluates its long-term stability over eight years.
Findings
Energy scale remained stable over 8 years
Calibration procedures adapted successfully
Consistent energy differences measured over time
Abstract
According to the international standard ISO 15472 the energy scale of an XPS instrument, type Physical Electronics Quantum 2000, was calibrated. It is shown, how the procedures of the ISO 15472 were adapted to the hardware and software design of the Quantum 2000. The long time stability of the energy scale calibration of the XPS instrument was investigated. The instrumented was operated with a satisfying energy scale calibration over a period of 8 years. All the time energy differences between certain peaks could be measured with the chosen precision of the energy scale calibration.