Improved Sputter Depth Resolution in Auger Thin Film Analysis Using In Situ Low Angle Cross-Sections
Uwe Scheithauer

TL;DR
This paper introduces a novel in situ low angle cross-section technique using standard Auger microprobe instruments, significantly improving depth resolution in thin film analysis.
Contribution
It presents a new method for depth profiling that enhances resolution by sputtering at nearly grazing incidence with a mask, enabling better analysis of thin films.
Findings
Significant improvement in depth resolution delta z / z.
Effective in situ preparation of low angle cross-sections.
Applicable with standard Auger microprobe instruments.
Abstract
A none conventional approach for depth profiling of thin film systems with enhanced depth resolution has been developed using standard Auger microprobe instruments. For the preparation of an in situ low angle cross-section, the sample is partly covered by an appropriate mask. Utilising the edge of this mask, the sample is sputtered in the Auger microprobe with ions at nearly grazing incidence. In the shadow of the mask, this produces a low angle cross section through the thin film system. Then, a conventional depth profile is measured at the point of interest where part of the thin film system is covered only by a thin top layer. As demonstrated a considerable improvement of depth resolution delta z / z can be obtained by this method.