Experimental studies of THGEM in different Ar/CO2 mixtures
Zhan-ying He, Jian-rong Zhou, Zhi-jia Sun, Gui-an Yang, Hong Xu,, Yan-feng Wang, Qian Liu, Hong-bang Liu, Shi Chen, Yi-gang Xie, Yang-heng, Zheng, Xiao-dong Wang, Xiao-dong Zhang, Bi-tao Hu, Yuan-Bo Chen

TL;DR
This study investigates the performance of a domestic THGEM detector in various Ar/CO2 gas mixtures, focusing on gain, energy resolution, and counting rate plateau, to inform neutron detection applications.
Contribution
It provides detailed experimental data on THGEM performance in different Ar/CO2 mixtures, aiding the development of neutron detectors.
Findings
Gain range from 100 to 1000 in tested mixtures
Wide counting rate plateau observed in Ar/CO2(90%/10%)
Energy resolution about 22% at 5.9 keV
Abstract
In this paper, the performances of a type of the domestic THGEM (THick Gaseous Electron Multiplier) working in the Ar/CO2 mixtures are reported in details. This kind of single THGEM can provide the gain range from 100 to 1000, which is very suitable for the application in the neutron detection. In order to study its basic characteristics as the references for the development of THGEM based neutron detector, the counting rate plateau, the energy resolution and the gain of the THGEM have been measured in the different Ar/CO2 mixtures with the change of the electrical fields. For the Ar/CO2(90%/10%) gas mixture, a wide counting rate plateau is got from 720V to 770V with the plateau slope of 2.4% / 100 V and the excellent energy resolution about 22% is obtained at the 5.9keV full energy peak of the 55Fe X-ray source.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.