Study on the imaging ability of the 2D neutron detector based on MWPC
Tian LiChao, Chen YuanBo, Tang Bin, Zhou JianRong, Qi HuiRong, Liu, RongGuang, Zhang Jian, Yang GuiAn, Xu Hong, Chen DongFeng, Sun ZhiJia

TL;DR
This paper presents the design, fabrication, and testing of a 2D neutron detector based on MWPC and 3He convertor, demonstrating high spatial resolution and linearity in neutron detection.
Contribution
The paper introduces a novel 2D neutron detector with optimized spatial resolution and validated performance through experimental testing.
Findings
Spatial resolution of 1.18 mm achieved
Good linearity in neutron detection confirmed
Experimental results align with theoretical predictions
Abstract
A 2D neutron detector based on 3He convertor and MWPC with an active area of 200 mm \times 200 mm has been successfully designed and fabricated. The detector has been tested with Am/Be neutron source and with collimated neutron beam with wavelength of {\lambda} = 1.37 {\AA}. A best spatial resolution of 1.18 mm (FWHM) and good linearity were obtained. This is in good agreement with the theoretical calculations.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.
Taxonomy
TopicsNuclear Physics and Applications · Radiation Detection and Scintillator Technologies · Medical Imaging Techniques and Applications