Simulations of a micro-PET System based on Liquid Xenon
A. Miceli, J. Glister, A. Andreyev, D. Bryman, L. Kurchaninov, P. Lu,, A. Muennich, F. Retiere, V. Sossi

TL;DR
This paper simulates a high-resolution microPET system using liquid xenon, demonstrating its potential for improved imaging performance with high spatial resolution and sensitivity in preclinical applications.
Contribution
It introduces a novel liquid xenon-based microPET design and evaluates its imaging performance through detailed simulations, highlighting its high resolution and sensitivity capabilities.
Findings
NECR peaks at 1326 kcps for mouse phantoms
Sensitivity of 12.6% at the center of FOV
Spatial resolution <= 0.8 mm throughout FOV
Abstract
The imaging performance of a high-resolution preclinical microPET system employing liquid xenon as the gamma ray detection medium was simulated. The arrangement comprises a ring of detectors consisting of trapezoidal LXe time projection ionization chambers and two arrays of large area avalanche photodiodes for the measurement of ionization charge and scintillation light. A key feature of the LXePET system is the ability to identify individual photon interactions with high energy resolution and high spatial resolution in 3 dimensions and determine the correct interaction sequence using Compton reconstruction algorithms. The simulated LXePET imaging performance was evaluated by computing the noise equivalent count rate, the sensitivity and point spread function for a point source, and by examining the image quality using a micro-Derenzo phantom according to the NEMA-NU4 standard. Results…
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.