Calibration and Interpixel Capacitance of a H2RG(2Kx2K) Near-IR Detector
G. Smadja, C. Cerna, A. Ealet

TL;DR
This paper presents a precise calibration of a H2RG near-IR detector, including noise analysis, non-linearity correction, and an extended correlation method to measure interpixel capacitance with high accuracy.
Contribution
It introduces an extended correlation method for interpixel capacitance measurement and achieves sub-percent accuracy in calibration.
Findings
Interpixel capacitance x = -0.0263 +/- 0.0020 (stat) +/- 0.0040 (syst)
Noise analysis and non-linearity corrections improve measurement precision
All measurements are consistent within sub-percent accuracy.
Abstract
A temporal analysis of the noise is performed, and non linearities are taken into account. We then extend the correlation method to groups of several pixels to derive the interpixel capacitance of a detector, found to be x = -0.0263 +/- 0.0020 (stat) +/- 0.0040 (syst). All measurements are consistent to a sub-percent accuracy.
Peer Reviews
No public reviews on file for this paper yet. If you reviewed it on a platform where reviews are public (OpenReview, ICLR, NeurIPS, ICML), you can paste yours below so the community can read it here.
Videos
No videos yet. Explain this paper in a talk, walkthrough, or lecture? Add one.